Maxum ii руководство по эксплуатации

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Maxum II

PD PA AP

Explosion Protection Safety

Standards

Manual

August 2017

A5E02220442001 Rev 8

General Information for the

User

Safety Systems — Purging

Safety Systems — Oven

Valves, Detectors, and

External Systems

Safe Operation

1

2

3

4

5

loading

Related Manuals for Siemens Maxum II

Summary of Contents for Siemens Maxum II

  • Page 1
    General Information for the User Safety Systems — Purging Safety Systems — Oven Maxum II Valves, Detectors, and External Systems PD PA AP Explosion Protection Safety Safe Operation Standards Manual August 2017 A5E02220442001 Rev 8…
  • Page 2
    Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    Table of contents General Information for the User……………………7 Contacts……………………..10 Approved Use……………………11 Qualified Personnel……………………11 Applicable Standards and Regulations………………12 Safety Protection Principles………………..13 Intrinsically Safe Devices…………………..14 Safety Systems — Purging………………………17 Purging Overview……………………17 Electronics Purging System (without Automatic Purge Unit)……….18 2.2.1 Overview……………………..18 2.2.2 Instrument Air Inlet and Regulator………………20 2.2.3 Fast Purge Switch……………………21 2.2.4…

  • Page 4
    4.4.2 Methanator Maintenance Considerations…………….55 Safe Operation……………………….57 Maintenance Panel Functional Description…………….57 Maintenance Considerations for Maintenance Panel………….58 Color Touchscreen Display Functional Description……………58 Maintenance Considerations for Control Interface Module Display……..59 Steps for Safe Startup of Maxum II………………60 5.5.1 Overview……………………..60 5.5.2 Procedure……………………..60 Glossary…………………………63 Figures Figure 1-1 Reference of Hardware (MAA) to Applicable Sections of this Manual……….8…
  • Page 5
    Figure 4-1 Siemens Liquid Injection Valve………………..45 Figure 4-2 Siemens Liquid Injection Valve (Installed)……………….46 Figure 4-3 Heater for Siemens Liquid Injection Valve………………46 Figure 4-4 Explosion Proof/Flameproof Thermal Conductivity Detector…………49 Figure 4-5 Thermal Conductivity Detector Beads on Modular Oven Application Module (Intrinsically Safe TCD)………………………50…
  • Page 6
    Table of contents Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 7: General Information For The User

    These standards are to ensure that the Maxum II and related products can be safely installed and operated in hazardous areas. Each individual detail and component of these safety systems is important to help ensure that the chromatograph does not ignite flammable vapors and gases which may be present in the environment surrounding the analyzer.

  • Page 8: Figure 1-1 Reference Of Hardware (Maa) To Applicable Sections Of This Manual

    General Information for the User Exhaust Di user Electronics Automatic Purge Unit (APU) Sections 2.2, 2.3 Enclosure (Purged) or Pressure Relief Valve Sections 2.2, 2.3 (Inside Enclosure) Sections 2.2, 2.3 Proportinal Valve for APU (Section 2.3) or Signal Disconnect Purge Control Module Relay for APU (Section 2.2) Section 2.3…

  • Page 9
    Trademarks All names identified by ® are registered trademarks of Siemens AG. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.
  • Page 10: Contacts

    General Information for the User 1.1 Contacts Contacts Register at the Siemens Industry Online Support (SIOS) website: https://support.industry.siemens.com International Siemens AG Siemens Industry, Inc. I IA SC PA PM Process Analytics 5980 West Sam Houston Parkway North Oestliche Rheinbrueckenstrasse 50…

  • Page 11: Approved Use

    Low voltages which are connected must also be generated using safe isolation. If any part of the Maxum II is opened, certain parts of the device are accessible which may carry dangerous voltages. Therefore, only suitably qualified personnel may work on this device as indicated in the next section which is titled «Qualified Personnel».

  • Page 12: Applicable Standards And Regulations

    The ATEX certificate for the Modular Oven Configuration is Sira 12ATEX1260X and comes with the following conditions for safe use: ● The Maxum II Modular Oven Gas Chromatograph shall be fitted with cable glands or conduit entries having an IP54 minimum rating.

  • Page 13: Safety Protection Principles

    The protection principles used depend on such factors as the type of device, the certifying organization that is applicable for the site, and the type of environment. Table 1-2 details some of the protection principles used in the Maxum II along with the hardware that uses those protection principles.

  • Page 14: Intrinsically Safe Devices

    IS‑TCD and the connections to the CIM touch screen display. Intrinsic safety protection for circuits and devices within the Maxum II enclosure is designed to conform to IEC EN60079-11. This is the standard that is applicable to factory installed electrical components and also for circuits that are additionally protected by other methods, such as a purged Maxum enclosure.

  • Page 15
    ● Maintenance for intrinsically safe devices, such as the IS-TCD, is limited to replacement. Repair or service on-site are not permitted. ● Intrinsically safe devices in the Maxum II that are equipped with a cover must have the cover attached during use. Leaving the cover off can violate requirements pertaining to mandatory separation from non-IS wiring.
  • Page 16
    General Information for the User 1.6 Intrinsically Safe Devices Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 17: Safety Systems — Purging

    The primary electronic devices of the Maxum II analyzer are housed inside the electronics enclosure (see Figures 1-1 and 1-2). All electronics in the Maxum II are designed to meet North American Division 2 and European Zone 2 safety requirements. This means that these electronics will not cause combustion of flammable vapors or gases under normal conditions.

  • Page 18: Electronics Purging System (Without Automatic Purge Unit)

    2.2.1 Overview The standard purging system of the Maxum II is controlled by the Power Entry Control Module (PECM) and the processor board (either SYSCON or CIM, depending on the Maxum II configuration). In the standard purging configuration, an alarm is generated whenever purging pressure is lost.

