Дрегер оксилог 1000 инструкция по применению

Dräger Medical Oxylog 1000 Technical Documentation Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Technical Documentation

Oxylog 1000

Emergency Ventilator

Revision 4.0

5503.400

9036037

Because you care

loading

Related Manuals for Dräger Medical Oxylog 1000

Summary of Contents for Dräger Medical Oxylog 1000

  • Page 1
    Technical Documentation Oxylog 1000 Emergency Ventilator Revision 4.0 5503.400 9036037 Because you care…
  • Page 3: Table Of Contents

    Contents General Symbols and Definitions Notes Function Description General information on the Oxylog 1000 Components of the Oxylog 1000 Pneumatic Control (old Oxylog 1000) Gas Supply ……………………… 10 Frequency Control …………………… 11 Air Mix/No Air Mix Selector Switch ………………12 Safety Devices ……………………

  • Page 4
    General points concerning the pressure regulator …………..34 Removing/replacing the pressure regulator …………….34 Schematics and Diagrams Flow chart of the Oxylog 1000 (old) Flow chart of the Oxylog 1000 (new) Tubing Diagram of the Oxylog 1000 (old) Tubing Diagram of the Oxylog 1000 (new)
  • Page 5
    General…
  • Page 7
    Oxylog 1000 General Symbols and Defini- tions WARNING A WARNING statement provides important information about a poten- tially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION A CAUTION statement provides important information about a potentially…
  • Page 8
    General Oxylog 1000 Know-how contained in this Technical Documentation is subject to ongoing change through research and development and Dräger Medical reserves the right to make changes to this Technical Documentation without notice. NOTE Unless otherwise stated, reference is made to laws, regulations or stan- dards (as amended) applicable in the Federal Republic of Germany for equipment used or serviced in Germany.
  • Page 9
    Function Description…
  • Page 11
    General information The Oxylog 1000 is a time-controlled, volume-constant emergency and trans- on the Oxylog 1000 port ventilator for patients weighing 7.5 kg or above. The Oxylog 1000 has the operating mode: controlled ventilation (IPPV = Intermittent Positive Pres- sure Ventilation).
  • Page 12
    Function Description Oxylog 1000 Components of the The Oxylog 1000 comprises the following components: Oxylog 1000 – Pressure reducer – ON/OFF switch – Inspiratory/expiratory switching valve – Minute volume control valve – Frequency control – Safety valve – Emergency air valve –…
  • Page 13
    Oxylog 1000 Function Description Figure 2 Block diagram of the Oxylog 1000 5503.400…
  • Page 14: Pneumatic Control (Old Oxylog 1000)

    The compressed gas is supplied from the central supply system (CS) or from an O cylinder with pressure regulator. The operating pressure of the Oxylog 1000 is between 2.7 bar and 6 bar. CAUTION Risk of damage to equipment! Dust, oil and humidity in medical gases may impair the functioning of the equipment or damage it.

  • Page 15: Frequency Control

    Oxylog 1000 Function Description Frequency Control When the unit is off, driving gas flows via the manifold and through the dosage valve D2 to the 3/2-way valve DMR5 and to DMR6. In consequence, DMR6 switches through and the switching valve WV1 switches to expiration.

  • Page 16: Air Mix/No Air Mix Selector Switch

    Function Description Oxylog 1000 Air Mix/No Air Mix The setting of the Air Mix/No Air Mix selector switch determines the O con- Selector Switch centration of the supply gas for the patient. When the Air Mix/No Air Mix selector switch is set to «Air Mix», the injector (I) takes in ambient air through the 3/2-way valve (WV2).

  • Page 17: Safety Devices

    The display range is -10 mbar to 80 mbar. 3.4.3 Emergency Air Valve The spontaneously breathing patient can breathe in ambient air through the emergency air valve if the Oxylog 1000 has failed due to a fault. 5503.400…

  • Page 18
    If the ventilation pressure is higher than 10 mbar, the 3/2-way valve (DMR7) switches. The 3/2-way valves (DMR9, DMR10 and DMR11) vent to atmo- sphere. The visual and audible alarms are suppressed. Ventilation pressure DMR10 DMR11 DMR9 DMR7 Figure 6 Flow chart of the Oxylog 1000: Disconnect Alarm 5503.400…
  • Page 19
    (DMR8) switches the 3/2-way valve (DMR18). The vent valve (EV) opens the supply gas path to atmosphere. This condition persists until the next inspiratory phase. DMR18 DMR12 DMR13 DMR14 DMR8 Figure 7 Flow chart of the Oxylog 1000: Stenosis alarm 5503.400…
  • Page 20
    Function Description Oxylog 1000 3.4.6 Gas Supply Failure If the gas supply fails or falls below a defined pressure, the Oxylog 1000 Warning (Pv) delivers two alarms: – Visual alarm indicator «Pv» – Audible alarm (alarm generator) The alarm is triggered under the following conditions: –…
  • Page 21
    Oxylog 1000 Function Description Alarm Silence When the pushbutton valve (T1) is pressed the 3/2-way valve (DMR15) fills the volume (V3) and activates the 3/2-way valve (DMR16). The 3/2-way valve (DMR17) vents the alarm generator to atmosphere. The alarm generator (L1) is inactive up to the switching point of the volume (V3) (approx.
  • Page 22
    Function Description Oxylog 1000 Pressure Sensor During ventilation, the pressure sensor controls and switches the stenosis and disconnection of the patient. The pressure sensor comprises two diaphragms. The ventilation pressure acts on the undersides of the diaphragms. The supply pressure (Insp.) is cen- tered on the craters on the pressure sensor during the inspiration phase.
  • Page 23: Pressure Sensor

