Руководство по эксплуатации beneheart

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

BeneHeart C & BeneHeart S Series

Automatic External Defibrillator

Operator’s Manual

(BeneHeart C1/BeneHeart C1A/BeneHeart C2/BeneHeart C2A/

BeneHeart C1 Fully Automatic/BeneHeart C1A Fully Automatic/

BeneHeart C2 Fully Automatic/BeneHeart C2A Fully Automatic/

BeneHeart S1/BeneHeart S1A/BeneHeart S2/BeneHeartS2A/

BeneHeart S1 Fully Automatic/BeneHeart S1A Fully Automatic/

BeneHeart S2 Fully Automatic/BeneHeart S2A Fully Automatic)

loading

Summary of Contents for Mindray BeneHeart C Series

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    1/168

    BeneHeart D6

    Defibrillator/Monitor

    Service Manual

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    2/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    3/168

    I

    Intellectual Property Statement

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter
    called

    Mindray) owns the intellectual property rights to this product
    and this manual. This manual

    may refer to information protected by copyrights or patents and
    does not convey any license

    under the patent rights of Mindray, nor the rights of others.
    Mindray does not assume any

    liability arising out of any infringements of patents or other
    rights of third parties.

    , , and are the registered trademarks or

    trademarks owned by Mindray in China and other countries.

    Revision History

    This manual has a revision number. This revision number changes
    whenever the manual is

    updated due to software or technical specification change.
    Contents of this manual are subject

    to change without prior notice.

      Version number 7.0

      Release time: May 2011

    © 2009-2011 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
    All rights reserved.

    Company Contact

    Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co.,
    Ltd.

    E-mail Address: [email protected]

    Tel: +86 755 26582479, +86 755 26582888

    Fax: +86 755 26582934, +86 755 26582500

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    4/168

    II

    Preface

    Manual Purpose

    This manual provides detailed information about the assembling,
    dissembling, testing and

    troubleshooting of the equipment to support effective
    troubleshooting and repair. It is not

    intended to be a comprehensive, in-depth explanation of the
    product architecture or technical

    implementation. Observance of the manual is a prerequisite for
    proper equipment

    maintenance and prevents equipment damage and personnel
    injury.

    Intended Audience

    This manual is for biomedical engineers, authorized technicians
    or service representatives

    responsible for troubleshooting, repairing and maintaining the
    defibrillator/ monitors

    Passwords

    Passwords may be required to access different modes. The
    passwords are listed below:

      Installation mode: 888888

    Service mode: 332888

      Configuration mode: 315666

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    5/168

    1

    Contents

    1
    Safety…………………………………………………………………………………………………………………
    1-1 

    1.1 Safety Information
    …………………………………………………………………………………………….
    1-1

    1.1.1 Dangers
    ………………………………………………………………………………………………..
    1-2

    1.1.2
    Warnings……………………………………………………………………………………………….
    1-2

    1.1.3 Cautions
    ……………………………………………………………………………………………….
    1-2

    1.1.4 Notes
    ……………………………………………………………………………………………………
    1-3

    1.2 Equipment Symbols
    …………………………………………………………………………………………..
    1-3

    2 Theory of Operation
    …………………………………………………………………………………………..
    2-1 

    2.1 The Basics
    ………………………………………………………………………………………………………..
    2-1

    2.1.1
    Overview………………………………………………………………………………………………
    2-1

    2.1.2 Main
    Functions………………………………………………………………………………………
    2-1

    2.2 Components
    ……………………………………………………………………………………………………..
    2-2

    2.3 Main Unit
    …………………………………………………………………………………………………………
    2-2

    2.4 Front Housing
    Assembly…………………………………………………………………………………….
    2-4

    2.5 Paddle Tray
    ………………………………………………………………………………………………………
    2-6

    2.6 Rear Housing
    Assembly……………………………………………………………………………………..
    2-6

    2.6.1 Power
    System………………………………………………………………………………………..
    2-6

    2.6.2 Main Control System
    ……………………………………………………………………………..
    2-7

    2.6.3 Therapy
    System……………………………………………………………………………………..
    2-7

    2.6.4 Parameter Measurement System
    ………………………………………………………………
    2-7

    2.7 External Device
    Connectors………………………………………………………………………………..
    2-8

    3 Unpacking and Installation
    …………………………………………………………………………………
    3-1 

    3.1 Unpacking the
    Equipment…………………………………………………………………………………..
    3-1

    3.2 Preparation for
    Installation………………………………………………………………………………….
    3-2

    3.2.1 Preparation for Installation
    Site………………………………………………………………..
    3-2

    3.2.2 Electrical Requirements
    ………………………………………………………………………….
    3-3

    3.3 Vehicle Mount Kit Installation
    …………………………………………………………………………….
    3-3

    3.4 Installing Hook Kit or Conductive Gel Container
    Kit…………………………………………….. 3-3

    3.5 Preparation for Power
    On……………………………………………………………………………………
    3-3

    3.6 User
    Test…………………………………………………………………………………………………………..
    3-4

    4 Testing and
    Maintenance…………………………………………………………………………………….
    4-1 

    4.1
    Introduction………………………………………………………………………………………………………
    4-1

    4.1.1 Test Report
    ……………………………………………………………………………………………
    4-2

    4.1.2 Preventative Maintenance
    ……………………………………………………………………….
    4-2

    4.1.3 Recommended
    Frequency……………………………………………………………………….
    4-3

    4.2 Preventative Maintenance Procedures
    ………………………………………………………………….
    4-4

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    6/168

    2

    4.2.1 Visual Test
    …………………………………………………………………………………………….
    4-4

    4.2.2 NIBP
    Tests…………………………………………………………………………………………….
    4-4

    4.2.3 CO2 Module
    Tests…………………………………………………………………………………..
    4-7

    4.2.4 Preventative maintenance test
    report…………………………………………………………
    4-8

    4.3 Power On Test
    …………………………………………………………………………………………………..
    4-9

    4.4 User
    Test…………………………………………………………………………………………………………
    4-10

    4.5 Password for Installation Mode
    ………………………………………………………………………….4-11

    4.6 Module Performance
    Tests…………………………………………………………………………………4-11

    4.6.1 Manual Defibrillation Test
    ……………………………………………………………………..4-11

    4.6.2 Pacing Test
    ………………………………………………………………………………………….
    4-13

    4.6.3 ECG Test
    …………………………………………………………………………………………….
    4-14

    4.6.4 Resp Test
    …………………………………………………………………………………………….
    4-15

    4.6.5 IBP
    Tests……………………………………………………………………………………………..
    4-15

    4.6.6
    SpO2 Test…………………………………………………………………………………………….
    4-17

    4.6.7 Temp Test
    ……………………………………………………………………………………………
    4-17

    4.7 Analogue Output
    Test……………………………………………………………………………………….
    4-18

    4.8 Electrical Safety Tests
    ………………………………………………………………………………………
    4-18

    4.9 Recorder
    Check……………………………………………………………………………………………….
    4-18

    4.10 Factory Service
    ……………………………………………………………………………………………..
    4-18

    4.10.1 Password for Service
    Mode………………………………………………………………….
    4-18

    4.10.2 Accessing Service Mode Menu
    …………………………………………………………….
    4-19

    4.10.3 Calibrating
    NIBP………………………………………………………………………………..
    4-19