  • Page 19: Figure 2-2 Electronics Pressurization System (Type Py) For Airbath/Airless Oven Model (Old Pressure Relief Valve Configuration)

    Safety Systems — Purging 2.2 Electronics Purging System (without Automatic Purge Unit) Instrument Air System Controller Input — “Clean (SYSCON) Regulator and Dry” Electronics Purge Enclosure Purge Signal Warning Cable Light PECM to Purge SYSCON Power Entry Purge Control Control Module Pressure Sensor Module (PCM) (PECM)

  • Page 20: Instrument Air Inlet And Regulator

    Depending on the configuration of the Maxum II, the regulator may be labeled as either «Purge» or «Isothermal Oven Air» (refer to the section of this manual regarding Isothermal Ovens for more information). The purge air is connected to the purged electronics enclosure through two possible paths, the «normal»…

  • Page 21: Fast Purge Switch

    Safety Systems — Purging 2.2 Electronics Purging System (without Automatic Purge Unit) 2.2.3 Fast Purge Switch The fast purge control switch, which is shown in the following figure, is located inside the detector enclosure (middle door) or near the purge air regulator as shown in the figure below. This switch is operated by the user and is used to re-purge the electronics enclosure for a period of 8 minutes after the sealed enclosure has been opened and then re-sealed.

  • Page 22: Figure 2-5 Fast Purge Switch (Shown In The Off Position)

    Safety Systems — Purging 2.2 Electronics Purging System (without Automatic Purge Unit) Modular Oven Airbath/Airless Oven Modular Oven fast-purge switch label for Div 2 Modular Oven fast-purge switch label for Div 1 / Zone 1 Figure 2-5 Fast Purge Switch (Shown in the Off Position) Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…

  • Page 23: Purge Control Module

    (either SYSCON or CIM, depending on configuration). The SYSCON or CIM then generates an alarm and flashes the purge alarm LED on the front door of the Maxum II analyzer electronics enclosure. The purge alarm LED flashes red when an alarm is present.

  • Page 24: Pressure Sensor And Atmospheric Pressure Reference

    Control Module (PECM). PECM may refer to either the PECM or PECM-DC, depending on the Maxum II configuration. Located on this module is a pressure sensor (Figure 2-7) that detects the pressure inside the enclosure and compares it to the pressure outside the enclosure.

  • Page 25: Fast Purge Pressure Relief Valve

    The pressure relief valve, which is shown in the following figure, is located either inside the detector enclosure (middle door or a Maxum II with Airbath/Airless oven) or inside the top left of the electronics enclosure for a Maxum II Modular Oven. The location of the pressure relief valve within the analyzer is shown in Figures 2-1 and 2-2.

  • Page 26: Electronics Purging System (With Automatic Purge Unit)

    Power and all other external wiring that enters the Maxum II is disconnected by the APU whenever purge is lost. This is referred to as purging with Type X pressurization (px). The APU is available in Maxum II Airbath/Airless Oven and the Maxum II Modular Oven configurations.

  • Page 27: Purge Air Inlet And Regulator

    Safety Systems — Purging 2.3 Electronics Purging System (with Automatic Purge Unit) 2.3.2 Purge Air Inlet and Regulator This is a pressure regulator with air input that is connected to an instrument air source. The regulator is necessary to ensure that the input pressure does not exceed a certain value. Excessive pressure could damage other components in the purging system.

  • Page 28: Proportional Valve

    8 minutes before power is connected to remove potentially hazardous levels of flammable gases and vapors from the analyzer before power up. The APU is mounted inside the electronics enclosure of the Maxum II as shown in the following figure. Explosion Protection Safety Standards…

  • Page 29: Maintenance Switch

    Safety Systems — Purging 2.3 Electronics Purging System (with Automatic Purge Unit) Figure 2-14 Automatic Purge Control Unit (APU) shown in Airless/Airbath model 2.3.5 Maintenance Switch This device allows a user to override the power cutoff function of the APU for maintenance purposes.

  • Page 30: Exhaust Diffuser

    Safety Systems — Purging 2.3 Electronics Purging System (with Automatic Purge Unit) Figure 2-15 Location and Close-Up of Maintenance Switch (MAA, Set for Normal Operation) The maintenance switch for the MMO configuration is on the top left of the regulator panel, to the right of the electronics enclosure.

  • Page 31: Signal Disconnect Relay

    Safety Systems — Purging 2.3 Electronics Purging System (with Automatic Purge Unit) Figure 2-17 Covered Purge Exhaust Diffuser 2.3.7 Signal Disconnect Relay The Signal Disconnect Relay is mounted externally to the analyzer. This relay disconnects all external wiring to the analyzer except for power to the APU. The Signal Disconnect Relay is controlled by the APU which deactivates the relay when purge pressure is lost.

  • Page 32: Maintenance Considerations For Purged Systems

    Signal Disconnect Relays Maintenance Considerations for Purged Systems The following practices must be observed in order to operate, maintain and prevent compromise of the Maxum II purging system. WARNING Avoid Equipment Contamination or Ignition of Explosive Gases Failure to correctly observe the following practices may result in a dangerous situation that could lead to severe personal injury or death and/or extensive property damage.

