FREQUENCY COUNTER
USER MANUAL
1. PRODUCT INTRODUCTION
The intelligent counters are measuring frequency and period from
the range of 0.01Hz to 120MHz on channel A, and 50MHz to
1.3(2.7)GHz on channel B.
The products also provide special features of high resolution and
sensitivity for value reading expediency.
1
Summary of Contents for GW Instek GFC-8131H
FREQUENCY COUNTER
USER MANUAL
CONTENTS PAGE
1.
PRODUCT INTRODUCTION………………………………………
1
2.
TECHNICAL SPECIFICATION……………………………
2
3.
PRECAUTIONS BEFORE OPERATION…….……………
3-1.Unpacking the Instrument………………….……………
3-2.Checking the Line Voltage…………………..………….
3-3.Equipment Installation and Operation…………………
4
4
4
5
4.
PANEL INTRODUCTION……………………..……………
6
5.
APPLICATION……………………………………………….
9
6.
MAINTENANCE……………………………………………..
6-1.Standard Method for Calibration………………………
6-2.Cleaning……………………………………………………
11
11
12
FREQUENCY COUNTER
USER MANUAL
1
1. PRODUCT INTRODUCTION
The intelligent counters are measuring frequency and period from
the range of 0.01Hz to 120MHz on channel A, and 50MHz to
1.3(2.7)GHz on channel B.
The products also provide special features of high resolution and
sensitivity for value reading expediency.
Loading…
FREQUENCY COUNTER |
|
USER MANUAL |
|
CONTENTS |
PAGE |
1. PRODUCT INTRODUCTION……………………………………… |
1 |
2.TECHNICAL SPECIFICATION…………………………… 2
3. |
PRECAUTIONS BEFORE OPERATION…….…………… |
4 |
3-1.Unpacking the Instrument………………….…………… |
4 |
|
3-2.Checking the Line Voltage…………………..…………. |
4 |
|
3-3.Equipment Installation and Operation………………… |
5 |
|
4. |
PANEL INTRODUCTION……………………..…………… |
6 |
5.APPLICATION………………………………………………. 9
6.MAINTENANCE…………………………………………….. 11
6-1.Standard Method for Calibration……………………… |
11 |
6-2.Cleaning…………………………………………………… |
12 |
FREQUENCY COUNTER
USER MANUAL
1. PRODUCT INTRODUCTION
The intelligent counters are measuring frequency and period from the range of 0.01Hz to 120MHz on channel A, and 50MHz to 1.3(2.7)GHz on channel B.
The products also provide special features of high resolution and sensitivity for value reading expediency.
1
FREQUENCY COUNTER
USER MANUAL
2. TECHNICAL SPECIFICATIONS |
||||||
CHANNEL A |
||||||
Range |
Coupling |
AC |
DC |
|||
FREQ A |
30Hz to 120MHZ 0.01Hz to 120MHz |
|||||
PRID A |
8ns to 30ms |
8ns to 100s |
||||
50mVrms Max. to 10kHz. |
||||||
Sensitivity |
25mVrms Max. to 80MHz. |
|||||
35mVrms Max. to 120MHz. |
||||||
(rms) |
||||||
(If frequency |
10MHz, sensitivity figures are for 4 |
|||||
stable digits of reading.) |
||||||
Coupling |
AC or DC switchable. |
|||||
Filter |
Low pass, switchable in or out on channel A. |
|||||
3dB point of nominally 100kHz. |
||||||
Impedance |
1MΩ nominal shunted by less than 40pF. |
|||||
Attenuator |
×1 or ×20 nominal. |
|||||
Trigger Level |
Variable between +2.5V DC and –2.5V DC. |
|||||
Damage Level |
AC & DC ×1 |
|||||
DC to 2.4kHz |
250V(DC+AC rms) |
|||||
2.4kHz to 100kHz |
600kV rms Hz |
FREQ |
||||
100kHz |
6V rms |
|||||
AC & DC × 20 |
||||||
DC to 20kHz |
500V(DC+AC peak) |
|||||
20kHz to 100kHz |
10MV rms Hz |
FREQ. |
||||
100kHz |
100V rms |
|||||
CHANNEL B |
||||||
Range |
50MHz to 1.3GHz (GFC-8131H) |
|||||
50MHz to 2.7GHz (GFC-8270H) |
