Easylyte plus na k cl анализатор электролитов инструкция

  1. Manuals
  2. Brands
  3. Medica Manuals
  4. Measuring Instruments
  5. EasyLyte Na/K
  6. Service manual
  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Quick Links

Cat. No. 2285 R3

EasyLyte

Service Manual

Na/K

Na/K/Cl

Na/K/Li

Analyzers, Serial Numbers 5000 and above

EasySampler

EDIC@

loading

Summary of Contents for Medica EasyLyte Na/K

  • Page 1
    Cat. No. 2285 R3 EasyLyte Service Manual Na/K Na/K/Cl Na/K/Li Analyzers, Serial Numbers 5000 and above EasySampler EDIC@…
  • Page 2
    Information contained herein is believed by Medica to be accurate; in any event, no responsibility, whether express or implied, is assumed hereby by Medica for or in connection with the use thereof, or for infringement of any third party rights which might arise therefrom, or from any represen-…
  • Page 3: Table Of Contents

    Service Manual Contents Product Description Product Overview ……………………..1 System Overview ……………………..1 Table of Symbols ……………………..2 Front Assembly Components ………………….3 Front Assembly Components – Analyzer Series below 6100 …………7 Electrode Housing Components ………………..11 Rear Assembly Components ………………..12 Replacement List ……………………13 Troubleshooting Troubleshooting Overview ………………….15 Troubleshooting Guide ……………………16 AC Power Verification ……………………17…

  • Page 4
    EasySampler Product Description ……………………79 Assembly and Component Overview………………80 Replacement List ……………………81 Troubleshooting Guide ……………………82 Alignment Verification……………………83 Tray Placement Verification ………………….84 Carousel Test ……………………..85 Tray Cycle Verification ……………………86 Sensor Verification ……………………87 QC Tests……………………….88 Repair Overview……………………..90 Recommended Tools ……………………90 Housing Removal ……………………91 Housing Installation ……………………93 Spindle Column Removal………………….95 Spindle Column Installation ………………….97…
  • Page 5: Product Description

    Replacement List in this section which lists replaceable assemblies, components, and appropriate catalog numbers. If repair to any electrical assembly is necessary, consult Schematics in the APPENDIX SECTION for further detail. Service personnel are encouraged to contact Medica directly with any problems or concerns relating to electrical assemblies.

  • Page 6: Table Of Symbols

    Product Description Table of Symbols NOTES Helpful information or references to other related sections of the manual. KEY INFORMATION Could take the EasyLyte analyzer out of operation. WARNING Possible biological hazard to the operator. 6100 EASYLYTE ANALYZER’S Information for EasyLyte analyzers with serial num- bers BELOW 6100 below 6100.

  • Page 7: Front Assembly Components

    Product Description Front Assembly Components This section describes the main assemblies and components of the EasyLyte front assembly. Refer to Figures P1, P2, and P3 for the location of each component. Front Housing front case of the analyzer Power Supply Assembly powers the analyzer Entry Module connection for the power cable Fuse Bay holds 2 fuses Power Supply Cable connects the power supply to the CPU assembly…

  • Page 8
    Product Description Printer Assembly Front Housing Display Assembly Electrode Jack(s) Sample Detector Cable Display Cable Probe Position Scale Assembly Firmware (EPROM) Probe Position Sensor Pump Motor Interface Ports Cable AUTOSAMPLER Sensor Cable Power Supply Cable Probe Motor Cable Printer Cable Voltage Selector Switch Entry Module…
  • Page 9
    Product Description Front Housing Printer Assembly Probe Arm Guide Rod Display Cable Probe Position Scale Probe Motor Assembly Lead Screw Probe Motor Mounting Bracket Figure P2 Front Assembly — Right Side View…
  • Page 10
    Product Description Front Housing LCD Screen Display Cable Pump Head Assembly Pump Assembly Printer Cable Pump Motor Figure P3 Front Assembly — Left Side View…
  • Page 11: Front Assembly Components — Analyzer Series Below 6100

    Product Description 6100 Front Assembly Components – Analyzers Series Below 6100 This section describes the main assemblies and components of the EasyLyte front assembly, analyzer series below 6100. Refer to Figures P4, P5, P6 for the location of each component. Front Housing front case of the analyzer Power Supply Assembly powers the analyzer Entry Module connection for the power cable…

  • Page 12
    Product Description Printer Assembly Front Housing Sample Detector Cable Electrode Jack(s) Display Cable Display Assembly Probe Position Scale Assembly Firmware (EPROM) Probe Position Sensor Pump Motor Cable AUTOSAMPLER Interface Ports Sensor Cable Power Supply Cable Probe Motor Cable Printer Cable Entry Module Voltage Selector Switch…
  • Page 13
    Product Description Front Housing Printer Assembly Guide Rod Probe Position Scale Probe Motor Assembly Lead Screw Probe Motor Mounting Bracket Voltage Selector Switch Figure P5 Front Assembly Right Side View 6100 — Analyzers with Serial Numbers below 6100…
  • Page 14
    Product Description Front Housing Display LCD Screen Cable Pump Head Assembly Pump Assembly Printer Cable Pump Motor Figure P6 Front Assembly Left Side View 6100 – Analyzers with Serial Numbers below 6100…
  • Page 15: Electrode Housing Components

    Product Description Electrode Housing Assembly Components This section describes the components of the EasyLyte electrode assembly. Refer to Figure P7 for the location of each component. Electrodes sodium, potassium, chloride, lithium, reference Sample Detector detects presence of air/liquid, sets sample position inside electrodes Internal Filling Solution 2 molar KCI solution, acts as ‘salt bridge’…

  • Page 16: Rear Assembly Components

    Product Description Rear Assembly Components This section describes the components of the EasyLyte Rear Assembly. Refer to Figure P8 for the location of each component. Rear Housing rear case Label identifies serial number Rear Housing Label with Serial Number Autosampler Serial Port Port…

  • Page 17: Replacement List

    Product Description Replacement List Consumable and assemblies items can be purchased directly from Medica. CONSUMABLE ITEMS Description Catalog No. Na Electrode 2102 K Electrode 2101 Cl Electrode 2113 Li Electrode 2106 Reference Electrode 2103 Sample Detector 2257 Sample Probe 2107…

  • Page 18
    EasyLyte Na/K, 1-line display 001937-007 EasyLyte Na/K, 2-line display 001937-008 EasyLyte Na/K, Chinese 001957-001 EasyLyte Na/K/Cl or Li, Chinese 001957-002 EasyLyte Na/K, Japanese 001957-003 EasyLyte Na/K/Cl or Li, Japanese 001957-004 The CPU Part Numbers include the EasyLyte CPU Board with software installed, programmed to operate specifically in EasyLytes described as above.
  • Page 19: Troubleshooting

    This section describes troubleshooting the mechanical and electrical functions of the EasyLyte analyzer. This section is based on the principle of replacing suspected faulty assemblies with new assemblies. Medica recommends keeping an inventory of functioning test assemblies for troubleshooting the EasyLyte. Refer to the Troubleshooting Guide in this section to deter- mine the cause and corrective action that must be taken to properly service the EasyLyte Analyzer.

  • Page 20: Troubleshooting Guide

    Troubleshooting Troubleshooting Guide Refer to the chart below to help determine the problem and corrective action needed. Problem Description Problem Source Corrective Action Steps Blank Display Power Malfunction or • AC Power Verification Defective Display • DC Power Verification • Display Verification No response Power Malfunction or •…

  • Page 21: Ac Power Verification

    Troubleshooting AC Power Verification WARNING: Careless handling of AC Power, especially 220V, can be hazardous or fatal. Only qualified personnel should work with AC Power. 1. Check that the voltage selector switch is set at the proper voltage (110V or 220V). This volt- age must match the power outlet voltage.