    Oxylog 1000 Function Description Location of Oxylog 1000 (old) Assem- blies Figure 11 Top view of the open Oxylog 1000 (old) Table 1 3/2-way valve (DMR7) Pressure gauge Indicator (Pv) 3/2-way valve (DMR8) Indicator (Silence) Air Mix/No Air Mix Pressure regulator…

  • Page 24: Pneumatic Control (New Oxylog 1000)

    The compressed gas is supplied from the central supply system (CS) or from an O cylinder with pressure regulator. The operating pressure of the Oxylog 1000 is between 2.7 bar and 6 bar. CAUTION Risk of damage to equipment! Dust, oil and humidity in medical gases may impair the functioning of the equipment or damage it.

  • Page 25: Frequency Control

    When the unit is switched off, valve WV1 and switch S1 are closed. No venti- lation is carried out and there is no pressure at the frequency control. If Oxylog 1000 is switched on using switch S1, the gas flows through switch S1 and manifold VL1 to DMR1 and DMR2 of the frequency control.

  • Page 26: Air Mix/No Air Mix Selector Switch

    Function Description Oxylog 1000 Air Mix/No Air Mix The setting of the Air Mix/No Air Mix selector switch determines the O con- Selector Switch centration of the supply gas for the patient. When the Air Mix/No Air Mix selector switch is set to «Air Mix», the injector (I) takes in ambient air through the 3/2-way valve (WV2).

  • Page 27: Safety Devices

    The display range is -10 mbar to 80 mbar. 5.4.3 Emergency Air Valve The spontaneously breathing patient can breathe in ambient air through the emergency air valve if the Oxylog 1000 has failed due to a fault. 5503.400…

  • Page 28
    The 3/2-way valves (DMR9, DMR10 and DMR11) vent to atmo- sphere. The visual and audible alarms are suppressed. DMR18 Ventilation pressure DMR9 DMR10 DMR11 DMR14 DMR1 DMR3 DMR7 DMR8 DMR12 DMR13 DMR2 DMR15 DMR16 DMR17 Figure 15 Flow chart of the Oxylog 1000: Disconnect alarm 5503.400…
  • Page 29
    This condition persists until the next inspiratory phase. DMR18 DMR9 DMR10 DMR11 DMR14 DMR1 DMR3 DMR7 DMR8 DMR12 DMR13 DMR2 DMR15 DMR16 DMR17 Figure 16 Flow chart of the Oxylog 1000: Stenosis alarm 5503.400…
  • Page 30
    Function Description Oxylog 1000 5.4.6 Gas Supply Failure If the gas supply fails or falls below a defined pressure, the Oxylog 1000 Warning (Pv) delivers two alarms: – Visual alarm indicator «Pv» – Audible alarm (alarm generator) The alarm is triggered under the following conditions: –…
  • Page 31
    Oxylog 1000 Function Description Alarm Silence When the pushbutton valve (T1) is pressed the 3/2-way valve (DMR15) fills the volume (V3) and activates the 3/2-way valve (DMR16). The 3/2-way valve (DMR17) vents the alarm generator to atmosphere. The alarm generator (L1) is inactive up to the switching point of the volume (V3) (approx.
  • Page 32: Pressure Sensor

    Function Description Oxylog 1000 Pressure Sensor During ventilation, the pressure sensor controls and switches the stenosis and disconnection of the patient. The pressure sensor comprises two diaphragms. The ventilation pressure acts on the undersides of the diaphragms. The supply pressure (Insp.) is cen- tered on the craters on the pressure sensor during the inspiration phase.

  • Page 33: Location Of Oxylog 1000 (New) Assemblies

    Oxylog 1000 Function Description Location of Oxylog 1000 (new) Assem- blies Figure 20 Top view of the open Oxylog 1000 (new) Table 2 3/2-way valve (DMR8) Indicator (Silence) On switch 3/2-way valve (DMR7) Alarm control valve Ventilator block Alarm silence…

  • Page 34
    Function Description Oxylog 1000 5503.400…
  • Page 35: Maintenance Procedures

    Maintenance Procedures…

  • Page 37: General Notes

    Maintenance Procedures General notes WARNING Danger of infection! Clean and disinfect the Oxylog 1000, or the rele- vant parts, according to hospital regulations prior to any servicing work or before returning the appliance or parts of it for repair pur- poses.