    4.10.4 Calibrating/Zeroing
    Impedance…………………………………………………………….
    4-19

    4.10.5 Device Information
    …………………………………………………………………………….
    4-21

    4.10.6 Checking Failure
    Code………………………………………………………………………..
    4-21

    4.10.7 Inputting Serial
    Number………………………………………………………………………
    4-22

    4.10.8 Paddle Open Circuit
    Display………………………………………………………………..
    4-22

    4.10.9 Selecting Recorder Type
    ……………………………………………………………………..
    4-22

    5 Hardware and Software
    Upgrade………………………………………………………………………..
    5-1 

    5.1 Hardware Upgrade
    …………………………………………………………………………………………….
    5-1

    5.1.1 Upgrading MPM module
    ………………………………………………………………………..
    5-1

    5.1.2 MPM Module upgrade procedure
    …………………………………………………………….
    5-5

    5.1.3 Upgrade the Therapy
    Module…………………………………………………………………..
    5-7

    5.1.4 Upgrade the Equipment with CO2
    Module………………………………………………..
    5-8

    5.1.5 Upgrade the Recorder
    …………………………………………………………………………..
    5-10

    5.2 Software Upgrade through a
    PC…………………………………………………………………………
    5-10

    5.2.1 Installing Mindray Patient Monitor Software Upgrade
    Tool………………………..5-11

    5.2.2 Software Upgrade Procedure
    …………………………………………………………………
    5-13

    5.3 Software Upgrade through a USB Memory
    …………………………………………………………
    5-14

    5.3.1
    Precautions………………………………………………………………………………………….
    5-14

    5.3.2 Software Upgrade Procedure
    …………………………………………………………………
    5-14

    6
    Troubleshooting………………………………………………………………………………………………….
    6-1 

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    7/168

    3

    6.1
    Overview………………………………………………………………………………………………………….
    6-1

    6.2 Part Replacement
    ………………………………………………………………………………………………
    6-1

    6.3 Checking Defibrillator/Monitor
    Status………………………………………………………………….
    6-1

    6.4 Checking Device
    Information……………………………………………………………………………..
    6-2

    6.5 Checking Technical
    Alarm………………………………………………………………………………….
    6-2

    6.6 Troubleshooting
    Guide……………………………………………………………………………………….
    6-3

    6.6.1 Defibrillation
    Problems…………………………………………………………………………..
    6-3

    6.6.2 Pacing
    Problems…………………………………………………………………………………….
    6-5

    6.6.3 Power On/Off
    Problems………………………………………………………………………….
    6-5

    6.6.4 Display Problems
    …………………………………………………………………………………..
    6-6

    6.6.5 Alarm
    Problems……………………………………………………………………………………..
    6-7

    6.6.6 Button and Knob Problems
    ……………………………………………………………………..
    6-8

    6.6.7 Recorder
    Problems…………………………………………………………………………………
    6-9

    6.6.8 Output Interface Problems
    ………………………………………………………………………
    6-9

    6.6.9 CF Card
    Problems………………………………………………………………………………..
    6-10

    6.6.10 Wireless Transmission Module Problems
    ……………………………………………… 6-10

    6.6.11 Power Supply Problems
    ……………………………………………………………………….6-11

    6.6.12 Software Upgrade
    Problems…………………………………………………………………
    6-12

    6.7 Technical Alarm
    Messages………………………………………………………………………………..
    6-12

    6.8 Error Codes
    …………………………………………………………………………………………………….
    6-13

    6.8.1 Therapy Module Error
    Codes…………………………………………………………………
    6-13

    6.8.2 Power Module Error
    Codes……………………………………………………………………
    6-14

    6.8.3 Main Control Error
    Codes……………………………………………………………………..
    6-15

    6.8.4 MPM Error Codes
    ………………………………………………………………………………..
    6-15

    7 Disassembly and Repair
    ……………………………………………………………………………………..
    7-1 

    7.1 Tools Required
    ………………………………………………………………………………………………….
    7-1

    7.2 Preparations for
    Disassembly………………………………………………………………………………
    7-2

    7.3 Disassembling the Main
    Unit………………………………………………………………………………
    7-3

    7.3.1 Removing Hook
    Mount…………………………………………………………………………..
    7-3

    7.3.2 Removing Paddle
    Tray……………………………………………………………………………
    7-4

    7.3.3 Separating the Housing
    …………………………………………………………………………..
    7-5

    7.3.4 Removing the Measurement Module Panel
    ………………………………………………. 7-6

    7.3.5 Removing the Power Supply Assembly
    …………………………………………………….
    7-8

    7.3.6 Discharging the
    Capacitor……………………………………………………………………….
    7-9

    7.3.7 Removing the Therapy Module High-voltage
    Board………………………………… 7-10

    7.3.8 Disassembling the MPM
    Module…………………………………………………………….7-11

    7.3.9 Removing the CO2 Module
    …………………………………………………………………..
    7-12

    7.3.10 Removing the CPU board
    Assembly……………………………………………………..
    7-12

    7.3.11 Removing the Therapy Port
    Assembly…………………………………………………..
    7-13