  • Page 22
    Troubleshooting 5. Remove the rear housing. Refer to Rear Housing Removal in the REPAIR SECTION. 6. Check for short circuits between the test points listed below: TP12 and TP2 TP12 and TP1 TP12 and TP3 TP12 and TP10 TP12 and TP11 If a short exists, replace the CPU board.
  • Page 23
    Troubleshooting 12. Check for short circuits between the test points below: TP11 and TP5 6100 TP11 and TP8 TP9 and TP6 TP9 and TP7 If a short exists, replace the CPU Board and return to normal 13. Disconnect the power supply cable (13-pin connector) from the CPU board, see Figure T4. Power Supply Pin 13 –>…
  • Page 24: Dc Power Verification

    Troubleshooting DC Power Verification 1. Remove the rear housing. See Rear Housing Removal in the REPAIR SECTION for details. 2. Connect the power cord first to an outlet, then to the entry module. 3. Connect the reference side of the voltage meter to TP12 on the CPU board. 4.

  • Page 25: Display Verification

    Troubleshooting Display Verification 1. Disconnect the display cable from the CPU board. See Display Assembly Removal in the REPAIR SECTION for details. 2. Connect a known good display assembly to the CPU board and connect the main power cord. Verify that the test display is operative. Based on the test results, take the appropriate corrective action.

  • Page 26: Probe Motor Verification

    Troubleshooting Probe Motor Verification 1. Disconnect the power cord, wait 5 seconds, then reinstall the power. The probe arm should move down and back up. If there are no error messages, resume normal operation. If an “MTS” error occurs, proceed to STEP 2. 2.

  • Page 27: Pump Motor Verification

    Troubleshooting Pump Motor Verification 1. Enter the QC TEST? function and proceed to the PUMP TEST? Refer to QC Tests in this section for further instructions. 2. At the PUMP TEST? prompt, press YES. 3. The pump should rotate ten revolutions. 4.

  • Page 28: Printer Verification

    Troubleshooting Printer Verification 1. Press the printer paper advance button. If the printer head moves proceed to STEP 2. If the printer head does not move, proceed to STEP 3. 2. Verify that the printer is on. Turn the printer on by pressing NO to the PRINTER OFF? ques- tion in the USER OPTIONS section.

  • Page 29: Sample Detector Verification

    Troubleshooting Sample Detector Verification 1. Remove the electrode housing from the analyzer. See the EasyLyte Operators Manual for fur- ther details. 2. Remove the sample detector from the front housing. 3. Install the sample detector into the sample detector receptacle. Do not install the electrode housing or pump tubing at this time.

  • Page 30: Signal Input Verification

    Troubleshooting Signal Input Verification 1. Remove and inspect electrodes and electrode housing for liquid leakage and salt build-up. All electrodes and the electrode housing surfaces MUST be completely dry. See EasyLyte Operator’s Manual for further information on consumable items. 2. Inspect the electrode pin receptacles on the EasyLyte analyzer front housing. Remove any liq- uid or salt from these areas.

  • Page 31
    Troubleshooting 6. Disconnect the black wire from the back connector on the Cl-/Li+ red wire. 7. Turn on the Voltage Source. 8. Set the Voltage Source to the group 1 settings listed in the table below. Three new individual values will appear on the screen. Wait a few seconds for the values to stabilize then verify that the values are within specification.
  • Page 32: Serial Communication Verification

    Troubleshooting Serial Communication Verification 1. Connect the EasyLyte analyzer to a device with a known working serial port. Use a known good EasyLyte serial cable, part no. 05001334. Make cross connections as required (e.g. 2 to 3; 4 to 5). 2.

  • Page 33: Qc Tests

    Troubleshooting QC Tests 1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis- play, as shown in the chart below. LCD Screen Displays Press CALIBRATE NOW? DAILY CLEANER? STANDBY MODE? SECOND MENU? * SEE RESULTS? * * PACK USAGE? * * QUAL CONTROLS? * * OPER FUNCTS? *…

  • Page 34
    CHECK CAL CONSTS? Displays the factory set electrode constants. The acceptable ranges are: Left column number: 0.92 to 1.08 Right column number: -800 to +800 If values are out of specification contact Medica. RAM/PROM TEST? Displays current software version and date installed. RAM CHECK OUT? Clears the EasyLyte memory.
  • Page 35
    Troubleshooting QC Test Function QC Test Function DIAGNOSTICS? Performs diagnostic testing of the EasyLyte Analyzer (See EasyLyte Operator’s Manual). LAST CAL VALUES? Checks last calibration values. Acceptable ranges are Na: 50-64, K: 50-64, Cl: 40-64, Li: 50-64. ELECTR’D VALS? Direct electrode millivolt readings. Typical values are Na: 35-125, K: 35-125 Cl: 35-125, Li: 30-60.
  • Page 36: Repair

    If repair to any electrical assembly is necessary, consult the Schematics in the APPENDIX SEC- TION for further detail. Service personnel are encouraged to contact Medica directly with any problems or concerns relating to electrical assemblies.

  • Page 37: Tools

    Repair Recommended Tools The following is a list of the tools required to perform repairs on the EasyLyte analyzer. Please refer to this list to determine the appropriate tool and size needed. Phillips Screw Driver, #1 Phillips Screw Driver, #2 Allen Wrench, 1/16 inch Scissors Hex Nut Driver, 11/32 inch, hollow stemmed…

  • Page 38: Rear Housing Removal

    Repair Rear Housing Removal 1. Disconnect the power cord from the entry module, see Figure R1. 2. Detach the four screws from the rear housing using a #2 Phillips screw driver, as shown in Figure R2. Power Cord Entry Module Figure R1 Screws Screws…

  • Page 39
    Repair 3. Separate the front and back housing. See Figure R3. Front Housing Rear Housing Figure R3…
  • Page 40: Rear Housing Installation

    Repair Rear Housing Installation 1. Align and attach the front and back housing. See Figure R4. Verify that all cable connectors are attached properly. Refer to Figures P1 and P6 in the PRODUCT DESCRIPTION SECTION for detail. Verify that all wires are contained within the front housing before proceeding.

  • Page 41
    Repair 2. Attach the four screws to the rear housing using a #2 Phillips screw driver, as shown in Figure R5. 3. Connect the power cord to the entry module, see Figure R6. Screws Screws Figure R5 Entry Module Power Cord Figure R6…
  • Page 42: Power Supply Removal

    Repair Power Supply Removal 1. Disconnect the power supply cable from the CPU board, see Figure R7. Refer to Figure R8 for EasyLyte analyzers with serial numbers below 6100. 2. Detach the four screws from the bottom of the front housing, using a #2 Phillips screw driver. See Figure R9.

  • Page 43
    Repair 3. Lift and remove the power supply assembly from under the CPU board, as shown in Figure R10. Refer to Figure R11 for EasyLyte analyzers with serial numbers below 6100. Power Supply Assembly Figure R10 Power Supply Assembly 6100 Figure R11…
  • Page 44: Power Supply Installation

    Repair Power Supply Installation 1. Slide the power supply assembly under the CPU board, as shown in Figure R12. Fit into the alignment holes in the bottom of the front housing. Refer to Figure R13 for EasyLyte analyzers with serial numbers below 6100. 2.