  • Page 38: Pressure Regulator

    Remove the scews (1) and washers. • Remve the grub screws (2). • Withdraw the connection socket. Connection socket View of the Oxylog 1000: Removing screws Fig. 1: • Withdraw the drawer unit from the housing of the Oxylog 1000. 5503.400…

  • Page 39
    Maintenance Procedures • Remove the screws (1). Fig. 2: Bottom view of the Oxylog 1000: Removing screws • Detach the small silicone hose (hose connection: supply block to pres- sure gauge) from the hose connection socket of the supply block.
  • Page 40
    Press in the circlip on the hose connection socket, at the same time with- drawing the PA tube (hose connection: supply block to working block). Supply block Pressure regulator Working block Circlip Pressure gauge Figure 3 View of the drawer unit of the Oxylog 1000 5503.400…
  • Page 41
    Secure the connection socket for the ventilation hose by the grub screw on the drawer unit of the Oxylog 1000. • Secure the connection socket for the compressed gas hose by the grub screw on the drawer unit of the Oxylog 1000. 5503.400…
  • Page 42
    Set the Air Mix/No Air Mix selector switch to „No Air Mix“. • Set the main switch of the Oxylog 1000 to „I“. With a flow of approx. 55 L/min (indicated on the flowmeter) the connected pressure gauge shows a pressure of approx. 1.5 bar.
  • Page 43
    Detach the hose from the pressure gauge and re-connect it to the con- nection socket of the main switch. • Check that there is no leakage in the hose system of the Oxylog 1000. • Install the drawer unit in the reverse sequence into the housing of the Oxylog 1000.
  • Page 44
    Maintenance Procedures Oxylog 1000 5503.400…
  • Page 45: Schematics And Diagrams

    Schematics and Diagrams…

  • Page 47: Flow Chart Of The Oxylog 1000 (Old)

    Oxylog 1000 Schematics and Diagrams Flow chart of the Oxylog 1000 (old) Figure 1 Flow chart of the Oxylog 1000 (old) 5503.400…

  • Page 48
    Schematics and Diagrams Oxylog 1000 Table 1 Key to flow chart (old) Compressed-gas connection Ventilation valve Restrictor D1 to D7 Pressure gauge DMR1 to DMR18 3/2-way valve Pressure regulator Vent valve Filter Frequency control valve Injector Alarm generator Minute volume control valve…
  • Page 49: Flow Chart Of The Oxylog 1000 (New)

    Oxylog 1000 Schematics and Diagrams Flow chart of the Oxylog 1000 (new) Figure 2 Flow chart of the Oxylog 1000 (new) 5503.400…

  • Page 50
    Schematics and Diagrams Oxylog 1000 Table 2 Key to flow chart (new) Compressed-gas connection Ventilation valve Restrictor D1 to D4 Pressure gauge DMR1 to DMR18 3/2-way valve Pressure regulator Vent valve Filter Frequency control valve Injector Alarm generator Minute volume control valve…
  • Page 51: Tubing Diagram Of The Oxylog 1000 (Old)

    Oxylog 1000 Schematics and Diagrams Tubing Diagram of the Oxylog 1000 (old) Figure 3 Tubing diagram of the Oxylog 1000 (old) 5503.400…

  • Page 52
    Schematics and Diagrams Oxylog 1000 Table 3 Key to the tubing diagram of the Oxylog 1000 (old) Silence control ON/OFF switch Frequency control Alarm logic Patient block Microswitch pin configuration Volume Pre-tubing Tubing for final assembly on total unit Plug…
  • Page 53: Tubing Diagram Of The Oxylog 1000 (New)

    Oxylog 1000 Schematics and Diagrams Tubing Diagram of the Oxylog 1000 (new) Figure 4 Tubing diagram of the Oxylog 1000 (new) 5503.400…

  • Page 54
    Schematics and Diagrams Oxylog 1000 5503.400…
  • Page 55: Annex

    Annex Parts catalog Test List…

  • Page 57
    Parts catalog Oxylog 1000 Revision: 2006-03-14 09:56:56 5503.400 Because you care Because you care Emergency Care — Perioperative Care — Critical Care — Perinatal Care — Home Care…
  • Page 59
    Product concerned Parts catalog Item Qty. Part No. Description Qty. Remark unit 5703300 Caddy 1.000 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5503.400 Revision: 2006-03-14 09:56:56…
  • Page 60
    Accessories/Consumables Parts catalog Item Qty. Part No. Description Qty. Remark unit 5703303 BAG OXYLOG 1000 FOR CADDY 1.000 5703304 BAG OXYLOG 2000 FOR CADDY 1.000 5703305 Bag for CompactCaddy 1.000 5703306 PROTECTION COVER 1.000 5703307 Carrying Belt 1.000 5704216 All-round Wall Holder 1.000…
  • Page 61
    Caddy Parts catalog Item Qty. Part No. Description Qty. Remark unit 5703310 CLAW_FOR_CADDY 1.000 5703329 PROTECTION BAR HIGH FOR CADDY 1.000 5703328 PROTECTION BAR LOW FOR CADDY 1.000 5703341 CYLINDER FIXATION SET 1.000 5703336 PROTECTION FOOT CADDY 1.000 5703340 GUIDING RAIL SET FOR CADDY 1.000 5703367 PROTECTION STRIP CADDY…
  • Page 62
    CompactCaddy Parts catalog Item Qty. Part No. Description Qty. Remark unit 5703310 CLAW_FOR_CADDY 1.000 5703333 GUIDING RAIL FOR COMPACTCADDY 1.000 5703335 PROTECTION FOOT Compact CADDY 1.000 Items that are shown in the illustration but are not listed below the illustration are not available as spare parts 5503.400 Revision: 2006-03-14 09:56:56…
  • Page 63
    Alduk II O2-DM G3/4 2M86632 Alduk II O2-DM Pin Index 2M86678 All-round Wall Holder 5704216 Bag for CompactCaddy 5703305 BAG OXYLOG 1000 FOR CADDY 5703303 BAG OXYLOG 2000 FOR CADDY 5703304 Breathing hose 150 cm 2M86511 Carrying Belt 5703307 COVER…
  • Page 65
    Test List…
  • Page 67
    Testing vent valve ……………………9 Testing I:E ratio ……………………10 Testing ventilatory frequency ………………..10 Testing minute volume ………………….11 Testing alarm logic and silence control …………….. 12 Testing O concentration …………………. 16 Deliver the Oxylog 1000 to the operators ready for operation. Tested…
  • Page 68
    Checking the I:E Ratio ………………….26 Checking the Ventilatory Frequency ………………26 Checking the Minute Volume ………………..27 Checking Warning Logics and Silence Control …………..28 Checking the O Concentration ……………….. 32 Supply Oxylog 1000 to user ready for operation. Tested…
  • Page 69
    Some function blocks of Oxylog 1000 have been modified. Therefore, this Test List is divided in two parts. Test List 1 (Oxylog 1000 «old») and Test List 2 (Oxylog 1000 «new»). Oxylog 1000 «new» can be recognized by its serial number (ARNL —…
  • Page 70
    Test List Oxylog 1000 5503.400…
  • Page 71
    Test List 1 (Oxylog 1000 „old“) Oxylog 1000 Folder no.: Serial no.: 5503.400 ————————— Device location: Issue: ————————— 02.99 This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.
  • Page 72
    Stopwatch Hose clamp Sealing plug This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer. Page 4 5503.400 Oxylog 1000 Test List…
  • Page 73: Function Tests