||||||
Sensitivity |
GFC-8131H |
GFC-8270H |
||||
25mVrms Max. to 80MHz |
25mVrms Max. to 80MHz |
|||||
15mVrms Max. to 700MHz |
15mVrms Max. to 1GHz |
|||||
25mVrms Max. to 1GHz |
25mVrms Max. to 2GHz |
|||||
40mVrms Max. to 1.3GHz |
50mVrms Max. to 2.7GHz |
FREQUENCY COUNTER
USER MANUAL
Coupling |
AC |
|
Impedance |
50Ω |
|
Max. Input Level |
3Vrms sine wave |
|
General |
||
Gate time displays at least 7 digits for 1s, 6 digits for |
||
100ms, 5 digits for 10ms respectively. |
||
Resolution |
The maximum resolution for frequency measurement is |
|
100nHz for 1Hz, and 0.1Hz for 100MHz input |
||
respectively, and for period measurement is 10nS for |
||
1Hz, 0.1fs (f=10-15 ) for 100MHz input respectively. |
||
Frequency————10MHz |
||
Time Base |
Aging Rate————1ppm per month |
|
Temperature———-5ppm, 23 ±5 |
||
Line Voltage———±0.005ppm for ±10% variation |
||
Accuracy |
±Resolution ± Time base error. |
|
Gate Time |
Continuously vary from 10ms to 10s or 1 period of the |
|
input depending on whichever is greater. |
||
Display |
8 digits and overflow indicators |
|
Operation |
0 ~40 |
|
Temperature |
||
Operation |
10~80% RH |
|
Humidity |
||
Power |
100/120/220/230±10%, 50/60Hz selectable internally. |
|
Requirement |
Max. consumption power is 15VA. |
|
Dimension |
Approx. 230(W) × 95(H) × 280(D) m/m. |
|
Weight |
Approx. 2.0 kgs. |
|
Instruction manual……………× 1 |
||
Accessories |
Power cord……………………× 1 |
|
Test lead GTL-101 …………..× 1 |
||
GTL-110 …………..× 1 |
FREQUENCY COUNTER
USER MANUAL
3.PRECAUTIONS BEFORE OPERATION
3-1.Unpacking the Instrument
The instrument has been fully inspected and tested before shipping from the factory. Upon receiving the instrument, please unpack and inspect it to check if there is any damages caused during transportation. If any sign of damage is found, notify the bearer and/or the dealer immediately.
3-2.Checking the Line Voltage
The instrument can be applied any kind of line voltage shown in the table below. Please check the line voltage indicated in the label attached on the real panel to replace correct fuses.
WARNING. To avoid electrical shock the power cord protective grounding conductor must be connected to ground.
When line voltages are changed, replace the required fuses shown as below:
Line voltage |
Range |
Fuse |
Line voltage |
Range |
Fuse |
100V |
90-110V |
T200mA |
220V |
198-242V |
T100mA |
120V |
108-132V |
250V |
230V |
207-253V |
250V |
WARNING. To avoid personal injury, disconnect the power cord before removing the fuse holder.
FREQUENCY COUNTER
USER MANUAL
3-3.Equipment Installation, and Operation
Ensure there is proper ventilation for the vents in the case. If this equipment is used not according to the specification, the protection provided by the equipment may be impaired.
You can only view or download manuals with
Sign Up and get 5 for free
Upload your files to the site. You get 1 for each file you add
Get 1 for every time someone downloads your manual
Buy as many as you need
FREQUENCY COUNTER
USER MANUAL
2. TECHNICAL SPECIFICATIONS
CHANNEL A
Coupling
Range
FREQ A
30Hz to 120MHZ
PRID A
8ns to 30ms
50mVrms Max. to 10kHz.