  • Page 45
    Repair 3. Connect the power supply cable to the CPU board, as shown in Figure R15. Refer to Figure R16 for EasyLyte analyzers with serial numbers below 6100. Connector to CPU Board Power Supply Cable Figure R15 Connector to CPU Board Power Supply Cable 6100…
  • Page 46: Printer Removal

    Repair Printer Removal 1. Remove the printer lid from the printer, see Figure R17. 2. Remove the paper roll from the printer bay, as shown in Figure R18. Printer Lid Printer Figure R17 Paper Roll Printer Bay Figure R18…

  • Page 47
    Repair 3. Disconnect the printer cable from the CPU board, see Figure R19. Refer to Figure R20 for EasyLyte analyzers with serial numbers below 6100. Connector to CPU Board Printer Cable Figure R19 Connector to CPU Board Printer Cable 6100 Figure R20…
  • Page 48
    Repair 4. Detach the printer cable from the retaining clamps located along the right side of the CPU board, as shown in Figure R21. 5. Detach the two screws using a # 2 Phillips screw driver. See Figure R22. Retaining Clamps Printer Cable Figure R21…
  • Page 49
    Repair 6. Remove the printer body as shown in Figure R23. 7. Pull the printer cable up from behind the CPU board, while holding the printer body as shown in Figure R24. Printer Body Figure R23 Printer Body Printer Cable CPU Board Figure R24…
  • Page 50: Printer Installation

    Repair Printer Installation 1. Slide the printer cable behind the CPU board, as shown in Figure R25. 2. Install the printer body into place. See Figure R26. 3. Attach the two screws using a #2 Phillips screw driver. See Figure R27. Printer Body Printer Cable CPU Board…

  • Page 51
    Repair 4. Connect the printer cable to the CPU board, see Figure R28. Refer to Figure R29 for EasyLyte analyzers with serial numbers below 6100. 5. Using scissors, cut the beginning of the paper roll into a “V” shape. See Figure R30. Insert the paper into the feeder slot inside the printer bay, and press the paper feed button.
  • Page 52
    Repair 7. Place the printer lid onto the printer, see Figure R32. 8. Attach the printer cable to the retaining clamps located along the right side of the CPU board, as shown in Figure R33. Printer Lid Printer Figure R32 Printer Cable Retaining Clamps Figure R33…
  • Page 53: Cpu Board Removal

    Repair CPU Board Removal 1. Disconnect the probe motor, sensor, pump motor, power supply and printer cables from the CPU board. See Figure R34. Refer to Figure R35 for EasyLyte analyzers with serial num- bers below 6100. 2. Remove the display cable from the top of the CPU board as shown in Figure R36. Probe Motor Cable Printer Cable Sensor Cable…

  • Page 54
    Repair 3. For EasyLyte Analyzers with serial numbers below 6100, push the display cable tabs to the side to unlock. Remove the cable from the connector. See Figure R37. 4. Disconnect the electrode jacks from behind the CPU board. See Figure R38 for the location of the electrode jacks.
  • Page 55
    Repair 5. Disengage the CPU board from the four plastic mounting posts, located at each corner of the CPU board. To disengage, press the mounting post tab in and pull the CPU board over the tab, as shown in Figure R39. 6.
  • Page 56
    Repair 7. Slide the CPU board off the mounting posts. See Figure R41. Mounting Mounting Posts Posts Figure R41…
  • Page 57: Cpu Board Installation

    Repair CPU Board Installation 1. Align the CPU board with the plastic mounting posts. Snap the right side mounting posts of the CPU board into place. See Figure R42. Do not snap the left or center mounting posts into place at this time. Mounting Posts Figure R42…

  • Page 58
    Repair 2. Connect the electrode jacks to the back of the CPU board. To connect, gently push the elec- trode jack straight into the connector pin. See Figure R43 to determine the proper connector pin position on the CPU board and the recommended installation order. See Figure R44 for EasyLyte Analyzer with serial numbers below 6100.
  • Page 59
    Repair 4. Attach the display cable to the CPU board. Push the display cable tabs to the front to secure cable in place. See Figure R46. 5. Connect the probe motor, sensor, pump motor, power supply and printer cables to the CPU board, see Figure R47.
  • Page 60: Display Removal

    Repair Display Removal 1. Detach the two short self tapping screws from the sample detector board using a #2 Phillips screw driver. See Figure R49. 2. Move and hold the display cable to expose the three display assembly screws. Detach the three short screws from the display assembly using a #2 Phillips screw driver.

  • Page 61
    Repair 3. Carefully remove the display assembly and the LCD screen, holding both items together. See Figure R51. Notice the orientation of the LCD screen which is extremely important when installing the display assembly. The one-line display has soldered leads located on the top of the LCD screen;…
  • Page 62: Display Installation

    Repair Display Installation 1. Place the display adapter into the front housing window. The smooth edge should be facing the front housing of the analyzer. For two-line displays there is an additional orientation notch. To install correctly, arrange the adapter so that the notch is located on the top inside edge of the adapter. 2.

  • Page 63
    Repair 4. Position the front housing so that the front of the analyzer is facing down. Align the display assembly and insert into place, as shown in Figure R53, using the YES/NO buttons as a guide. 5. Hold the display in place and attach the three screws using a #2 Phillips screw driver. See Figure R54.
  • Page 64
    Repair 6. Align the sample detector board so the sample detector cable is positioned on the right side. Attach the two screws lightly, using a #2 Phillips screw driver, see Figure R55. 7. Install the sample detector in the front of the analyzer to verify the sample detector board alignment.
  • Page 65: Pump Motor Removal

    Repair Pump Motor Removal 1. Remove the pump tubing from the front of the analyzer. 2. Using a wire cutter, cut the tie wrap that secures the pump motor cable wires, as shown in Figure R57. 3. Detach the four screws from the front of the analyzer, using a #1 Phillips screw driver. See Figure R58.

  • Page 66: Pump Motor Installation

    Repair Pump Motor Installation 1. Install the pump motor assembly through the front housing, as shown in Figure R60. The pump motor wires are positioned below the pump when properly aligned. 2. Attach the four screws into the front of the analyzer, using a #2 Phillips screw driver. See Figure R61.

  • Page 67: Probe Motor Assembly Removal

    Repair Probe Motor Assembly Removal 1. Detach the screw at the top of the probe position scale, using a #2 Phillips screw driver, as shown in Figure R62. Remove the washer. Some EasyLyte analyzers use a white spacer that should be removed also. 2.