    Checking general condition of Oxylog 1000 • Check that the accompanying documents are with the Oxylog 1000. • Check the following components for damage: Housing of the Oxylog 1000 Ventilation pressure gauge Ventilation hose Ventilation valve Compressed gas connecting hose 1.5 m…

  • Page 74: Testing Supply Pressure And Safety Valve

    • Unscrew the compressed gas connecting hose from the connection socket (compressed gas supply). • Remove the drawer unit of the Oxylog 1000 by removing the screws with washers (1). • Remove the screws (2). • Withdraw the connection sockets (ventilation hose and compressed gas supply).

  • Page 75
    0 to 6 bar Main switch Fig. 2: Oxylog 1000 with test set-up: Testing the supply pressure • Set the frequency selector (Freq.) to 10 1/min. • Set the minute volume adjuster (MV) to full right. • Set the ventilation pressure (Paw) to approx. 60 mbar (hPa).
  • Page 76
    • Close off the hose to the double-membrane relay with the hose clamp; see following illustration. Fig. 3: View of the drawer unit of the Oxylog 1000: Clamping the hose • Set the minute volume adjuster (MV) to full right.
  • Page 77: Testing Vent Valve

    Test pressure reducer Flowmeter 0 to 10 L/min Fig. 4: Oxylog 1000 with test set-up: Testing the vent valve • With the test pressure reducer feed a flow of 10 L/min into the connection socket for the ventilation hose. The maximum pressure rise is 8 mbar (hPa) (read off the pressure rise from the pressure gauge of the Oxylog 1000).

  • Page 78: Testing I:e Ratio

    • Turn the maximum ventilation pressure (Paw) to full right. • Set the Air Mix/No Air Mix selector switch to „No Air Mix“. • Switch on the Oxylog 1000 at the main switch. • Use a stopwatch to determine the I:E ratio.

  • Page 79: Testing Minute Volume

    Adjustable Flowmeter resistor Fig. 5: Oxylog 1000 with test set-up: Testing the minute volume • Set the ventilation pressure selector (Paw) to full right. • Set the frequency selector (Freq.) to 10 1/min. • Set the minute volume adjuster (MV) variably; see following table.

  • Page 80: Testing Alarm Logic And Silence Control

    Ventilation valve Test lung Fig. 6: Oxylog 1000 with test set-up to test the supply pressure alarm • Set the minute volume adjuster (MV) to approx. 10 L/min. • Set the frequency selector (Freq.) to 10 1/min. • Set the ventilation pressure selector (Paw) to approx. 40 mbar (hPa).

  • Page 81
    „rot“ indicator (Paw low) lights up. The „Paw low“ switching point is at approx. 10 mbar (hPa). (Read off the switching point from the pressure gauge of the Oxylog 1000 and from the separate pressure gauge). This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.
  • Page 82
    Ventilation hose resistor T-piece Fig. 8: Oxylog 1000 with test set-up: Testing the stenosis alarm • Set the minute volume adjuster (MV) to 10 L/min. • Set the frequency selector (Freq.) to 10 1/min. • Set the Air Mix/No Air Mix selector switch to „Air Mix“.
  • Page 83
    • Set the frequency selector (Freq.) to 10 1/min. • Detach the test lung from the ventilation valve. The Oxylog 1000 emits an audible alarm. The „Paw low“ indicator lights up red. • During the inhalation phase press the „Alarm silence“ key.
  • Page 84: Testing O Concentration

    Vol% Test adapter Test lung Ventilation valve Fig. 10: Oxylog 1000 with test set-up: Testing the O concentration • Set the frequency selector (Freq.) to 10 1/min. • Set the ventilation pressure (Paw) to full right. • Set the Air Mix/No Air Mix selector switch to „Air Mix“.

  • Page 85: Deliver The Oxylog 1000 To The Operators Ready For Operation

    Deliver the Oxylog 1000 to the operators ready for operation. Tested Date:_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name:_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.