  • Page 33
    A reference tube (see Figure 2-8) is used for this purpose. This tube must not be obstructed. Air does not flow through this tube. The tube is intended to sample the pressure immediately outside the Maxum II enclosure, and it should not be lengthened, shortened, or altered in any way.
  • Page 34
    Safety Systems — Purging 2.4 Maintenance Considerations for Purged Systems ● The Maxum II Analyzer should be visually inspected at regular intervals to ensure the integrity of the purge system. The following inspections should be made: – Instrument air lines and fittings should be inspected for leaks and damage.
  • Page 35: Safety Systems — Oven

    This is accomplished by the oven heater system. A variety of oven configurations are available for the Maxum II. For isothermal (single set temperature) analyses air bath, airless, and modular ovens are available. Since the Maxum II…

  • Page 36: Isothermal Air Bath Oven Heater System (Single And Split)

    Isothermal Air Bath Oven Heater System (Single and Split) The isothermal air bath oven is the most common oven used in the Maxum II. It is available in single and split configurations to accommodate a wide range of chromatographic applications.

  • Page 37: Temperature Control

    Safety Systems — Oven 3.2 Isothermal Air Bath Oven Heater System (Single and Split) Figure 3-2 Timing Coil for Pressure Switch See also Safety Systems — Purging (Page 17) 3.2.2 Temperature Control As mentioned in section 3.1, the temperature of the oven is controlled via an RTD sensor. The temperature of the air itself is controlled via two independent sensors which work with the two set point resistors mentioned in section 3.1 to ensure that all temperatures remain below possible ignition limits.

  • Page 38: Programmed Temperature Air Bath Oven Heater System (Split)

    Within the Maxum II, temperature control and air purging of the air bath heater is the same for the programmed temperature oven as for the isothermal air bath oven.

  • Page 39: Isothermal Airless Oven Heater System (Split)

    3.4 Isothermal Airless Oven Heater System (Split) Isothermal Airless Oven Heater System (Split) The Maxum II airless oven is used to provide a very stable isothermal oven temperature without the need to supply oven air to the analyzer. This allows for excellent chromatographic results while reducing utility costs.

  • Page 40
    Safety Systems — Oven 3.4 Isothermal Airless Oven Heater System (Split) Insert Location for RTD Probe (Same as on Opposite Side) Channel for Heating Element Figure 3-5 Location of Temperature Probes and Heaters on Airless Oven The heating elements are enclosed in hollow tubes running diagonally down the side of the aluminum casing.
  • Page 41: Maintenance Considerations For Air Bath And Airless Ovens

    (including programmed temperature oven) and airless ovens of Maxum II. Modular ovens are described in the next section. ● It must be ensured that the analyzer T-rating is suitable for the area in which the analyzer is to be installed.

  • Page 42: Modular Oven Heater System

    3.6 Modular Oven Heater System Modular Oven Heater System The Maxum II Modular Oven is used to provide a stable isothermal oven temperature for certain lower temperature applications. This allows for excellent chromatographic results at a greatly reduced product cost and simplified and low cost operation and maintenance for the applicable range of applications.

  • Page 43: Maintenance Considerations Modular Ovens

    Maintenance Considerations Modular Ovens The following practices must be observed in order to operate, maintain and prevent compromise of the Maxum II Modular Oven heater system. ● The installaion of the heater elements and RTDs should be inspected to verify that they are fully inserted into the tubes as shown in the previous section.

  • Page 44
    Safety Systems — Oven 3.7 Maintenance Considerations Modular Ovens Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 45: Valves, Detectors, And External Systems

    4.1.1 SLIV Functional Description Some applications require the use of a heated Siemens Liquid Injection Valve (SLIV) which is shown in figure 4-1. This valve is generally mounted through a side wall of the analyzer oven (refer to figure 4-2). The narrow portion of the valve that extends into the oven is known as the vaporizer.

  • Page 46: Sliv Maintenance Considerations

    4.1.2 SLIV Maintenance Considerations The Siemens Liquid Injection Valve (SLIV) should be inspected to assure that it is free of physical damage. All tubing and conduit connections should be correctly installed and tightened as prescribed by local safety requirements. Whenever maintenance is performed on the valve, care must be taken not to damage the temperature sensors and the heater element which are inserted in the valve body.

  • Page 47: Detectors

    Detector Functional Descriptions In order to analyze the chemical compositions of different samples, four major types of detectors are used in the Maxum II Gas Chromatograph (refer to figures 4-4 through 4-6). The types of detectors are: ● Explosion Proof/Flameproof Thermal Conductivity Detector (TCD) ●…

  • Page 48: Detector Maintenance Considerations

    For most detectors within the Maxum II, protection is accomplished by enclosing the active elements in a flameproof/explosion proof housing. In these cases electrical, optical and gas connections must penetrate the housing of the detector.

  • Page 49
    Valves, Detectors, and External Systems 4.2 Detectors The circuits for the IS-TCD require special considerations to ensure that the circuit remains intrinsically safe. For important information on the safe use of intrinsically safe circuits, refer to the sections of this manual labeled «Intrinsically Safe Devices». Figure 4-4 Explosion Proof/Flameproof Thermal Conductivity Detector Explosion Protection Safety Standards…
  • Page 50
    Valves, Detectors, and External Systems 4.2 Detectors Figure 4-5 Thermal Conductivity Detector Beads on Modular Oven Application Module (Intrinsically Safe TCD) Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 51
    Valves, Detectors, and External Systems 4.2 Detectors Figure 4-6 Flame Ionization Detector (without Insulation and Cover) Figure 4-7 Flame Photometric Detector (without Insulation and Cover) Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 52: Air Treater

    4.3.1 Air Treater Functional Description To provide clean air for the Flame Ionization Detector (FID), the Maxum II uses a catalytic device known as an Air Treater (figure 4-8) to remove hydrocarbons from ordinary plant instrument air. The Air Treater is an independent unit mounted near the Maxum II. It consists of an extremely hot combustion chamber and a catalyst.