  • Page 68
    Repair 3. Remove the C-shaped retaining ring from the actuator, using the retaining ring pliers. DO NOT stretch out the retaining ring with the pliers. Disengage the actuator from the top of the lead screw assembly. See Figure R64. Save the retaining ring. 4.
  • Page 69
    Repair 5. Detach the three screws from the probe motor bracket, using a #2 Phillips screw driver, as shown in Figure R67. Remove the washers. The other two screws are located on the bottom edge of the bracket. 6. Remove the Probe Motor Assembly, as shown in Figure R68. Probe Motor Bracket Figure R67…
  • Page 70: Probe Motor Assembly Installation

    Repair Probe Motor Assembly Installation 1. Place the probe motor assembly into place. See Figure R69. 2. Attach the three washers and screws to the probe motor assembly bracket, using a #2 Phillips screw driver. See Figure R70. The other two screws are located on the bottom edge of the bracket. Probe Motor Assembly Figure R69…

  • Page 71
    Repair 3. Insert the actuator through the top of the lead screw assembly as shown in Figure R71. Install the retaining ring onto the actuator using the retaining ring pliers. DO NOT stretch out the retaining ring with the pliers. 4.
  • Page 72
    Repair 6. Install the sample probe as follows. Refer to Figure R73 for details. A. With the probe hole facing out, align the probe collar with the white line on the top of the probe arm. When aligned, snap the probe into the probe arm. B.
  • Page 73
    Repair 7. Hold the sample probe against the alignment valve tool to keep in place. Install the probe position scale through the top of the housing, see Figure R74. Verify the probe position alignment slot is located on the bottom right side of the scale. Probe Position Scale Probe Position…
  • Page 74
    Repair Place the scale into position. The scale should be seated inside the probe position sensor and inside the mounting bracket slot. See Figure R75. Attach and slightly tighten the washer and screw at the top of the position scale. Do not fully tighten the screw.
  • Page 75
    Repair 11. When alignment is complete, fully tighten the probe position screw into place using a #2 Phillips screw driver, as shown in Figure R76. Screw Figure R76…
  • Page 76: Guide Rod Removal

    Repair Guide Rod Removal 1. Remove the sample probe from the probe arm and the solutions valve. See Figure R77. Remove the solutions valve. 2. Detach the screw from the bottom of the guide rod, using a #2 Phillips screw driver, as shown in Figure R78.

  • Page 77
    Repair 3. Remove the guide rod and probe arm assembly, down and away from the housing. See Figure R79. Guide Rod Probe Arm Figure R79…
  • Page 78: Guide Rod Installation

    Repair Guide Rod Installation 1. Install the guide rod as shown in Figure R80. To install, first insert the guide rod into the alignment hole located in the top wall of the front assembly. 2. Attach the washer and screw to the bottom of the guide rod, using a #2 Phillips screw driver.

  • Page 79: Electrode Jack Removal

    Repair Electrode Jack Removal 1. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R82. 2. Remove the hex nuts using the nut driver. See Figure R83. 3. Pull the electrode jacks from the front housing of the analyzer, as shown in Figure R84. Electrode Jack Wires Nut Driver…

  • Page 80: Electrode Jack Installation

    Repair Electrode Jack Installation 1. Push the electrode jacks through the holes in the front of the analyzer, see Figure R85. 2. Insert the electrode jack wires into the 11/32 inch nut driver. See Figure R86. 3. Attach the hex nut using the nut driver, and tighten, as shown in Figure R87. Analyzer Front Housing Electrode Jack…

  • Page 81
    EasySampler Product Description The EasySampler when used with the EasyLyte analyzer provides automated batch sampling of serum, plasma or diluted urine for electrolyte analysis. The EasySampler can also perform emer- gency (STAT) analysis of blood samples on demand. The EasySampler contains a 25 position sample tray which includes a STAT and two quality con- trol positions.
  • Page 82: Assembly And Component Overview

    EasySampler Assembly and Component Overview The EasySampler is made up of the following mechanical assemblies and associated compo- nents. Refer to Figure A1 to determine the location of each assembly or component in relation to the EasySampler. EasySampler Housing case of the EasySampler IR Windows covers the IR beam transmitter and receiver Chassis Assembly frame which supports the mechanical assemblies PC Board Assembly contains the PC board and IR transmitter and receiver…

  • Page 83
    EasySampler EasySampler Replacement List The following assemblies and components can be purchased directly from Medica. Description Part No. EasySampler Housing 001074 IR Windows 001192 Chassis Assembly 001370 PC Board Assembly 001233 Motor Assembly 001328 Spindle Column Assembly 001680 Sensor Assembly…
  • Page 84: Tray Placement Verification

    EasySampler EasySampler Troubleshooting Guide Refer to the chart below to determine the problem and corrective action needed. Problem Description Problem Source Corrective Action Steps LCD screen displays SAFETY BEAM • Verify there is no IR obstruction “OBSTRUCTION” and OBSTRUCTION • Sensor Verification analyzer is beeping LCD screen displays “MTS”…

  • Page 85: Alignment Verification

    EasySampler Alignment Verification 1. Remove and reinstall the power cord. The sample tray will rotate and then stop. 2. ALIGN PROBE/CUP? appears on the LCD screen. Press YES to verify the probe alignment. The sample probe will descend and stop above the “Daily Cleaner” cup position. 3.

  • Page 86
    EasySampler Tray Placement Verification 1. Advance to the DAILY CLEANER? display; press YES. TRAY IN PLACE? appears on the LCD screen; press YES. 2. The sample tray should rotate and stop with the probe at cup position “12”. 3. CLEANER IN TRAY? appears on the LCD screen; press NO. 4.
  • Page 87: Carousel Test

    EasySampler Carousel Test 1. Enter the QC TEST? menu and proceed to TEST CAROUSEL?. Refer to QC Test in this section for further instructions. 2. At the TEST CAROUSEL? display, press YES. 3. CALIB CAROUSEL? appears on the LCD screen; press YES. The sample tray should spin one revolution;…

  • Page 88: Tray Cycle Verification

    EasySampler Tray Cycle Verification 1. Enter the QC TEST? menu and proceed to the TEST CAROUSEL? test. Refer to QC Test in this section for further instructions. 2. At the TEST CAROUSEL? display, press YES. At the CALIB CAROUSEL? display, press NO.

  • Page 89: Sensor Verification

    EasySampler Sensor Verification 1. Remove the rear housing of the EasyLyte analyzer as detailed in Rear Housing Assembly Removal in the REPAIR SECTION. 2. Connect a voltage meter to the EasyLyte analyzer. Attach one reference lead to TP11 (DGND) and the other to U13 PIN #2 on the analyzer CPU board. 3.

  • Page 90
    EasySampler QC Tests 1. To access the QC TEST? functions, answer the questions that appear on the EasyLyte dis- play, as shown on the chart below. LCD Screen Displays Press CALIBRATE NOW? DAILY CLEANER? STANDBY MODE? SECOND MENU? * SEE RESULTS? * * PACK USAGE? * * QUAL CONTROLS? * * OPER FUNCTS? *…
  • Page 91
    EasySampler 7. Consult the table below which lists all the QC tests and their functions. To select a test press YES. To bypass a test press NO. QC Test QC Test Function QC TEST? Access to the QC Test Functions. TURN ON MV’S? START CLOCK? Analyzer QC Test Functions.
  • Page 92
    EasySampler Repair Overview Always consult Troubleshooting in this section first to determine the problem before attempting to service the EasySampler. Please note that complete mechanical assemblies can be replaced. Refer to the Replacement List in this section which lists replaceable assemblies and appropriate catalog numbers. Recommended Tools The following is a list of the tools required to perform repairs on the EasySampler.
  • Page 93: Housing Removal

    EasySampler Housing Removal 1. Turn the EasySampler upside down and gently place on a flat surface. 2. Remove the two screws from the cable retaining plate using a #2 Phillips screw driver, as shown in Figure A2. Cable Retaining Screws Plate Figure A2…

  • Page 94
    EasySampler 3. Remove the EasySampler cable from the groove in the bottom of the housing. 4. Remove the three screws from the metal base using a #2 Phillips screw driver, as shown in Figure A3. 5. Turn the EasySampler to the upright position. 6.
  • Page 95: Housing Installation

    EasySampler Housing Installation 1. Align and place the housing over the chassis assembly, as shown in Figure A5. Be careful not to damage the IR receiver or transmitter on the PCB. 2. Turn the EasySampler upside down and gently place on a flat surface. 3.