  • Page 86
    This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer. Page 18 5503.400 Oxylog 1000 Test List…
  • Page 87
    Test List 2 (Oxylog 1000 „new“) Oxylog 1000 File no.: Serial no.: 5503.400 ————————— Installation site: Edition: ————————— 02/2000 This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.
  • Page 88
    Hose clamp Sealing plug This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer. Page 20 5503.400 Oxylog 1000 Test List 2…
  • Page 89: Functional Tests

    Carry out the following functional tests using an O test pressure regulator or the central gas supply system. The supply pressure of Oxylog 1000 is 2.7 bar to 6 bar. This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.

  • Page 90: Checking The Supply Pressure And The Safety Valve

    • Unscrew compressed-gas connecting hose from connecting piece (compressed-gas supply). • To dismount the plug-in unit of Oxylog 1000, remove the screws and washers (1). • Remove the screws (2). • Pull out the connecting piece (ventilation hose and compressed-gas supply).

  • Page 91
    120 L/min «Pv» indicator ON/OFF switch Fig. 2: Oxylog 1000 with test set-up: Checking the supply pressure • Set the frequency adjuster (Freq.) to 10 1/min. • Set the minute volume adjuster (MV) to right limit stop. • Set the airway pressure (Paw) to right end stop.
  • Page 92
    Reading on the pressure gauge: Pplat = 80 mbar ( 10 mbar). (Peak pressure Pmax <100 mbar) • Place the ON/OFF switch of Oxylog 1000 to position «O». • Remove the hose clamp. • Disconnect the compressed-gas connector from the socket of the central gas supply system/compressed-gas cylinder.
  • Page 93: Checking The Vent Valve

    Test pressure regulator Flowmeter 0 to 10 L/min Fig. 4: Oxylog 1000 with test set-up: Checking the vent valve • Using the test pressure reducer, feed a flow of 10 L/min into the connecting piece of the ventilation hose. The increase in pressure should not exceed 8 mbar (hPa) (read pressure on pressure gauge of Oxylog 1000).

  • Page 94: Checking The I:e Ratio

    • Set the maximum airway pressure (Paw) to right end stop. • Set the Air Mix/ No Air Mix selector to the «No Air Mix» position. • Switch on Oxylog 1000 using the ON/OFF switch. • Use the stopwatch to measure the I:E ratio.

  • Page 95: Checking The Minute Volume

    Adjustable Flowmeter restrictor Fig. 5: Oxylog 1000 with test set-up: Checking the minute volume • Set the airway pressure adjuster (Paw) to right end stop. • Set the frequency adjuster (Freq.) to 10 1/min. • Set the minute volume adjuster (MV) to various steps, see the following table.

  • Page 96: Checking Warning Logics And Silence Control

    Test pressure regulator Ventilation valve Test lung Fig. 6: Oxylog 1000 with test set-up for supply pressure warning check • Set the minute volume adjuster (MV) to left limit stop. • Set the frequency adjuster (Freq.) to 10 1/min. • Set the airway pressure adjuster (Paw) to approx. 40 mbar (hPa).

  • Page 97
    (Paw low). The «Paw low» switching point is at approx. 10 mbar (hPa) (read the switching point on the pressure gauge of Oxylog 1000 and on the separate pressure gauge). This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.
  • Page 98
    (Paw high). The «Paw high» switching point is at 50 to 60 mbar (hPa) (read the switching point on the pressure gauge of Oxylog 1000 and on the separate pressure gauge).
  • Page 99
    Test pressure regulator Ventilation valve Test lung Fig. 9: Oxylog 1000 with test set-up: Checking the alarm sound silence • Set the minute volume adjuster (MV) to 10 L/min. • Set the frequency adjuster (Freq.) to 10 1/min. • Disconnect the test lung from the ventilation valve.
  • Page 100: Checking The O 2 Concentration

    Vol% Test adapter Test lung Ventilation valve Fig. 10: Oxylog 1000 with test set-up: Checking the O concentration • Set the frequency adjuster (Freq.) to 10 1/min. • Set the airway pressure (Paw) to right end stop. • Set the Air Mix/ No Air Mix selector to the «Air Mix» position.

  • Page 101: Supply Oxylog 1000 To User Ready For Operation

    Supply Oxylog 1000 to user ready for operation. Tested Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer.

  • Page 102
    This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and maintenance work carried out by the manufacturer. Page 34 5503.400 Oxylog 1000 Test List 2…
  • Page 104
    Manufacturer: Support: Dräger Medical b.v. Dräger Medical AG & Co. KG Kanaaldijk 29 Moislinger Allee 53 – 55 D5683 CR BEST D-23542 Lübeck The Netherlands Germany Tel: (++31) 499 331 — 331 Phone: (++49) (0) 1805-3723437 Fax: (++31) 499 331 — 335 Fax: (++49) 451/882 — 3779 Subject to change without notice Will not be replaced in the event of modifications.


  • Описание


  • Характеристики


  • Производитель


  • Доставка и оплата

Преимущества Draeger Oxylog 1000

Система для более безопасной рабочей среды.

Аппарат ИВЛ Oxylog 1000 служит для вентиляции при оказании первой помощи в экстренных ситуациях. Предназначенный для работы вне помещений, обладающий интуитивно понятным управлением, надежный и мобильный Oxylog 1000 является единственным в своем роде. Для обеспечения безопасности пациентов аппарат оборудован звуковой и визуальной сигнализацией, контролирующей давление в дыхательных путях и давление газоснабжения.