25mVrms Max. to 80MHz.
Sensitivity
35mVrms Max. to 120MHz.
(rms)
(If frequency < 10MHz, sensitivity figures are for 4
stable digits of reading.)
Coupling
AC or DC switchable.
Low pass, switchable in or out on channel A.
Filter
3dB point of nominally 100kHz.
1MΩ nominal shunted by less than 40pF.
Impedance
Attenuator
×1 or ×20 nominal.
Trigger Level
Variable between +2.5V DC and –2.5V DC.
AC & DC ×1
Damage Level
DC to 2.4kHz
2.4kHz to 100kHz
>100kHz
AC & DC × 20
DC to 20kHz
20kHz to 100kHz
>100kHz
CHANNEL B
Range
50MHz to 1.3GHz (GFC-8131H)
50MHz to 2.7GHz (GFC-8270H)
Sensitivity
GFC-8131H
25mVrms Max. to 80MHz
15mVrms Max. to 700MHz
25mVrms Max. to 1GHz
40mVrms Max. to 1.3GHz
AC
DC
0.01Hz to 120MHz
8ns to 100s
250V(DC+AC rms)
600kV rms Hz FREQ
6V rms
500V(DC+AC peak)
10MV rms Hz FREQ.
100V rms
GFC-8270H
25mVrms Max. to 80MHz
15mVrms Max. to 1GHz
25mVrms Max. to 2GHz
50mVrms Max. to 2.7GHz
2
FREQUENCY COUNTER
USER MANUAL
AC
Coupling
50Ω
Impedance
Max. Input Level 3Vrms sine wave
General
Gate time displays at least 7 digits for 1s, 6 digits for
100ms, 5 digits for 10ms respectively.
The maximum resolution for frequency measurement is
Resolution
100nHz for 1Hz, and 0.1Hz for 100MHz input
respectively, and for period measurement is 10nS for
1Hz, 0.1fs
(f=10
Frequency————10MHz
Aging Rate————1ppm per month
Time Base
Temperature———-5ppm, 23℃±5℃
Line Voltage———±0.005ppm for ±10% variation
±Resolution ± Time base error.
Accuracy
Continuously vary from 10ms to 10s or 1 period of the
Gate Time
input depending on whichever is greater.
Display
8 digits and overflow indicators
Operation
0℃~40℃
Temperature
Operation
10~80% RH
Humidity
100/120/220/230±10%, 50/60Hz selectable internally.
Power
Requirement
Max. consumption power is 15VA.
Dimension
Approx. 230(W) × 95(H) × 280(D) m/m.
Weight
Approx. 2.0 kgs.
Instruction manual……………× 1
Power cord……………………× 1
Accessories
Test lead GTL-101 …………..× 1
GTL-110 …………..× 1
-15
) for 100MHz input respectively.
3
FREQUENCY COUNTER
USER MANUAL
CONTENTS PAGE
1. PRODUCT INTRODUCTION…………………………..…………. 1
2. TECHNICAL SPECIFICATION…………………………… 2
3. PRECAUTIONS BEFORE OPERATION…….…………...
3-1.Unpacking the Instrument………………….……………
3-2.Checking the Line Voltage…………………..………….
3-3.Equipment Installation and Operation…………………
4
4
4
5
4. PANEL INTRODUCTION……………………..…………… 6
5. APPLICATION………………………………………………. 9
6. MAINTENANCE……………………………………………..
6-1.Standard Method for Calibration………………………
6-2.Cleaning……………………………………………………
11
11
12
FREQUENCY COUNTER
USER MANUAL
1
1. PRODUCT INTRODUCTION
The intelligent counters are measuring frequency and period from
the range of 0.01Hz to 120MHz on channel A, and 50MHz to
1.3(2.7)GHz on channel B.
The products also provide special features of high resolution and
sensitivity for value reading expediency.