  • Page 96
    EasySampler 4. Thread the EasySampler cable into the groove in the bottom of the housing. See Figure A7. 5. Attach the two screws to the cable retaining plate using a #2 Phillips screw driver, as shown in Figure A7. Cable Retaining Plate EasySampler Cable…
  • Page 97: Spindle Column Removal

    EasySampler Spindle Column Removal 1. Disconnect the sensor cable and the motor cables from the PC Board. See Figure A8. 2. Turn the adjustment knob counterclockwise to loosen. Remove the adjustment knob from the bottom of the chassis assembly, as shown in Figure A9. Some EasySamplers use a spacing disk that should be removed and retained.

  • Page 98
    EasySampler 3. Remove the spindle column assembly, as shown in Figure A10. Spindle Column Assembly Figure A10…
  • Page 99: Spindle Column Installation

    EasySampler Spindle Column Installation 1. Place the spindle column assembly down into place, as shown in Figure A11. 2. Attach the adjustment knob, see Figure A12, and turn clockwise to tighten. Some EasySamplers use a spacing disk that should be attached before the knob is installed.

  • Page 100
    EasySampler 3. Attach the sensor cable and the motor cable to positions J1 and J3 on the PC board, see Figure A13. Sensor Connector (J3) Sensor Motor Connector (J1) Cable Motor Cable Figure A13…
  • Page 101: Sensor Removal

    EasySampler Sensor Removal 1. Loosen the two slotted screws from the motor bracket, using a #2 Phillips screw driver, as shown in Figure A14. 2. Remove the sensor and mounting block, as shown in Figure A15. 3. Remove the two slotted screws from the mounting block, using a #2 Phillips screw driver. 4.

  • Page 102: Sensor Installation

    EasySampler Sensor Installation 1. Place the sensor into the mounting block. 2. Attach the two slotted screws to the mounting block, using a #2 Phillips screw driver. 3. Place the sensor and mounting block into the motor bracket. See Figure A16. 4.

  • Page 103: Motor Removal

    EasySampler Motor Removal 1. Loosen the socket screw from the brass damper using a 1/16 inch Allen wrench. See Figure A18. 2. Remove the damper from the top of the motor shaft, as shown in Figure A19. Damper Socket Screw Figure A18 Damper Motor…

  • Page 104
    EasySampler 3. Remove the four slotted screws from the top of the motor using a 7/32” flat head screw driv- er. See Figure A20. 4. Remove the motor from the motor shelf, See Figure A21. Slotted Screws Motor Figure A20 Motor Shelf Motor Figure A21…
  • Page 105: Motor Installation

    EasySampler Motor Installation 1. Place the motor into the motor shelf. 2. Pull the belt around the gear of the motor, see Figure A22 for location of gear and belt. 3. Attach the four slotted screws to the top of the motor using a 7/32” flat head screw driver, as shown in Figure A22.

  • Page 106
    EasySampler 5. Tighten the socket screw to the damper using a 1/16 inch Allen wrench. See Figure A24. Damper Socket Screw Figure A24…
  • Page 107: Pc Board Removal

    EasySampler PC Board Removal 1. Disconnect the power cable from the PC board. See Figure A25. 2. Remove the hex nut from the heat-sink plate using a 1/4 inch open end wrench. See Figure A26. Power Cable Figure A25 Hex Nut Heat Sink Plate Figure A26…

  • Page 108
    EasySampler 3. Remove the four screws from the PC board using a #2 Phillips screw driver. See Figure A27. The lower left screw may be hidden by the EMI suppressor box. If so, peel back the tape to reveal the screw. 4.
  • Page 109: Pc Board Installation

    EasySampler PC Board Installation 1. Place the PC Board, as shown in Figure A29, onto the base housing. 2. Install the four screws to the CPU board using a #2 Phillips screw driver, see Figure A30. Figure A29 Screws Screws Figure A30…

  • Page 110
    EasySampler 3. Install the hex nut to the heat-sink plate using a 1/4 inch open end wrench, see Figure A31. 4. Connect the power cable to position J2 on the PC board, see Figure A32. Hex Nut Heat Sink Plate Figure A31 PC Board Power…
  • Page 111
    Specifications Specifications Analyzer Specifications Moderate Complexity CLIA Classification: Whole Blood, Serum, Plasma or Urine. Sample: Direct (undiluted) measurement Ion Selective Electrode 100 μl Whole Blood, Serum, Plasma or 400 μL diluted (1:10) Urine Sample Size: Blood Na + : 20-200 mmol/L Urine Na + : 40-1000 mmol/L Detection K + : 0.2-40…
  • Page 112: Easysampler Specifications

    Specifications Analyzer Specifications (continued) 15-32 o C (60-90 o F), <85% humidity Ambient Conditions: 100/115~VAC, 50-60 Hz, 0.8A/250 V fuse or 220~VAC, 50-60 Hz 0.4A/250 V fuse Power: Refer to the label on the back of the analyzer for preset power and fuse settings.

  • Page 113: Display Messages

    Appendix Display Messages “!!” Na/K/Cl result(s) out of analyzer measurement range. ***** Result not reported because electrode is not calibrated. ABNORMAL CONTROL? Allows operator to select analysis of abnormal control. ABNORMAL STATS? Allows operator to display or print statistics for stored abnormal control results. AIR IN CLEANER Cleaning solution not detected correctly.

  • Page 114
    Appendix BLOOD K LIMITS? Allows operator to set normal K range. BLOOD Li LIMITS? Allows operator to set normal Li range. BLOOD NA LIMITS? Allows operator to set normal Na range. CAL VALUES? Allows operator to display last calibration values. CALIBRATE NOW? Allows operator to calibrate analyzer.
  • Page 115
    Appendix *DIAGNOSTICS?* Allows operator to check key analyzer functions. DILUTED 1:10? Analyzer requests confirmation that urine sample has been diluted. DRIFT Cl Chloride electrode is drifting outside of preset electrode operating range. DRIFT K Potassium electrode is drifting outside of preset electrode operating range. DRIFT Li Lithium electrode is drifting outside of preset electrode operating range.
  • Page 116
    Appendix LAST RESULT? Allows operator to display last analysis result. Li BELOW 0.2mM/L Lithium sample concentration is below EasyLyte measurement range. Li HIGH.. Lithium result obtained is higher than the preset normal range (flashing display). Li LOW… Lithium result obtained is lower than the preset normal range (flashing display). MEAN VALUES? Allows operator to display mean values of quality control results.
  • Page 117
    Appendix NORMAL STATS? Allows operator to display or print statistics for stored normal control results. NOT ENOUGH DATA! In order to calculate statistics, the Analyzer requires 5 stored control results of both normal and abnormal levels. *OPER FUNCTS?* Allows operator to select “On Demand” flush, purge or calibration functions. *PACK USAGE?* Allows operator to display % pack expended.
  • Page 118
    Appendix PUMPCAL OK Successful completion of Pumpcal Test. PUMPCAL TEST? Allows operator to test pump calibration. PURGING STD A Standard A fluid line is being purged. PURGING STD B Standard B fluid line is being purged. PURGING FLUSH Flush Solution fluid line is being purged. *QUAL CONTROL? Allows operator to choose quality control functions.
  • Page 119
    Appendix SOL’N PURGE? Allows operator to purge air from fluid lines. ***STANDBY*** Appears when Analyzer is in Standby mode. STANDBY IN 1 MIN Analyzer will go into standby in 1 minute (Flashing Display). STANDBY MODE? Allows operator to put analyzer in standby mode to conserve fluids. STATISTICS? Allows operator to display statistics for all control results.
  • Page 120: Qc Test? Flowcharts

    QC TEST? Flowchart (Service Diagnostics) DATE/RANGES? BLOOD Na LIMITS? 246.0 Yes/No QC TEST? TURN ON MV’S? CHECK SUM CHECK CAL CONSTS? RELEASE DATE RAM/PROM TEST? RAM CHECKOUT? CLEAR MEMORY SIPPER TEST? CALIB CAROUSEL? FIND HOME? PUMP TEST? CYCLE TRAY? TEST CAROUSEL? CAL VALUES? SOLUTION PURGE? ELECTRODES VALUES?