    7.3.12 Checking Waterproof Material on the Rear Housing
    ………………………………. 7-14

    7.4 Disassembling the Front Housing
    Assembly………………………………………………………..
    7-15

    7.4.1 Removing the Keypad
    board………………………………………………………………….
    7-15

    7.4.2 Removing Display
    Assembly…………………………………………………………………
    7-17

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    8/168

    4

    7.4.3 Removing the Speaker
    ………………………………………………………………………….
    7-20

    7.4.4 Removing the Indicating Lamp Board and Alarm Lamp Board
    …………………. 7-20

    7.4.5 Removing the Mode Select Knob
    …………………………………………………………..
    7-21

    7.4.6 Removing the Rotary Encoder
    ……………………………………………………………….
    7-22

    7.4.7 Removing the
    Recorder…………………………………………………………………………
    7-22

    7.5 Disassembling the MPM
    Module……………………………………………………………………….
    7-23

    7.5.1 Removing the Fan
    ………………………………………………………………………………..
    7-23

    7.5.2 Removing the SpO2 board
    …………………………………………………………………….
    7-24

    7.5.3 Removing the MPM Module Analog Board
    ……………………………………………. 7-25

    7.5.4 Removing the MPM Module Digital
    Board…………………………………………….. 7-26

    7.5.5 Removing the NIBP
    Module………………………………………………………………….
    7-27

    7.6 Disassembling the Power Supply Assembly
    ………………………………………………………..
    7-28

    7.6.1 Removing the AC/DC board
    ………………………………………………………………….
    7-28

    7.6.2 Removing the Power Supply Sheet Metal and the Grounding
    Terminal………. 7-29

    7.7 Disassembling the CO2 Module
    ………………………………………………………………………..
    7-30

    7.7.1 Disassembling the Microstream CO2 Module
    …………………………………………. 7-30

    7.7.2 Disassembling Mindray CO2
    Module……………………………………………………..
    7-31

    7.8 Disassembling the Measurement Module Panel
    Assembly……………………………………. 7-32

    7.8.1 Disassembling the Measurement Module Panel with Mindray
    CO2 Module.. 7-32

    7.8.2 Disassembling the Measurement Module Panel with
    Microstream CO2 Module

    …………………………………………………………………………………………………………………..
    7-33

    7.8.3 Disassembling the Measurement Module Panel without CO2
    Module ……….. 7-33

    7.9 Disassembling the
    Recorder………………………………………………………………………………
    7-34

    7.9.1 Disassembling the TR6F
    Recorder………………………………………………………….
    7-34

    7.9.2 Disassembling the TR8A Recorder
    …………………………………………………………
    7-35

    7.10 Disassembling the External
    Paddle…………………………………………………………………..
    7-37

    7.10.1 Disassembling the Adult
    Paddle……………………………………………………………
    7-37

    7.10.2 Disassembling the Apex Pediatric
    Paddle………………………………………………
    7-37

    7.10.3 Disassembling the Sternum Pediatric
    Paddle…………………………………………. 7-38

    8 Parts
    ………………………………………………………………………………………………………………….
    8-1 

    8.1
    Introduction………………………………………………………………………………………………………
    8-1

    8.2 Main Unit
    …………………………………………………………………………………………………………
    8-2

    8.2.1 Exploded View
    ………………………………………………………………………………………
    8-2

    8.2.2 Parts List
    ………………………………………………………………………………………………
    8-2

    8.3 Front Housing
    Assembly…………………………………………………………………………………….
    8-4

    8.3.1 Exploded View
    ………………………………………………………………………………………
    8-4

    8.3.2 Parts List
    ………………………………………………………………………………………………
    8-4

    8.4 Rear Housing
    Assembly……………………………………………………………………………………..
    8-8

    8.4.1 Exploded View
    ………………………………………………………………………………………
    8-8

    8.4.2 Parts List
    ………………………………………………………………………………………………
    8-8

    8.5 Rear
    Housing…………………………………………………………………………………………………..
    8-10

    8.5.1 Exploded View
    …………………………………………………………………………………….
    8-10

    8.5.2 Parts List
    ……………………………………………………………………………………………..8-11

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    9/168

    5

    8.6 Measurement Module Panel
    Assembly……………………………………………………………….
    8-12

    8.6.1 Exploded View
    …………………………………………………………………………………….
    8-12

    8.6.2 Parts List
    …………………………………………………………………………………………….
    8-12

    8.7 Power Supply Assembly
    …………………………………………………………………………………..
    8-14

    8.7.1 Exploded View
    …………………………………………………………………………………….
    8-14

    8.7.2 Parts List
    …………………………………………………………………………………………….
    8-15

    8.8 MPM
    Module………………………………………………………………………………………………….
    8-16

    8.8.1 Exploded View
    …………………………………………………………………………………….
    8-16

    8.8.2 Parts List
    …………………………………………………………………………………………….
    8-17

    8.9 Capacitor Assembly
    …………………………………………………………………………………………
    8-18

    8.9.1 Exploded View
    …………………………………………………………………………………….
    8-18

    8.9.2 Parts List