Note for Owners:
Guidesimo.com webproject is not a service center of GW Instek trademark and does not carries out works for diagnosis and repair of faulty GW Instek GFC-8131H equipment. For quality services, please contact an official service center of GW Instek company. On our website you can read and download documentation for your GW Instek GFC-8131H device for free and familiarize yourself with the technical specifications of device.
-
Fluke 93
FLUKE AND PHILIPS THE GLOBAL ALLIANCE IN TEST & MEASUREMENTFluke 93/95/97Philips PM93/95/97SCOPEMETERService ManualRuker 915970Philipe: 4622 872 05349920121Warning: These servicing instructions are foruseby qualified personneloniy. To reduce the riskof eiecthc shocks do not perform any servicing otherthan that specified in the operatinginstructions uniessyou are fully qualified to do so.FLUKE …
93 Measuring Instruments, 267
-
Milwaukee C12 CMH
Original instructionsOriginalbetriebsanleitungNotice originaleIstruzioni originaliManual originalOorspronkelijke gebruiksaanwijzingOriginal brugsanvisningOriginal bruksanvisningBruksanvisning i originalAlkuperäiset ohjeetΠρωτότυπο οδηγιών χρήσηςOrijinal işletme talimatıPůvodním návodem k používáníPôvodný návod na použitieInstrukcją oryginalnąEredeti használat …
C12 CMH Measuring Instruments, 43
-
Datakom DPR-350 Series
DPR-350 User Manual Ver_01 Firmware V-1.0 DPR-350 is a state of the art network analyzer and protection relay in 96×96 standard panel dimensions. Thanks to its low depth, it provides space economy. The DPR-350 features dual current measurement circuits. It is both a precise analyzer and a protection relay. Protection curves can be …
DPR-350 Series Measuring Instruments, 72
-
Aventech AIMMS-20
Rev: 1.00 Date: 11.11.2013 Page: 1 of 48AIMMS-20 INSTALLATION MANUAL Document-P/N: AIMMS20-IM © Aventech Research Inc. 2013. All Rights Reserved. AIMMS-20 Aircraft Integrated Meteorological Measurement System INSTALLATION MANUAL Document-P/N: AIMMS20-IM Aventech Research Inc. 110 Anne Street South, Unit 23 Barrie, Ontario, Canada L4N 2E3 Tel: (705) 722-4288 Fax: (705 …
AIMMS-20 Measuring Instruments, 48
-
AXE MMX-L
Rev 1.0 AL1 RPM M/min Y/min Ft/min AXE MICROPROCESS RPM&LINE-SPEED CONTROLLER METER MMX-L ■ Features ◎ Accept more type sensors(switch, encoder, proximity switch, …etc) finish RPM/LINE SPEED control ◎ Diameter(LINE-SPEED)/scale(RPM) can be modified (0.0001~9.9999) ◎ Accuracy 0.03% F.S. ◎ One alarm function ◎ Input range(0~30KHz),Readout range(0~99999) ◎ Man-m …
MMX-L Measuring Instruments, 3
-
Avtech H.265 NVR
h517_h8536_d8536_8536l_h8516_h2116_d2116_h1109_d1109_h1004_d1104_manual_V1.3 AVTECH H.265 NVR SERIES User Manual Please read instructions thoroughly before operation and retain it for future reference. For the actual display & operation, please refer to your device in hand. Free PC CMS software (CMS Lite) …
H.265 NVR Measuring Instruments, 94
-
Rigol DS1000E series
User’s Guide RIGOL Publication number UGA07118-1110 Dec. 2015 DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D © 2008 RIGOL Technologies, Inc. All Rights Reserved www.GlobalTestSupply.comFind Quality Products Online at: [email protected] …
DS1000E series Test Equipment, 166
-
LAPLACE INSTRUMENTS AC2000A
LAPLACE INSTRUMENTS LTD AC2000A USER MANUAL ISSUE 14 LAPLACE INSTRUMENTS LTD Tudor House, Grammar School Road, North Walsham, Norfolk NR28 9JH UK T: +44 (0) 16 92 40 20 70 E: [email protected] W: www.laplace.co.uk …
AC2000A Measuring Instruments, 59