  • Page 121: Cpu Board — Analyzer Series Below 6100

    CPU Board – Analyzer Series below 6100…

  • Page 123
    CPU Board…
  • Page 125
    Power Supply – Analyzer Series under 6100…
  • Page 126: Transformer Assembly

    Transformer Assembly…

  • Page 127: Easysampler Pcb

    EasySampler PCB…

  • Page 130
    For Service Or Technical Assistance Call: 1-800-777-5983 (In Continental US) 1-617-275-4892 (International) Medica Corporation 5 Oak Park Drive Bedford, MA 01730 EDIC@…


Is the category for this document correct?


Thank you for your participation!

No more boring flashcards learning!

Learn languages, math, history, economics, chemistry and more with free Studylib Extension!

  • Distribute all flashcards reviewing into small sessions
  • Get inspired with a daily photo
  • Import sets from Anki, Quizlet, etc
  • Add Active Recall to your learning and get higher grades!

Add to Chrome
It’s free

Особенности анализатора EASYLYTE

• используются сменные пакеты, содержащие рабочие растворы и емкость для отходов (количество исследований зависит от потока, может достигать 1500)
• управление двумя клавишами: «Да» и «Нет»
• сменные электроды, не требующие техобслуживания
• быстрая замена компонентов без использования инструментов
• калибровка: автоматическая одноточечная и двухточечная с установленным промежутком времени, по требованию
• сохранение в памяти результатов измерений 125 проб и по 20 контрольных материалов каждого из трех уровней
• автоматическая подача проб при подключении устройства EasySampler
• возможность исследования биопроб животных – сыворотки, цельной крови или плазмы (EasyLyte и EasyLyte Plus)
• русифицированный интерфейс
• встроенный термопринтер (L=57 мм)

Производительность
до 60 проб в час

Пробы
• 100 мкл (плазма, сыворотка, цельная кровь) или 40 мкл (моча в разведении 1:10, кроме EasyLyte Calcium); для EasyLyte Calcium рекомендуется цельная кровь; системы взятия крови с антикоагулянтом – литиевая или натриевая соль гепарина
• для приборов EasyLyte, EasyLyte Plus, EasyLyte Lithium подача пробы цельной крови возможна через капилляр, объем пробы – 60 мкл

Габариты: (ДхШхВ) 24х20х42 см
Вес: 5,8 кг (с пакетом реагентов)

Flow Chart

Whole Blood, Serum, or Plasma Analysis

End of Day Cleaning

Fluid Conservation, Maintenance

Urine Analysis

Stored Results Displayed, Printed, Deleted

Check % Pack Usage, Install New Pack,

Reset % Countdown

Run Controls, See Last Control,

See/Print Statistics, Delete Results

«On Demand» Calibration, Wash, Solutions Purge,

User Options, Probe Wiping, I.D.#, Syringe Sample,

Set Correlation,Vet Mode, Select Language,

AutoSampler, Printer Off

Check: Calibration Values

Electrode Values

Fluid Flow

Sample Detector

Pump Calibration

Change Day/Time, AutoCal,

Set Na/K/Cl/Li Ranges

Copyright ©1995 Medica

Information contained herein is believed by Medica to be accurate; in any event, no responsibility, whether express or implied, is assumed hereby

by Medica for or in connection with the use thereof, or for infringement of any third party rights which might arise therefrom, or from any represen-

tation or omissions contained therein. Information is subject to change and/or updating without notice.

CALIBRATE NOW?

Yes

ANALYZE BLOOD?

Yes

No

DAILY CLEANER?

Yes

No

STANDBY MODE?

Yes

No

✽SECOND MENU?✽

Yes

No

✽ ANALYZE URINE? ✽

Yes

No

✽ SEE RESULTS? ✽

Yes

No

✽ PACK USAGE? ✽

Yes

No

✽ QUAL CONTROL? ✽

Yes

No

✽ OPER FUNCTS? ✽

Yes

No

✽ DIAGNOSTICS? ✽

Yes

No

✽ DATE/RANGES? ✽

Yes

No

Анализатор электролитов крови EasyLyte ф.Medica Corporation

scratch2010 Дата: Понедельник, 30.Янв.2012, 09:35 | Сообщение # 1

инженер

На заявках

Российская Федерация



волгоград

Анализатор электролитов крови EasyLyte компании Medica Corporation, США.
У кого есть документация, сервис-мануал, каталог запчастей (с каталожными номерами для заказов). Имеется частая проблема в протекании, после промывки и продувки клапана и всей гидравлики проблема временно исчезает. Вот думаю на клапан (распределительный), в который вставляется пак. либо засор либо подсос воздуха. Хотелось бы попробовать его поменять, но не знаю как правильно называется и где лучше заказать.

.
Есть темы:
— Анализатор электролитов крови EasyLyte Calcium (Na/K/Ca/pH) (Неполадки/проблемы, опыт и советы по их устранению);
— Анализатор электролитов крови EasyLyte Plus (Na/K/Cl) (Неполадки/проблемы, опыт и советы по их устранению);
— Анализаторы ProLyte и EasyLyte — в чём различие?.

 
Serg74 Дата: Понедельник, 30.Янв.2012, 11:52 | Сообщение # 2

Инженер лаборатория ИВЛ

На заявках

Российская Федерация



Челябинск

scratch2010,
не забывай что клапан сменный, гарантированная наработка 12 месяцев, посмотри сколько прибору. Правда стоят по 2-4 года, но проблема еще и в объемах, чем больше тем быстрее отказывает, хотя там тупо резинки, но кривизна рук сказывается, да и если фильтр для иглы не используют, то тоже грязи насобирывает. Попробуй помой жесткой промывкой от гематологов, понемногу, шприцом. Сервисной нет, по расходке есть две картинки из пользовательской, прилагаю. По ценам надо искать, на вскидку http://www.unimedao.ru/price_list?id=11547#1 скачай каталог, там есть все каталожные номера и паков и электродов и расходки.

Сообщение отредактировал Serg74Понедельник, 30.Янв.2012, 11:55

 
scratch2010 Дата: Понедельник, 30.Янв.2012, 15:35 | Сообщение # 3

инженер

На заявках

Российская Федерация



волгоград

Во спасибо, а пользовательская в электронном виде есть?

 
Serg74 Дата: Понедельник, 30.Янв.2012, 16:54 | Сообщение # 4

Инженер лаборатория ИВЛ

На заявках

Российская Федерация



Челябинск

scratch2010,
не нашел, а это так, щелкнул для информации.