  • Page 53: Air Treater Maintenance Considerations

    Two versions of the methanator exist for use in the Maxum. The original is protected by purging/ pressurization. The more recent version is protected in an explosion proof enclosure. Both versions are designed to install in the detector enclosure of the Maxum II Airbath/Airless Oven configuration (refer to figure 1-1).

  • Page 54
    When the purged/pressurized version of the methanator is installed, if the purge air supply of the Maxum II is interrupted during operation, it is imperative that the analyzer be powered down. Once powered down, the electronics enclosure door must be kept closed for at least 30 minutes.
  • Page 55: Methanator Maintenance Considerations

    Valves, Detectors, and External Systems 4.4 Methanator Figure 4-10 Explosion Proof Methanator (Both Versions) As protection measure, the gas inlets and outlets that enter and leave the methanator are configured as flame arrestors (refer to figure 4-8 and 4-9). In addition, both versions of the methanator assembly are insulated to prevent the surface temperature of the methanator from exceeding 180°C (surface temperatures for the explosion proof/flameproof methanator are significantly lower than 180°C).

  • Page 56
    Valves, Detectors, and External Systems 4.4 Methanator never be cut or shortened or bent or bypassed in any way. If there is any physical damage, the device should be shut down until it can be replaced. WARNING The purged/pressurized methanator must not be operated without purge/pressurization unless it has been verified that the area is free of explosive gases and vapors.
  • Page 57: Safe Operation

    PC-based System Manager software and also via a user access display panel that is mounted on the front door of the Maxum II electronics enclosure. Two types of display panel exist, both of which can be used to perform the same user functions. The newest version is the Color Touchscreen display, which is described in section 5-3.

  • Page 58: Maintenance Considerations For Maintenance Panel

    If an APU is not installed, the «Purge» LED flashes when purge pressure is lost. The «Power» LED is lit at all times when power is applied to the analyzer. Note that the «Fault» and «Warning» LEDs are related to the trouble reporting systems of the Maxum II but are not applicable to the safety systems.

  • Page 59: Maintenance Considerations For Control Interface Module Display

    Maxum II Use of the display for maintenance and operation functions is covered in the Reference Manual and the Maxum II Reference Manual, Modular Oven Configuration . Note In addition to the display installed on the analyzer, there is an emulator (called HMI emulator) available through the PC-based workstation software.

  • Page 60: Steps For Safe Startup Of Maxum Ii

    Overview The startup procedure described in this section explains the principle steps necessary to power the Maxum II in a manner that does not compromise the safety systems that have been described in this manual. This section does not explain the startup procedures required that relate to software functions or the analytical application of the analyzer.

  • Page 61
    Safe Operation 5.5 Steps for Safe Startup of Maxum II Note The remainder of the startup process is dependent on whether an APU is installed. (If APU is not equipped) Turn on the fast purge switch (figure 2-4) and allow fast purge air to flow for at least 8 minutes.
  • Page 62
    Safe Operation 5.5 Steps for Safe Startup of Maxum II Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…
  • Page 63: Glossary

    115 VAC or 230VAC and converts to 24V DC for distribution within the analyzer. Air Bath Oven One of three primary types of oven available for the Maxum II. Heating of the oven is accomplished by passing air over a heating element and into the oven. Available as a single oven or it can be split into two independent oven compartments.

  • Page 64
    Glossary Detector Compartment The middle section of a Maxum II Chromatograph that is equipped with either Airbath or Airless oven. It houses various detectors that may be equipped in the Maxum II Airbath/Airless Oven. It also houses the Methanator, if equipped.
  • Page 65
    CSA term for this concept is non-incendive. Oven Compartment The bottom area of a Maxum II that houses the ovens. Depending on the configuration of the Maxum II, this may be a Modular Oven Compartment or an Airbath/Airless Oven Compartment.
  • Page 66
    Temperature Programming The process of increasing oven temperature incrementally in order to reduce the time needed to separate a sample using a chromatograph. In the Maxum II, a variant of the air bath oven may be used to employ temperature programming.
  • Page 67
    A purged system where an alarm is generated whenever purge pressure is lost. In addition, any equipment which would exceed the T-Rating of the location upon loss of air flow (such as the Maxum II air bath oven heater) must be de-energized when purge is lost. Zone 1 An atmosphere where a mixture of air and flammable substances in the form of gas, vapor, or mist is likely to occur in normal operation occasionally.
  • Page 68
    Glossary Explosion Protection Safety Standards Manual, August 2017, A5E02220442001 Rev 8…

Назначение
Описание
Программное обеспечение
Технические характеристики
Знак утверждения типа
Комплектность
Поверка
Сведения о методах измерений

Назначение

Хроматографы газовые промышленные Maxum edition II предназначены для измерения содержания компонентов, входящих в состав анализируемых проб веществ и материалов.

Описание

Принцип действия хроматографов основан на разделении компонентов пробы при её прохождении в потоке газа-носителя через хроматографическую колонку и регистрации аналитического сигнала от компонента с помощью детектора.

Хроматографы представляют собой стационарные промышленные приборы и включают в себя электронный блок, один или два термостата с хроматографическими колонками и детекторами, блоки дозирования пробы и контроля газовых потоков, а также системы управления и обработки данных.

Для ввода пробы и переключения потоков между колонками, в зависимости от измерительной задачи, могут использоваться следующие устройства: мембранный кран, поворотный кран с пневмоприводом, пистонный кран жидкостного впрыскивания. В ряде случаев, для распределения потоков между колонками возможно использование бесклапанного переключения по патентованной LIVE технологии.

На передней панели хроматографов имеется сенсорный дисплей, предназначенный для управления прибором и отображения служебной и измерительной информации.