Все функции осуществляются с помощью пневматики, поэтому вентилятор не зависит от источников электроэнергии. Благодаря своей конструкции Oxylog 1000 обеспечивает большую надежность и готов к работе в сложных условиях.

Oxylog 1000 – аппарат ИВЛ для чрезвычайных ситуаций с переключением по времени, с управлением объемом и ограничением давления для принудительной вентиляции пациентов, которым требуется минутный объем, по крайней мере, в 3 л/мин.

Характеристики Draeger Oxylog 1000

Габаритные размеры (Ш x В x Г) 215 x 90 x 215 мм / 8,5 x 3,5 x 8,5 дюйм (без ручки)
Вес 3,15 кг (7,3 фунта)
Рабочий газ
Медицинский O2 или в исключительных случаях сжатый воздух
Давление подачи От 2,7 до 6 бар / от 40 до 88 фунт/кв. дюйм при 60 л/мин
Рабочие характеристики
Режим вентиляции IPPV/ CMV
PEEP-вентиляция с опционным PEEP-клапаном
Принцип действия Прерывание потока
Управление Переключение по времени, постоянный объем
Частота вентиляции, плавная регулировка от 4 до 54 л/мин
Минутный объем, плавная регулировка от 3 до 20 л/мин
Соотношение вдох : выдох (фиксированное) 01:01,5
Макс. давление в воздушных путях (Pmax), плавная регулировка от 25 до 60 мбар / см H2O
Концентрация O2 в вентиляционном газе при подаче O2
Переключение на „Air Mix“ (Воздушная смесь) около 60% по объему O2
Переключение на „No Air Mix“ (Без воздушной смеси) 100% по объему O2
Расход приводного газа прибл. 1,0 л/мин.
Объем мертвого пространства прибл. 12 мл
Комплайнс аппарата прибл. 1 мл/мбар / см H2O
Давление открытия предохранительного клапана 80 мбар / см H2O
Дисплей манометра от -10 до +80 мбар / см H2O
Функции сигнализации
Низкое давление подачи (Psupply) Давление подачи падает ниже 2,7 бар / 40 фунт/кв.дюйм
Давление в дыхательных путях высокое (Paw high) Фактическое значение превышает установленное значение (Pmax)
Давление в дыхательных путях низкое (Paw low) Давление 10 мбар/см H2O во время вдоха не превышается
Визуальные и звуковые сигналы тревоги. Обеспечиваются пневматическими компонентами и не требуют источника питания.
Условия эксплуатации
Диапазон температуры от -18 до +50 °C / от 0 до 122 °F
Относительная влажность отн. влажность от 15% до 95%
Давление окружающей среды от 700 до 1100 гПа
Выполнены вибрационные испытания в соотв. с MIL STD 810 F, метод 514.5
Использование в летательных аппаратах в соотв. с RTCA DO-160 D, раздел 8
Классификация согласно директиве ЕС 93/42/EEC Класс IIb
Код UMDNS 18-098
Типичное время работы MV = 10 л/мин
  • баллон – 2,5 л / 200 бар: прибл. 90 мин для „Air Mix“, прибл. 45 мин для „No Air Mix“
  • баллон с O2 типа E: прибл. 112 мин для „Air Mix“, прибл. 56 мин для „No Air Mix“
  • баллон с O2 типа D: прибл. 64 мин для „Air Mix“, прибл. 32 мин для „No Air Mix“

Компания Draeger специализируется на производстве медицинского оборудования и технологий для обеспечения безопасности. Продукция компании предназначена для защиты, поддержки и спасения человеческих жизней в больницах, пожарных и спасательных службах, правительственных учреждениях, горнодобывающей промышленности других отраслях по всему миру.

Компания основана в 1889 году в Любеке, Германия, и уже на протяжение пяти поколений является семейным предприятием, которое выросло до масштабов международной корпорации, акции которой котируются на фондовой бирже. Компания Draeger представлена в более чем 190 странах и насчитывает более 15 000 сотрудников по всему миру. В 2020 году чистый объем продаж компании составил около 3,4 млрд евро.

Oxylog 1000 — аварийный аппарат искусственной вентиляции легких с временным циклом, контролируемым объемом и ограниченным давлением. Он используется для контролируемой вентиляции пациентов в случаях, когда требуется минутный объем не менее 3 л/мин.

Oxylog используется для оказания неотложной медицинской помощи на протяжении 25 лет. Oxylog 1000 можно смело назвать самым маленьким и мобильным аппаратом ИВЛ в мире.

Благодаря своей фундаментальной конструкции Oxylog 1000 отличается высокой долговечностью и прочностью, готов к работе в тяжелых условиях.

Использование в любой рабочей среде

Oxylog 1000 предлагает первую медицинскую вентиляцию легких в форс-мажорных ситуациях. Оборудование может использоваться на открытом воздухе.

Надежность и транспортабельность делают Oxylog 1000 лучшим в своем классе. Его управление просто и понятно на интуитивном уровне. Вентилятор имеет встроенную сигнализацию с аудио- и видеоканалами. Она тщательно контролирует давление подачи и давление в дыхательных путях, чтобы гарантировать безопасную и успешную процедуру.