 
KarelMT Дата: Понедельник, 30.Янв.2012, 18:39 | Сообщение # 5

Инженер по обслуживанию всего

На заявках

Российская Федерация



Северо-Запад

Клапан растворов он называется, дело процентов на 95 в нем. Сервисная у меня есть, только там ничего интересного нет. Медиковские приборы отличаются крайней простотой в обслуживании и замене запчастей, с этим, по идее, любой лаборант справится. А если что-то с электроникой, то без имитатора электродов все равно прибор не настроишь, а половина сервисной инструкции этому посвящена.

 
Saimon47sa Дата: Понедельник, 30.Янв.2012, 23:58 | Сообщение # 6

Стажер

Инженер

На заявках

Российская Федерация



Москва

У кого-нибудь при откручивании двигателя пробозаборника, крепления к корпусу отламывались?
Сейчас пока на супер клее с обезжиривание сидит или переклеить их на поксипол?

Сообщение отредактировал Saimon47saВторник, 31.Янв.2012, 00:00

 
Serg74 Дата: Среда, 05.Сен.2012, 08:17 | Сообщение # 7

Инженер лаборатория ИВЛ

На заявках

Российская Федерация



Челябинск

Добрый день!
Аппарат Na/K/Ca/pH, отработал на новых электродах 3 месяца, начались проблемы с калибровкой Na, первоначально помогала промывка, но перестала, электрод почистил штатными средствами, откалибровался, но с дрейфом, отработал день и снова перестал. Попробовал поставить электрод с другого, картина таже. Поменял референсный, Na заработал, но перестал калиброваться Са, меняем референсный — все наоборот. Да, принтер умер уже давно, поэтому уточнить, что при калибровке происходит, не получается. Уже сломал голову mad
Может у кого-нибудь есть идеи и мысли. unknown

 
chesterblack Дата: Среда, 05.Сен.2012, 14:39 | Сообщение # 8

Заглянувший

руководитель организации

На заявках

Российская Федерация



Сыктывкар

Может звонок в сервисную службу Аналитики (Москва) поможет 8-495-737-03-63

 
Serg74 Дата: Среда, 05.Сен.2012, 14:50 | Сообщение # 9

Инженер лаборатория ИВЛ

На заявках

Российская Федерация



Челябинск

chesterblack,
спасибо за номер. Еще завтра помучаю и отзвонюсь от самого прибора. wink

 
gi7128 Дата: Среда, 05.Сен.2012, 16:00 | Сообщение # 10

Стажер

сервис-инженер

На заявках

Российская Федерация



СПб

(495) 774-46-77 круглосуточный сервис для анализаторов неотложных состояний MEDICA Corp. и др

 
Bardey Дата: Среда, 05.Сен.2012, 18:22 | Сообщение # 11

Завсегдатай

инженер

На заявках

Российская Федерация



Воронеж

Serg74,Привет,обычно мы меняем все электроды или по истории замены электродов.

Обчно по крутизне и напряжению станет ясно на скольколько работоспобны электроды.

Сообщение отредактировал BardeyСреда, 05.Сен.2012, 18:28

 
Svetaslav Дата: Понедельник, 14.Янв.2013, 10:51 | Сообщение # 12

Участник

инженер-электроник

На заявках

Российская Федерация



Владимир

Добавил EasyLyte Operator’s Manual — Rus в хранилище, еще бы сервис мануал кто бы положил, со схемой. После Нов Года явился на работу и обнаружил что протекла крыша над EasyLyte Calcium. Вода попала в блок электродов,дорожки кое-где отгорели в уголь. Восстановил дорожки, собрал-идёт нагрев, на кнопки не реагирует, по плате полазил,обнаружил отсутствие U=+5V — питание микрух, идет с U12(78L05)-между Вых и Землёй R=10 Ом. Искал КЗ, обнаружил в микроконтроллере U10 P80C552EBA, смог заказать EFA, надеюсь подойдет.

 
vlmil Дата: Понедельник, 18.Мар.2013, 16:40 | Сообщение # 13

Стажер

сервисный инженер

На заявках

Российская Федерация



Рузаевка

Присоединяюсь. Аналогично с электродом Na. Менял на новый промывкой дотягивает до 47, маловато. Подскажите какое напряжение должно быть на нормальном Na, у меня 125.
Заказали весь комплект электродов хочу заменить.

 
Alice Дата: Вторник, 02.Июл.2013, 16:56 | Сообщение # 14

Стажер

инженер

На заявках

Российская Федерация



Москва

Здравствуйте коллеги! На Еasylyte calcium после ежедневной очистки,в калибровке выдаётся ошибка «ПОМПА НЕ В НОРМЕ»? Перепробовали всё! В чем проблема?

 
Alexium Дата: Вторник, 02.Июл.2013, 17:08 | Сообщение # 15

Инженер медтехник

На заявках

Российская Федерация



Саратов

Цитата (Alice)

Перепробовали всё!

Это что-то из Камасутры?

 
doctor_farsh Дата: Среда, 03.Июл.2013, 01:09 | Сообщение # 16

Передовик

инженер по ремонту медоборудования

На заявках

Российская Федерация



Благовещенск

у нас тупо часто его засирают кровякой, сгустками и прочей хренью матерился , ругался с ними сейчас стали поаккуратней относиться и забыл я про этот аппарат только электроды хожу меняю

 
Alice Дата: Вторник, 09.Июл.2013, 13:57 | Сообщение # 17

Стажер

инженер

На заявках

Российская Федерация



Москва

Цитата (Alexium)

Это что-то из Камасутры?
6808860.jpg(74Kb)

-не помогло,увы

 
Alexium Дата: Вторник, 09.Июл.2013, 14:36 | Сообщение # 18

Инженер медтехник

На заявках

Российская Федерация



Саратов

Цитата (Alice)

-не помогло,увы

я так понимаю, что информации что делали и какой результат не получить.

 
SAI Дата: Четверг, 11.Июл.2013, 10:43 | Сообщение # 19

Стажер

Инженер медтехник

На заявках

Российская Федерация



Череповец

Цитата (Svetaslav)

Здравствуйте коллеги! На Еasylyte calcium после ежедневной очистки,в калибровке выдаётся ошибка «ПОМПА НЕ В НОРМЕ»? Перепробовали всё! В чем проблема?

Проблема в «cмачиваемости » трубки помпы насоса.
Промойте внутри трубку помпы гипохлоридом (или белизной из магазина бытовой химии — это тоже 5% водный р-р гипохлорида),
если трубка имеет механические дефекты, поменяйте её на новую.

 
Касатик Дата: Пятница, 12.Июл.2013, 08:45 | Сообщение # 20

Инженер

На заявках

Российская Федерация



СПб

Коллеги, обращаюсь с вопросом. Устанавливал EasyLyte с кальцием и pH. При калибровке сразу выдает «ошибка 11». Судя по мануалу проблема с датчиком считывания RFID метки. подскажите, куда смотреть, чего крутить?

 
OBEY Дата: Понедельник, 05.Авг.2013, 22:03 | Сообщение # 21

Участник

лабораторная техника

На заявках

Украина



Киевский Хунтанат

Ребя, кто-то пробывал ставить на EasyLyte Plus Na/K/Cl производства Medica Corp., USA электроды от Ilyte (Instrumentation Laboratory, Italy)?

 
Alexium Дата: Понедельник, 05.Авг.2013, 22:15 | Сообщение # 22

Инженер медтехник

На заявках

Российская Федерация



Саратов

Цитата (OBEY)

электроды от Ilyte (Instrumentation Laboratory, Italy)

Это один и тот же прибор. Производитель Medica.