В хроматографах могут быть использованы как насадочные, так и капиллярные колонки.

В зависимости от решаемой аналитической задачи, хроматографы могут быть оснащены одним или несколькими детекторами из следующего списка:

•    ДТП -детектор по теплопроводности;

•    ПИД -пламенно-ионизационный детектор;

•    ПФД- пламенно-фотометрический детектор.

Хроматографы выпускаются в двух исполнениях, которые отличаются габаритными размерами и количеством модулей внутри корпуса. В исполнении 1 в хроматографе могут устанавливаться два типа термостатов: воздушный, термостатирование в котором осуществляется циркулированием горячего воздуха, либо один или два безвоздушных, термостатирование в котором происходит без циркуляции воздуха. В исполнении 2 в хроматограф устанавливаются один или два модуля, каждый из которых представляет собой смонтированный блок из мембранного крана переключения потоков и ввода пробы, хроматографических колонок и термокондуктометрического детектора.

При необходимости установки хроматографа во взрывоопасной зоне, хроматограф может выполняться в специальном взрывозащищенном исполнении с соответствующей маркировкой, указанной в таблице 1. Взрывозащита обеспечивается как продувкой блока электроники хроматографа воздухом КИП под избыточным давлением, так и взрывозащищенным исполнением отдельных узлов и компонентов (детекторов, нагревателей, электромагнитных клапанов и т.п.). Внешний вид хроматографов приведен на рисунках 1 и 2.

Исполнение

хроматографа

Ех-маркировка

Соответствуют требованиям стандартов

Maxum II Исполнение 1

1Ex d e py IIB+H2 T4…T1 Gb X или

1Ex d e px py IIB+H2 T4…T1 Gb X или

1Ex d e ib px py IIB+H2 T4.T1 Gb X или

1Ex d e m px IIB+H2 T4…T1 Gb X или

1Ex d e ib m px IIB+H2 T4.T1 Gb X

ГОСТ Р МЭК 60079-0-2011, ГОСТ IEC 60079-1-2011, ГОСТ IEC 60079-2-2011, ГОСТ Р МЭК 60079-7-2012, ГОСТ Р МЭК 60079-11-2010, ГОСТ Р МЭК 60079-18-2012

Maxum II Исполнение 2

1Ex py nA nC ib IIC T4 Gb X или

1Ex px ib IIC T4 Gb X

ГОСТ Р МЭК 60079-0-2011, ГОСТ IEC 60079-2-2011, ГОСТ Р МЭК 60079-11-2010, ГОСТ Р МЭК 60079-15-2010

Место нанесения знака поверки

Программное обеспечение

Хроматографы оснащены встроенным ПО, работающим под управлением операционной системой Windows CE, которое управляет работой прибора и отображает, обрабатывает и хранит полученные данные. Идентификационные данные ПО приведены в таблице 2.

Таблица 2

Идентификационные данные (признаки)

Значение

Идентификационное наименование ПО

САС_8И4_МАХиМ

Номер версии (идентификационный номер) ПО

не ниже 3.00.00.

Все ПО является метрологически значимым и выполняет следующие функции:

■    управление прибором;

■    настройка режимов работы прибора;

■    получение хроматограмм;

■    обработка и хранение результатов измерений

■    проведение калибровки прибора

■    проведение диагностических проверок прибора и отдельных его блоков; Уровень защиты ПО от непреднамеренных и преднамеренных изменений соответствует

уровню «средний» по Р 50.2.077-2014. Влияние ПО на метрологические характеристики учтено при их нормировании.

Технические характеристики

1 Предел детектирования и предельное допускаемое значение относительного изменения выходного сигнала за цикл измерений 48 часов (в зависимости от типа детектора):

Детек

тор

Предел детектирования, не более

Контрольное вещество

Предельное допускаемое значение относительного изменения выходного сигнала за цикл измерений 48 часов (по площади пика), %

ДТП

0,5 • 10-10 г/см3

Бензол1

±2,0

Пропан

ПИД

0,5 10-13 г/с

Бензол1

±2,0

110-12 г/с

Пропан

ПФД

3 10-12 г/c (по сере)

Сера меркаптановая1

±4,0

Сероводород

2 Относительное СКО выходного сигнала (в зависимости от детектора), %, не более:

Детектор

По времени удерживания

По высоте пика

По площади пика

ДТП

1,0

2,0

2,0

ПИД

1,0

2,0

2,0

ПФД

2,0

4,0

4,0

3.    Напряжение сетевого питания частотой 50±1 Гц, В    220 22-зз

4.    Потребляемая мощность, В А, не более:    2000

5.    Средний срок службы, лет, не менее    8

6. Наработка на отказ, ч, не менее

7. Габаритные размеры (ДхШхВ), мм, не более: -исполнение 1 -исполнение 2

16000

662x411x1010

662x411x683

8. Масса, кг, не менее: -исполнение 1 -исполнение 2

77

60

9. Условия эксплуатации:

о/~’

от -18 до +50 80

от 84 до 106

-диапазон температур окружающего воздуха, С -диапазон относительной влажности окружающего воздуха (при 25 оС), %, не более -диапазон атмосферного давления, кПа

Знак утверждения типа

наносится на титульный лист руководства по эксплуатации методом компьютерной графики и на левую боковую панель корпуса хроматографа в виде наклейки.

Комплектность

определяется заказом и отражается в спецификации.

Основной комплект включает:

—    хроматограф;

—    руководство по эксплуатации;

—    методику поверки МП-242-1896-2015.

Поверка

осуществляется по документу МП-242-1896-2015 «Хроматографы газовые промышленные Maxum edition II. Методика поверки», утвержденному ГЦИ СИ ФГУП «ВНИИМ им. Д.И.Менделеева» 05.07.2015 года.