    …………………………………………………………………………………………….
    8-18

    8.10 Sidestream CO2 Module Kit
    ……………………………………………………………………………
    8-19

    8.10.1 Exploded View
    …………………………………………………………………………………..
    8-19

    8.10.2 Parts List
    …………………………………………………………………………………………..
    8-19

    8.11 Microstream CO2 Module Kit
    …………………………………………………………………………
    8-20

    8.11.1 Exploded View
    …………………………………………………………………………………..
    8-20

    8.11.2 Parts
    List……………………………………………………………………………………………
    8-20

    8.12 Paddle Tray
    Assembly…………………………………………………………………………………….
    8-21

    8.12.1 Exploded View
    …………………………………………………………………………………..
    8-21

    8.12.2 Parts List
    …………………………………………………………………………………………..
    8-21

    8.13 External Paddle
    ……………………………………………………………………………………………..
    8-22

    8.13.1 Exploded View
    …………………………………………………………………………………..
    8-22

    8.13.2 Parts List
    …………………………………………………………………………………………..
    8-22

    8.14 Sternum Pediatric Paddle Kit
    …………………………………………………………………………..
    8-23

    8.14.1 Exploded View
    …………………………………………………………………………………..
    8-23

    8.14.2 Parts List
    …………………………………………………………………………………………..
    8-23

    8.15 Sternum Adult Paddle Kit
    ……………………………………………………………………………….
    8-24

    8.15.1 Exploded View
    …………………………………………………………………………………..
    8-24

    8.15.2 Parts List
    …………………………………………………………………………………………..
    8-24

    8.16 Apex Pediatric Paddle Kit
    ……………………………………………………………………………….
    8-25

    8.16.1 Exploded View
    …………………………………………………………………………………..
    8-25

    8.16.2 Parts List
    …………………………………………………………………………………………..
    8-25

    8.17 Apex Adult Paddle Kit
    ……………………………………………………………………………………
    8-26

    8.17.1 Exploded View
    …………………………………………………………………………………..
    8-26

    8.17.2 Parts List
    …………………………………………………………………………………………..
    8-26

    8.18 External Paddle Cable
    …………………………………………………………………………………….
    8-27

    8.18.1 Exploded View
    …………………………………………………………………………………..
    8-27

    8.18.2 Parts List
    …………………………………………………………………………………………..
    8-27

    8.19 Hook Mount
    ………………………………………………………………………………………………….
    8-28

    8.19.1 Exploded View
    …………………………………………………………………………………..
    8-28

    8.19.2 Parts List
    …………………………………………………………………………………………..
    8-28

    8.20 Replacement
    Parts………………………………………………………………………………………….
    8-29

    8.20.1 Main Unit
    ………………………………………………………………………………………….
    8-29

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    10/168

    6

    8.20.2 Connecting Cables
    ……………………………………………………………………………..
    8-32

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    11/168

    1-1

    1 Safety

    1.1 Safety Information

    DANGER

      Indicates an imminent hazard that, if not avoided, will
    result in death, serious

    personal injury or property damage.

    WARNING

      Indicates a potential hazard or unsafe maintenance
    practice that, if not avoided,

    could result in death, serious personal injury, product /
    property damage.

    CAUTION

      Indicates a potential hazard or unsafe maintenance
    practice that, if not avoided,

    could result in minor personal injury or product/property
    damage

    NOTE

      Provides application tips or other useful information to
    ensure that you can better

    service your product.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    12/168

    1-2

    1.1.1 Dangers

    WARNING

      Do not open the equipment cases to avoid shock hazard.
    All servicing and future

    upgrades must be carried out by the personnel trained and
    authorized by our

    company only.

    1.1.2 Warnings

    WARNING

      To avoid high voltage shock, disconnect the
    defibrillator/monitor from AC adapter

    and remove the batteries before disassembly.

      The equipment must be connected to a properly installed
    power socket with

    protective earth contacts only. If the installation does not
    provide a protective

    earth conductor, do not use this socket and operate the
    equipment on rechargeable

    batteries.

      When disposing of the packaging material, be sure to
    observe the applicable waste

    control regulations and keep it out of children’s reach.

    1.1.3 Cautions

    CAUTION

      Make sure that no electromagnetic radiation interferes
    with the performance of the

    equipment when preparing to carry out performance tests. Mobile
    phone, X-ray

    equipment or MRI devices are a possible source of interference
    as they may emit

    higher levels of electromagnetic radiation.