 
OBEY Дата: Пятница, 09.Авг.2013, 17:31 | Сообщение # 23

Участник

лабораторная техника

На заявках

Украина



Киевский Хунтанат

Цитата (Alexium)

Это один и тот же прибор. Производитель Medica.

Вот что ответил сервис Instrumentation Laborotary:

Dear Sergey,
Even if the two instruments ( EasyLyte Plus and ILyte) are manufactured by the same company (Medica Corp.), the parts and the solutions pack are different. The electrodes of the ILyte doesn’t fit in the EasyLyte instruments, and also the position of the calibrators inside of the solution pack, are different. If you try to use a solution pack of the ILyte on the EasyLyte (and also the contrary) surely you will find an error during the calibration, on all channels, since the solutions for calibration (Cal1 and Cal2) have not the concentration that the software expects to “see».
Best Regards

Думаю, надо брать электроды и пробывать. Пак понятно, что если отверстие для слива и стандарт А перепутаны, то не подойдут. Может не хотят палится просто — судя с их ответа.

Сообщение отредактировал OBEYПятница, 09.Авг.2013, 17:32

 
Alexium Дата: Четверг, 05.Сен.2013, 09:27 | Сообщение # 24

Инженер медтехник

На заявках

Российская Федерация



Саратов

Вот что в реальной жизни, если не слушать барыг

 
san-che30 Дата: Понедельник, 18.Ноя.2013, 11:21 | Сообщение # 25

Участник

Биомедицинские аппараты и системы

На заявках

Российская Федерация



Москва

Прибор калибруется, ошибок не выдает.Напряжение электродов попадает в допустимый диапазон.После взятия пробы крови не выдает значения Na,вместо значения натрия выдает **** ,а расчет К в норме .С чем это может быть связано(электроды новые)?

 
eroigor Дата: Понедельник, 18.Ноя.2013, 15:13 | Сообщение # 26

Завсегдатай

Cервисный инженер

На заявках

Российская Федерация



IzheVWsk

san-che30, Какое значение крутизны Na электрода при калибровке?
Контроль гонял?

 
san-che30 Дата: Понедельник, 18.Ноя.2013, 16:44 | Сообщение # 27

Участник

Биомедицинские аппараты и системы

На заявках

Российская Федерация



Москва

eroigor,Что то я не припомню,чтобы при калибровке выводилось значение крутизны на дисплей.Измерения проводил с помощью контроля

 
babysh-m Дата: Понедельник, 18.Ноя.2013, 19:57 | Сообщение # 28

сервисный инженер

На заявках

Российская Федерация



Екатеринбург

san-che30, Значения крутизны можно посмотреть в подразделе меню, смотри описание анализатора. Да, а после калибровки на экране что пишет?

 
eroigor Дата: Вторник, 19.Ноя.2013, 08:05 | Сообщение # 29

Завсегдатай

Cервисный инженер

На заявках

Российская Федерация



IzheVWsk

san-che30, Если уверен в том,что электрод бракованный, смотри вложение….

 
san-che30 Дата: Вторник, 19.Ноя.2013, 09:07 | Сообщение # 30

Участник

Биомедицинские аппараты и системы

На заявках

Российская Федерация



Москва

babysh-m ,»Измерять кровь?»

 
babysh-m Дата: Вторник, 19.Ноя.2013, 20:24 | Сообщение # 31

сервисный инженер

На заявках

Российская Федерация



Екатеринбург

san-che30, Если нормально прошла калибровка, на печать должна быть выдана надпись что калибровку прошли оба электрода.

 
GUNIA Дата: Среда, 30.Мар.2016, 23:02 | Сообщение # 32

Участник

ИНЖЕНЕР

На заявках

Российская Федерация



МОСКВА

а что за красная муфта клапана два прибора ,но не видел на трубке их??

 
black_fury Дата: Воскресенье, 03.Апр.2016, 10:43 | Сообщение # 33

Техник

Инженер по ремонту мед.оборудования

На заявках

Российская Федерация



Челябинск

Alexium, это вроде прародитель)))
GUNIA, это вы где такое видели?

 
Константин1969 Дата: Четверг, 08.Сен.2016, 22:50 | Сообщение # 34

Завсегдатай

Инженер-приборостроитель

На заявках

Российская Федерация



Сергиев-Посад

Периодически прибор не показывал калий. Сначала калибровка была как на фото 1. Потом стали появляться звёздочки. Электроды не меняли давно. Сегодня заменил все электроды, трубки, раствор. Планово. Но калий упорно не хочет калиброваться. Пробовал и старый и новый электрод. Ткните почему.

 
black_fury Дата: Пятница, 09.Сен.2016, 04:02 | Сообщение # 35

Техник

Инженер по ремонту мед.оборудования

На заявках

Российская Федерация



Челябинск

Константин1969, электрод сравнения меняли?

 
Константин1969 Дата: Пятница, 09.Сен.2016, 08:20 | Сообщение # 36

Завсегдатай

Инженер-приборостроитель

На заявках

Российская Федерация



Сергиев-Посад

Электроды заменены все, годны до 5. 05. 2017.Такое ощущение что контакта на калии нет. Или пропадает. Но там отвёртку нужно длинную, вчера под рукой не было.

 
black_fury Дата: Пятница, 09.Сен.2016, 08:30 | Сообщение # 37

Техник

Инженер по ремонту мед.оборудования

На заявках

Российская Федерация



Челябинск

Константин1969, отвёрткой и прочими (надфилями) не вздумай залазить туда, имеется печальный опыт (не мой). Можешь взять зубочистку, не много намочить (спиртом) и протереть внутри, максимум скрепкой. Но на практике такая неисправность была только на референсном электрод…

 
fred1 Дата: Пятница, 09.Сен.2016, 15:01 | Сообщение # 38

инженер BeckmanCoulter, Siemens MRT

На заявках

Российская Федерация



Анапа

Проверь контакт ! До платы !

Для информации, на этих приборах со временем перестает работать референтный электрод, причина, наслоение несмываемых отложений на мембранной сборке, эта беда довольно легко ремонтируется!
Надо вытащить мембранную сборку , аккуратно слив хлористый калий, отклеить липкую ленту, под ней есть маленькая дырочка,надо просто сместить ленту на новое место или в сторону и заклеить все на место ! Проверено неоднократно работает как новая !!!

Удачи !

Сообщение отредактировал fred1Пятница, 09.Сен.2016, 15:01

 
doctor_farsh Дата: Суббота, 10.Сен.2016, 00:23 | Сообщение # 39

Передовик

инженер по ремонту медоборудования

На заявках

Российская Федерация



Благовещенск

fred1, +1 Константин1969, ищи проблему в мембране такая же фигня, только с Натрием была. Лента в потраха на мембране размоталась, пришлось использовать вместо нее пару витков фумленты предварительно размоченной в растворе, что в камеру заливаешь. так мы проработали 4 месяца biggrin

 
Константин1969 Дата: Суббота, 10.Сен.2016, 01:23 | Сообщение # 40

Завсегдатай

Инженер-приборостроитель

На заявках

Российская Федерация



Сергиев-Посад

Референтный электрод новый, в том-то и дело. Что новые что старые электроды ставил, картина одна. Естественно все способы попробую что советовали. Всем спасибо за отклики. Судя по схеме измерений если реф сдох то по идее калибровки пл Хл и Натр тоже быть не должно. Так? На мембране обычная лента клейкая или что-то специальное?
Отвёртку естественно имел ввиду для снятия задней стенки, два верхних винта. biggrin

Сообщение отредактировал Константин1969Суббота, 10.Сен.2016, 01:25

 