Основные средства поверки: стандартный образец массовой и объемной доли бензола в нефтепродуктах ГСО 10185-2013, стандартный образец содержания меркаптановой серы в нефтепродуктах ГСО 8418-2003. Стандартные образцы — поверочные газовые смеси: 103222013 (пропан/гелий), ГСО 10538-2014 (сероводород/ метилмеркаптан/ этилмеркаптан/азот) или ГСО 8529-2004 (сероводород/азот) или аналогичные по составу и метрологическим характеристикам ГСО.

Сведения о методах измерений

приведены в документе «Хроматографы газовые промышленные Maxum edition II. Руководство по эксплуатации»

Нормативные и технические документы, устанавливающие требования к хроматографам газовым промышленным Maxum edition II

Техническая документация фирмы-изготовителя.

Siemens Maxum II: Available Instructions

Note for Owners:

Guidesimo.com webproject is not a service center of Siemens trademark and does not carries out works for diagnosis and repair of faulty Siemens Maxum II equipment. For quality services, please contact an official service center of Siemens company. On our website you can read and download documentation for your Siemens Maxum II device for free and familiarize yourself with the technical specifications of device.

  • Agilent Technologies Varian CP-8400

    Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Varian, Inc. 2003 Printed in the Netherlands 03-914852-00:Rev. 4.1 CP-8400 AutoSampler and CP-8410 AutoInjector Operator’s Manual …

    Varian CP-8400 Laboratory Equipment, 62

  • Leica CM3050 S

    Leica CM3050SCryostat Instructions for UseEnglishOrder No.: 14044380101 — Revision GAlways keep this manual with the instrument.Read carefully before working with the instrument.Version 2.0, Revision G — 11/2017 …

    CM3050 S Laboratory Equipment, 86

  • Munchkin Steam Guard

    CCD-0053-001 REV 2IF THE SUPPLY CORD IS DAMAGED, IT MUST BE REPLACED BY A SPECIAL CORD OR ASSEMBLY AVAILABLE FROM THE MANUFACTURER OR ITS SERVICE AGENT.SI LE CORDON D’ALIMENTATION EST ENDOMMAGÉ, IL DOIT ÊTRE REPLACÉ PAR UN CORDON SPÉCIAL POUVANT ÊTRE OBTENU AUPRÈS DU FABRICANT OU D’UN RÉPARATEUR AGRÉÉ. …

    Steam Guard Laboratory Equipment, 8

  • Velp Scientifica F201A0500

    1 Instruction Manual Manuale di istruzioni Manuel d’instructions Manual de instrucciones Bedienungsanleitung CONTROLLER Advance F201A0500 General Information / Informazioni Generali / Informations Générales / Información General / Allgemeine Hinweise Before using the unit, please read the following instruction manual carefully. Prima dell’utilizzo dello strumento si raccomanda d …

    F201A0500 Laboratory Equipment, 36

  • REITEL RETOCAST T

    Operating Instructions LABORATORY EQUIPMENTIN STAINLESS STEEL RETOCAST T REITEL Feinwerktechnik GmbH Senfdamm 20 — 49152 Bad Essen Phone: +49(0)5472-9432-0 — Fax: +49(0)5472-9432-40 http://www.reitel.com — [email protected] …

    RETOCAST T Laboratory Equipment, 28

  • Bio-Works BabyBio S

    IN 45 100 010 • AA • 1 / 12 INSTRUCTION IN 45 100 010 BabyBio S BabyBio Q BabyBio DEAE BabyBio™ S, BabyBio Q and BabyBio DEAE are ready-to-use ion exchange chromatography columns for easy and convenient purification of proteins, peptides and oligonucleotides, by utilizing the difference in these molecules surface charge. BabyBio S works as a strong cation exchanger, BabyBio Q a …

    BabyBio S Laboratory Equipment, 12

  • Drinkpod 1000 Pro Series

    Use & Care GuideEnglish 1000 Pro Series Countertop Purier·Dispenser·SterilizerModel: DP1000(B)Customer Assistance 1-844-374-6576www.drinkpod.comDrinkpod and the Drinkpod logo are the registeredtrademarks of Drinkpod, LLC. All Rights Reserved. …

    1000 Pro Series Laboratory Equipment, 27

  • SciCan statim 5000 G4

    95-112505 Rev. 4.0 Copyright 2011 SciCan Ltd . All rights reserved.STATim 2000/5000 G4CASSETTE AUTOCLAVE™•  Setting Up and Using Your Web Portal•  Paramétrage et utilisation de votre portail Web•  Einrichtung und Verwendung Ihres Web-Portals•  Impostazione e uso del Vostro portale web•  Conguración y uso de su portal …

    statim 5000 G4 Laboratory Equipment, 56

Popular Laboratory Equipment User Guides:

Table of Contents for Siemens Maxum II:

  • Compare Airflow Loss Shutdown PECM Air Bath Oven High-Wattage Heater Analog SSR Pair Air Pressure Switch Digital Temp Control, Temp Limit, Overtemp RTDs Temp Setpoint Modules PWM I 2 C Temperature Control Airflow Loss Shutdown Heater AC Line SSRa SSRb SYSCON Figure 2-9 PECM Heater Control Functions Each circuit consists of two series-connected solid-state relays. One of these relays controls the 1400-Watt AC heater to maintain the set point

  • 2.10.3 Replacement Procedures Overview Replacing the color touchscreen panel requires removal and disassembly of the door. WARNING Follow Safety Precautions. Failure to follow proper procedures may result in equipment damage, personal injury or death. Full safety precautions must be followed throughout all sections of this procedure to prevent possible injury, equipment damage, or death. Verify that the area is clear of flammable gases and vapors and that appropriate authorization is obtained to do the