Все функции имеют пневматическое управление. Устройство полностью автономно и не зависит от подключения к электросети.

Благодаря своей фундаментальной конструкции Oxylog 1000 также отличается высокой надежностью и готов к работе в тяжелых условиях.

Oxylog 1000 во время вентиляции

Датчик давления контролирует и переключает стеноз и отключение пациента. Датчик давления состоит из двух диафрагм. Давление действует на нижнюю часть диафрагм. Давление подачи сосредоточено на кратерах на датчике давления во время фазы вдоха. Встроенные в датчик давления механические пружины оказывают на диафрагмы физические силы.

Сжатый газ подается из центральной системы подачи (ЦС) или из баллона с O2 с регулятором давления. Рабочее давление Oxylog 1000 составляет от 2,7 до 6 бар.

Частотный контроль

Когда установка выключена, клапан и выключатель закрыты. Вентиляция не проводится, на частотном регуляторе нет давления. Если Oxylog 1000 включен с помощью переключателя, газ проходит через переключатель и коллектор на регуляторы частоты. Газ проходит к клапану регулировки частоты и к управляющему патрубку.

В зависимости от того, как установлен клапан регулировки частоты, объем заполняется быстрее или медленнее, что определяет продолжительность вдоха. Как только объем заполнен, клапан переключается. Содержащийся в объеме газ выходит через клапан регулировки частоты. В зависимости от того, как он установлен, объем опорожняется быстрее или медленнее, что определяет продолжительность выдоха.

Клинические исследования, подтверждающие эффективность Dräger Oxylog 1000

Вентиляторы Dräger Oxylog 1000 оценивались с использованием смеси воздуха (60% кислорода) или 100% кислорода и тестировались на моделях, имитирующих нормальное легкое, легкое с низкой податливостью (острый респираторный дистресс-синдром) и легкое с высоким сопротивлением (астма).

Вентиляторы были испытаны в диапазоне смоделированных высот от уровня моря до 3048 м. В этом диапазоне дыхательный объем, выдаваемый Oxylog 1000, увеличился на 68%, а частота дыхания снизилась на 28%.

Изменения соответствовали аппарату ИВЛ независимо от содержания кислорода или модели легких. Важно, чтобы врачи, занимающиеся вентиляцией легких в отделениях интенсивной терапии в гипобарической среде, знали, что отдельные аппараты ИВЛ работают по-разному и чтобы они знали о характеристиках конкретного аппарата ИВЛ, который они используют.

Как купить аппарат ИВЛ Dräger Oxylog 1000?

Мы предлагаем выгодные цены на Dräger Oxylog 1000. Располагаем штатом профессиональных менеджеров, которые всегда готовы помочь с выбором оборудования или оформлением заказа. Если вы искали, где в Москве купить Dräger Oxylog 1000, свяжитесь с нами для оформления заказа.

Если вы сомневаетесь в выборе или не знаете, какие сопутствующие товары понадобятся, наши опытные консультанты помогут приобрести лучшее оборудование. Свяжитесь с нами для оформления заказа или получения подробной информации по телефону: 8-987-582-09-99.

Мы предлагаем стационарные и портативные модели с большими функциональными возможностями, а также:

  • доступные цены на современное оборудование;
  • поставки от известных производителей;
  • предложения по лизингу и аренде диагностической техники;
  • гарантии на любую модель.

← Вернуться в каталог


Артикул:
9601


Наличие:
Нет в наличии


Доставка:
Невозможна

Добавить к сравнению

Способы оплаты —
узнать больше

  • Наличный расчет
  • Безналичный расчет
  • В рассрочку или в кредит
  • Банковскими картами Visa / MasterCart / МИР

Запросите коммерческое предложение и технические характеристики для авторизации проекта

Быстрая и удобная эксплуатация с понятным интерфейсом

Конструкция аппарата отличается прочностью и удобными размерами, поэтому способна прослужить долго, может быть использована в любых условиях. Аппарат пневмоприводный (не зависит от источника питания). Это позволяет использовать его при любых погодных условиях и в любом транспорте/помещении. При предельном давлении/нарушении газоснабжения/соединений, срабатывает система сигналов тревоги, что обеспечивает безопасность в использовании.

1

Необходимые режимы и их функциональность

Переключатели Paw Low и Paw High контролируют давление на входе в аппарат, просты в использовании. Кроме того устройство оснащено режимом Pmax (вентиляция с ограничением по давлению с заданным временем вдоха), режим HLW (постоянная вентиляция при реанимации).

Особенности модели

  • Портативность и мобильность (21,5/9/21,5). Самый компактный аппарат из всей линейки Oxylog;
  • До 112 минут непрерывной работы;
  • Режимы IPPV/CMV;
  • Частота вентиляции плавно регулируется от 4 до 54 л/мин;
  • Частота вентиляции плавно регулируется от 4 до 54 л/мин;
  • Объем за минуту от 3 до 20 л/мин. Соотношение вдох/выдох 1:1,5.

Видео презентация

Тип ИВЛ Неинвазивный
Класс аппарата Профессиональный
Вес аппарата 3,15 кг без тележки
Дыхательный объем 100 – 2000 мл
Размеры аппарата 215 x 90 x 215 мм
Зарядка батареи 4-5 часов
Категории пациентов Взрослые, Дети
Аппарат ИВЛ Drager Oxylog 1000 1 шт
Тележка 1 шт
Инструкция по эксплуатации для пользователя 1 шт
Крепеж 1 шт
Гарантия 1 год
Производитель Dräger

Oxylog 1000 – аппарат ИВЛ для чрезвычайных ситуаций с переключением по времени, с управлением объемом и ограничением давления для принудительной вентиляции пациентов, которым требуется минутный объем, по крайней мере, в 3 л/мин.