      Before connecting the equipment to the power line, check
    that the voltage and

    frequency ratings of the power line are the same as those
    indicated on the

    equipment’s label or in this manual.

      Protect the equipment from damage caused by drop, impact,
    strong vibration or

    other mechanical force during servicing.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    13/168

    1-3

    1.1.4 Notes

    NOTE

      Refer to Operation Manual for detailed operation and
    other information.

    1.2 Equipment Symbols

    Attention: Please read thismanual carefully before

    servicing.

    Equipotential terminal

    Danger: High-voltage Service indicator

    Alternating current(AC) Network connector

    Battery Video output

    USB connector Analog input/out

    ESD warning symbol for Electrostatic sensitive devices.

    Type CF applied part. Defibrillator-proof protection against
    electric shock.

    Type BF applied part. Defibrillator-proof protection against
    electric shock.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    14/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    15/168

    2-1

    2 Theory of Operation

    2.1 The Basics

    2.1.1 Overview

    BeneHeart D6 defibrillator/monitor (hereinafter called the
    equipment) provides four

    operating modes: Manual Defib, AED, Pacer, and Monitor. The
    equipment is for use in

    hospital and pre-hospital settings. It adopts the most advanced
    biphasic defibrillation

    technology and can deliver up to 360J of defibrillation
    energy.

    BeneHeart D6 has an 8.4″ LCD display.

    2.1.2 Main Functions

    The equipment has the following main functions:

      Manual Defib Mode

    In Manual Defib Mode, the operator analyzes the patient’s ECG,
    and, if appropriate, follows

    this procedure:

    1 Select the Manual Defib mode, adjust the energy level if
    necessary

    2 Charge; and

    3 Deliver the shock.

    Defibrillation may be performed through external paddles or
    multifunction electrode pads. In

    Manual Defib Mode, you can also perform synchronized
    cardioversion.

      AED Mode

    In AED mode, the equipment automatically analyzes the patient’s
    ECG rhythm and

    indicates whether or not a shockable rhythm is detected. Voice
    prompts provide

    easy-to-follow instructions and patient information to guide you
    through the

    defibrillation process. Messages and flashing buttons are also
    presented to reinforce the

    voice prompts.

      Pacer Mode

    The Pacer Mode offers non-invasive transcutaneous pacing
    therapy. Pace pulses are

    delivered through multifunction electrode pads using a
    monophasic square waveform.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    16/168

    2-2

      Monitor Mode

    In Monitor Mode, the equipment is intended for monitoring,
    displaying, reviewing,

    storing and printing multiple physiological parameters and
    waveforms including ECG,

     pulse oximetry (SpO2), temperature (Temp), non-invasive
    blood pressure (NIBP),

    invasive blood pressure (IBP) and carbon dioxide (CO2).

    2.2 Components

    The equipment consists of a main unit, accessories and PC
    software.

    The main unit is the core of the equipment. It provides:

    Overall system control;

      System power supply;

      Display;

      Defibrillation and pacing;

      AED ;

      Man-mahcine interface;

      Audible and visible alarms;

    Multiple parameter measurements;

      External connectors and communication; and

      Recording, printing and data storage.

    2.3 Main Unit

    The main unit is composed of the front housing assembly, rear
    housing assembly and the

     paddle tray assembly. External paddles are rested in the
    paddle tray.

      The front housing assembly mainly consists of LCD, keypad
    board, recorder, speaker,

    microphone, Mode Select knob, navigation knob, backlight
    inverter, alarm lamp board,

    indicating lamp board, front housing and front housing sheet
    metal, etc.

      The rear housing assembly consists of CPU board, therapy
    module, high voltage

    capacitors, MPM module, CO2 module, power management board,
    wireless network

    adapter, fan, measurement module panel, therapy port, and rear
    housing, etc.

      The paddle tray is for holding the external paddles.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    17/168

    2-3

    The main unit consists of the following subsystem:

      Input subsystem: Its input includes keypad board,
    microphone, Mode Select knob,

    navigation knob, and paddle handle controls.

    Output subsystem: includes display screen, alarm lamp board,
    recorder, and speaker

      Processing and communication subsystem: includes CPU
    board, therapy module, MPM

    module, and CO2 module.

      Power management subsystem: includes batteries, AC/DC
    board and power

    management board.

      External device connection subsystem: includes USB
    connector, network connector,

    VGA connector and multifunction connector for analog output and
    synchronous input.