  • 11.Ensure the correct fuses are in the correct positions, as shown in PECM AC Power Distribution illustration. 12.When the procedure is completed, follow the steps in the General Analyzer Startup Procedure . 2.3 System Controller Version 2.1 (SYSCON2.1) 2.3.1 Description The System Controller (SYSCON2.1) is a combination of two interconnected boards that together function as the control processor and motherboard for the Maxum analyzer. The SYSCON2.1 consists of two boards, the Communication and Analytical Control

  • 2.3.3.2 CAC3 Status Indicator LEDs The CAC3 is equipped with several LEDs that communicate useful information about the operating status of the CAC3. Link Acknowledge LED4 LED5 LED7 LED1 LED2 LED3 Link Status Figure 2-13 CAC3 LED Locations LED1 Debug LED1 Green – On during normal operation. LED2 Debug LED2 Green – On during normal operation. Off during bootload. LED3 Power Good Green – Power to CAC3 is functional.

  • 2.7 Solid State Relay Module The Solid State Relay (SSR) Module is made up of two pairs of high wattage heater relays that are used for controlling the oven air bath heaters. One pair controls ABH1 and the other ABH2. Each pair of relays controls Temperature Limit and Oven Temperature shut down. If the over temperature limit is exceeded, the power to the air bath heater is shutdown. Two different configurations of SSR are available, the SSR and the

  • 2.3.4 Maintenance Overview Board Replacement A failure in the SYSCON2.1 will generally interrupt communication between the GCP networkbased software and the detectors. A simplified view of the system is shown below. CAC3 CAN RS485 RS232 Digital Communication DC Supply AC Supply Analyzers GCP Ethernet Ethernet Switch (optional) I/O Boards (I 2 C) 120/240VAC 120/240VAC 120/240VAC 24VDC Heaters 24 V Supply PECM SSR Board 24VDC 24VDC I

  • Vapor or High Pres‐ sure Liquid Samples Model 20 The air-pressure actuated, diaphragm valve provides uniform sample volume, low internal volume, high pressure up to 1500 psi, 10350 kPa, fast switching (millisec‐ onds), reliability, and durability. It functions equally well as a liquid or vapor sample valve, column switching valve, or column back flush valve. Liquid Sample LIV The liquid injection valve can be used to automatically inject a constant quantity of liquid sample followed by fast, complete vaporization. Small gas quantities can also be injec

  • Note Verify proper position of line-voltage selector switch and fuse value. Incorrect settings can damage the equipment. See the information packet that was shipped with the analyzer for information on the individual analyzer. Electronic Compartment Component Descriptions and Maintenance Procedures 2.1 Power Supplies Maxum II Reference Manual Manual, 7/2017, 2000596-001 17

  • 8. Lift the unit up and out of the enclosure and place on an ESD-safe work surface. 9. Remove the two mounting screws near the upper corners and remove the old unit from the DPM cage. Configuring the New IS-TCD3 DPM Set the location ID switch and reference-selector switches on the replacement unit to match those on the unit being removed. Note The newer units use slide switches in place of the 2-pos

  • Overview of DPM Functions The Base 3 Detector Personality Module (DPM) combines these functions in a single module: Including Mezzanine Modules FID Input from detector via mezzanine module Ignite signal / glow-plug output Range-select output 300-V bias output Flame-sense input (used in Maxum I analyzers only) FPD Input from detector via mezzanine module Ignite signal / glow-plug output Range-select output Enable signal output 300V bias output Flame-sense input (used in Maxum I analyzers

  • Digital inputs ● Optically coupled with a common for all inputs. ● Self powered floating-contact input, or configurable for sinking or sourcing current. ● Sourcing current mode: 24 V internal isolated supply, with positive terminal of supply at common. ● Sinking current mode: 5 V internal isolated supply, with negative terminal of supply at common. Digital outputs Floating double-throw contacts, maximum contact load rating 1 A at 30 V (AC or DC). External diode shunt suppression should be used for induct

  • 2.4.4 Analog and Digital IO Board………………………………………………………………………………………60 2.5 Detector Personality Modules……………………………………………………………………………………61 2.5.1 DPM Types……………………………………………………………………………………………………..

  • Electronic Compartment Component Descriptions and Maintenance Procedures 2.10 Color Touchscreen Maxum II Reference Manual 92 Manual, 7/2017, 2000596-001

  • 2.4.1 IO Card Common Features Available IO Boards The newest version of IO board communicates over the I 2 C bus. The I 2 C IO boards are the type available for new installation. ● Analog IO board (AIO_I2C, Part Number A5E02486267001): has 8 analog output channels, 8 analog input channels, and 2 digital input channels ● Digital IO board (DIO_I2C, Part Number A5E02486268001): has 8 digital outputs and 6 digital inputs ● Analog and Digital IO board (ADIO_I2C, Part Number A5E02359491001): has 4 digital outputs

  • Install wire tie for the LCD connector through hole adjacent to the cutout. Install wire tie on the lower TIB standoff near the intrinsic safety ground lugs. Install wire tie for backlight connector through small hole at the corner of the panel plate. Figure 2-49 Wire Tie Installation Detail Electronic Compartment Component Descriptions and Maintenance Procedures 2.10 Color Touchscreen Maxum II Refe

  • Analyzer Overview 1 1.1 Introduction The Maxum edition II system, also called the “Maxum II”, represents a significant advance in process chromatography. The Maxum II combines the best of the Siemens Advance Maxum and PGC 302 gas chromatographs into a single platform analyzer. From oven and electronic components to software and communication networks, the system is modular. Pre-configured application modules are available for many common measurements. A Maxum II system offers a wide range of detector modules including Thermal Conductivity,

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of Siemens Maxum II device using right now.