Oxylog 1000 – вентилятор с самыми компактными размерами в линейке аппаратов ИВЛ от Draeger.

Аппарат ИВЛ Draeger Oxylog 1000. Особенности

Аппарат для экстренной вентиляции Oxylog 1000 отличается компактными размерами и предназначен для оказания экстренной помощи пострадавшим на месте аварий, несчастных случаев или вне помещений клиники.

Аппарат работает на пневматике. Ему не требуется подключение к источникам энергопитания, поэтому нет необходимости заряжать аккумуляторы или использовать провода.

Интуитивно понятное управление ускоряет оказание медицинской помощи, что важно в критических ситуациях.

Контроль давления в дыхательных путях и давление газоснабжения позволяет отслеживать безопасность ИВЛ. Звуковая и световая сигнализация привлекает внимание медицинских работников к нарушению параметров вентиляции.

Прочный корпус выдерживает даже самые суровые условия эксплуатации. При необходимости можно дополнить аппарат ИВЛ Oxylog 1000 сумкой-переноской и другими аксессуарами.

Технические характеристики:

  • Инвазивная вентиляция.
  • Давление подачи от 2,7 до 6 бар.
  • Режимы вентиляции — IPPV/ CMV.
  • Частота вентиляции плавно регулируется в диапазоне от 4 до 54 л/мин.
  • Минутный объем регулируется в диапазоне от 3 до 20 л/мин.
  • Фиксированное соотношение вдох/выдох — 1:1,5.

Страна изготовления

Германия

Класс аппарата

Условный класс, определяемый суммой характеристик, функций и цены.

Базовый

Тип аппарата

Различают стационарные и портативные аппараты. Принадлежность аппарата к определенному типу зависит от его веса и габаритов.

Портативный аппарат ИВЛ может весить не более 1,5 кг. Он применяется при оказании помощи на дому или в экстремальных условиях. Стационарные аппараты используются в медицинских учреждениях.

Транспортные ИВЛ, как правило, используются в условиях внебольничной транспортировки, могут иметь трансопортный и летный сертификат, портативные с длительной автономностью, устойчивы к тряске, падениям и сниженному давлению в воздухе.

  • Транспортный

Тип вентиляции (ИВЛ)

При применении неинвазивного способа на лицо больного накладывается плотно прилегающая маска, через которую поступает воздух в тело пациента.

Инвазивный способ подразумевает внедрение трубки в трахею через рот или нос (интубация).

  • Инвазивный
  • Неинвазивный (опция)

Управление

Сенсорное — если большинство кнопок являются сенсорными на дисплее.

Кнопки — это жесткие кнопки на корпусе рядом с дисплеем.

  • Кнопки/поворотный регулятор

Тип привода

Способ подведения энергии к аппарату ИВЛ для осуществления автоматического вдоха.

Ручной. Вентиляция производится по средствам сжатия мешка или меха рукой. Преимущества способа: простота устройства, минимальные потери мощности и получение оператором ощущения непосредственного контакта с легкими пациента.

Пневматический — привод его от заранее сжатого газа. Преимуществом аппарата с пневмоприводом является возможность выполнения его автономным, т.е. независящим от внешнего источника энергии, что имеет первостепенное значение для экстренной помощи пациенту в труднодоступных районах. Однако системы газоснабжения являются сложным инженерным сооружением, требующим непрерывного обслуживания, соблюдения строгих правил техники безопасности, периодических освидетельствований.

Электропривод (турбинный/поршневой) — питание аппарата от электросети. Для подачи воздуха служит турбина/поршень. Преимущества: стабильность электросети гораздо выше, чем пневмосети. В качестве минусов приводятся сложности при ремонте аппарата, высокий уровень призводимого им шума, необходимость соблюдения правил техники безопасности при работе с электричеством.

Комбинированный/гибридный привод используется в аппаратах для длительной ИВЛ. В них работа управляющих цепей обеспечивается электропитанием, а в качестве генератора вдоха используется непосредственное поступление газов от внешней пневмосети. Питание от двух источников позволяет исключить из состава аппарата генератор вдоха, что значительно упрощает конструкцию, снижает размеры и стоимость аппарата, увеличивает надежность его работы. Одновременно значительно снижается создаваемый аппаратом шум. Часто электропривд в таких машинах является резервным, а также может служить для транспортировки.

  • Пневмопривод

Тип пациентов

Возможность работы с новорожденными пациентами весом от 500 грамм до 3 кг.

  • Взрослые
  • Дети

Дыхательный объем, мл

Величина одного автоматического вдоха в режиме объемной вентиляции (VCV/SIMV)

100-2000

Размеры (ШxВxГ), мм

215х90х215

Найти похожие

Понравилась статья? Поделить с друзьями:
  • Должностная инструкция зам декана по учебной работе
  • Starline s96 v2 инструкция по настройке приложения
  • Инструкция по использованию клеевого пистолета для рукоделия
  • Huawei watch gt 2 pro руководство пользователя
  • Руководство 351904 04