Download Manual  of Siemens Maxum II Laboratory Equipment, Measuring Instruments for Free or View it Online on All-Guides.com.

Siemens Maxum II Manual  - Page 1

1

Siemens Maxum II Manual  - Page 2

2

Siemens Maxum II Manual  - Page 3

3

Siemens Maxum II Manual  - Page 4

4

Siemens Maxum II Manual  - Page 5

5

Siemens Maxum II Manual  - Page 6

6

Siemens Maxum II Manual  - Page 7

7

Siemens Maxum II Manual  - Page 8

8

Siemens Maxum II Manual  - Page 9

9

Siemens Maxum II Manual  - Page 10

10

Siemens Maxum II Manual  - Page 11

11

Siemens Maxum II Manual  - Page 12

12

Siemens Maxum II Manual  - Page 13

13

Siemens Maxum II Manual  - Page 14

14

Siemens Maxum II Manual  - Page 15

15

Siemens Maxum II Manual  - Page 16

16

Siemens Maxum II Manual  - Page 17

17

Siemens Maxum II Manual  - Page 18

18

Siemens Maxum II Manual  - Page 19

19

Siemens Maxum II Manual  - Page 20

20

Siemens Maxum II Manual  - Page 21

21

Siemens Maxum II Manual  - Page 22

22

Siemens Maxum II Manual  - Page 23

23

Siemens Maxum II Manual  - Page 24

24

Siemens Maxum II Manual  - Page 25

25

Siemens Maxum II Manual  - Page 26

26

Siemens Maxum II Manual  - Page 27

27

Siemens Maxum II Manual  - Page 28

28

Siemens Maxum II Manual  - Page 29

29

Siemens Maxum II Manual  - Page 30

30

Siemens Maxum II Manual  - Page 31

31

Siemens Maxum II Manual  - Page 32

32

Siemens Maxum II Manual  - Page 33

33

Siemens Maxum II Manual  - Page 34

34

Siemens Maxum II Manual  - Page 35

35

Siemens Maxum II Manual  - Page 36

36

Siemens Maxum II Manual  - Page 37

37

Siemens Maxum II Manual  - Page 38

38

Siemens Maxum II Manual  - Page 39

39

Siemens Maxum II Manual  - Page 40

40

Siemens Maxum II Manual  - Page 41

41

Siemens Maxum II Manual  - Page 42

42

Siemens Maxum II Manual  - Page 43

43

Siemens Maxum II Manual  - Page 44

44

Siemens Maxum II Manual  - Page 45

45

Siemens Maxum II Manual  - Page 46

46

Siemens Maxum II Manual  - Page 47

47

Siemens Maxum II Manual  - Page 48

48

Siemens Maxum II Manual  - Page 49

49

Siemens Maxum II Manual  - Page 50

50

Siemens Maxum II Manual  - Page 51

51

Siemens Maxum II Manual  - Page 52

52

Siemens Maxum II Manual  - Page 53

53

Siemens Maxum II Manual  - Page 54

54

Siemens Maxum II Manual  - Page 55

55

Siemens Maxum II Manual  - Page 56

56

Siemens Maxum II Manual  - Page 57

57

Siemens Maxum II Manual  - Page 58

58

Siemens Maxum II Manual  - Page 59

59

Siemens Maxum II Manual  - Page 60

60

Siemens Maxum II Manual  - Page 61

61

Siemens Maxum II Manual  - Page 62

62

Siemens Maxum II Manual  - Page 63

63

Siemens Maxum II Manual  - Page 64

64

Siemens Maxum II Manual  - Page 65

65

Siemens Maxum II Manual  - Page 66

66

Siemens Maxum II Manual  - Page 67

67

Siemens Maxum II Manual  - Page 68

68

Siemens Maxum II Manual  - Page 69

69

Siemens Maxum II Manual  - Page 70

70

Siemens Maxum II Manual  - Page 71

71

Siemens Maxum II Manual  - Page 72

72

Siemens Maxum II Manual  - Page 73

73

Siemens Maxum II Manual  - Page 74

74

Siemens Maxum II Manual  - Page 75

75

Siemens Maxum II Manual  - Page 76

76

Siemens Maxum II Manual  - Page 77

77

Siemens Maxum II Manual  - Page 78

78

Siemens Maxum II Manual  - Page 79

79

Siemens Maxum II Manual  - Page 80

80

Siemens Maxum II Manual  - Page 81

81

Siemens Maxum II Manual  - Page 82

82

Siemens Maxum II Manual  - Page 83

83

Siemens Maxum II Manual  - Page 84

84

Siemens Maxum II Manual  - Page 85

85

Siemens Maxum II Manual  - Page 86

86

PD PA AP

Maxum edition II Analyzer General

Maintenance

Manual

General Analyzer Functions, Troubleshooting, and

Maintenance for Maxum II Gas Chromatographs. The

information in this manual supercedes the applicable

topics in previous manuals.

August 2018

A5E42019842001

Introduction

1

Safety notes

2

System Functions

3

General Maintenance and

Troubleshooting

4

Appendix A — Contact

Information

A

Понравилась статья? Поделить с друзьями:
  • Национальное руководство по эндокринологии 2020 скачать
  • Кондиционер electrolux инструкция к пульту теплый воздух
  • Руководство дежурством в уголке природы
  • Вебасто thermo top c дизель инструкция по эксплуатации
  • Как вести журнал учета выдачи инструкций по охране труда