    System Structure

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    18/168

    2-4

    System Signal Flow

    2.4 Front Housing Assembly

    The front housing assembly consists of display assembly
    (including the backlight inverter), a

    keypad board, a recorder, a speaker, a microphone, a Mode Select
    knob, a navigation knob,

    an alarm lamp board, an indicating lamp board, a front housing
    and front housing sheet metal,

    etc.

    Navigation Knob

    You can rotate the knob clockwise or counterclockwise and then
    press it to confirm a

    selection. The knob is connected to the keypad board.

    Mode Select Knob

    A 8-position encoder is used to select the operating mode
    (Monitor, Manual Defib, AED and

    Pacer) and power-off. The unused positions are mechanically
    disabled.

    Recorder

    The recorder receives data from the CPU board and then sends the
    data to a thermal head for

     printing. The recorder front panel has a key for starting/
    stopping the recorder and a green

    indicator which is lit when working normally. The recorder is
    connected to the keypad board

    which board provides connection for the TR6F recorder. The block
    diagram and functional

    modules of the recorder are shown as below.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    19/168

    2-5

    Module Description

    Power Interface Introduces DC power supply from the CPU
    board.

    Recorder Power Module Adjusts input voltage to run each
    module.

    Recorder CPUCoordinates module communication, controls and
    processes module

    status.

    Keypad board InterfaceServes as the data communication channel
    between the keypad board

    and the recorder CPU.

    Motor Drive CircuitReceives control signals sent by the recorder
    CPU to drive the step

    motor.

    Keypad and Indicator

    Interface

    Sends keypad commands to CPU and receives CPU commands to

    control the indicator.

    FPC InterfaceSends print head information to CPU and receives
    CPU commands to

    control the print head.

    Speaker

    The speaker emits alarm tones, key-stroke tone, heart beats and
    PR sound. It supports the

    functions of PITCH TONE and the multi-level volume. The speaker
    is connected to the

    keypad board.

    Microphone

    It provides the function of voice recording.

     Alarm Lamp Board

    The keypad board interfaces with the alarm lamp board. The alarm
    lamp transmits signals to

    drive the green and yellow alarm lamp.

    Indicating Lamp Board

    The keypad board interfaces with the indicating lamp board. On
    the equipment’s front panel,

    there are 3 indicators: AC power indicator, battery indicator
    and service indicator, each has an

    icon aside.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    20/168

    2-6

    2.5 Paddle Tray

    The paddle tray is used to hold paddles. It has a 50 Ω test
    load inside. When the equipment

    runs self tests, test current will pass through the test
    load.

    2.6 Rear Housing Assembly

    Rear housing assembly consists of the CPU board, the therapy
    module, high voltage

    capacitors, a MPM module, a CO2 module, a power management
    board, a fan, a rear housing,

    a measurement module panel, and a therapy port, etc.

    2.6.1 Power System

    The power system consists of the following items:

      AC/DC board: The AC mains is the input, and the outputs
    is 18VDC.

      Battery: 14.8V, 4500mAh.

      Power Management Board

    It is intended for power transform and battery charge control.
    The system outputs four

    types of power supply: 18V (when AC mains is used) or 14.8V
    (when a battery is used),

    12V, 5V, and 3.3V.

    The priority of system power supply is rated in the order of AC
    mains, Battery 1 and

    Battery 2. That is to say, when AC is not available, Battery 1
    will be used; if Battery 1 is

    defective or depleted, Battery 2 will be used.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    21/168

    2-7

    2.6.2 Main Control System

    The CPU board is connected with the power management board with
    stacking connectors, as

    shown below.

    The main control module mainly consists of the CPU and FPGA. CPU
    is used to provide

    least required internal storage, program memory, large capacity
    non-volatile storage, and the

    watch dog. It connects EEPROM and other peripheral ICs such as
    Ethernet PHY chip. FPGA

     performs the main functions of display and audio. Besides,
    it has the function of adapting

    interfaces from CPU to MPM module, the keypad board and the
    recorder. CPU controls

    FPGA via Flexbus.

    2.6.3 Therapy System

    The therapy system provides the functions of defibrillation,
    pacing and AED analysis. The

    therapy module is an unseparated assembly.

    The therapy module adopts DSP+MCU framework. MCU is responsible
    for therapy control

    while DSP for ECG and impedance detection, AED algorithm,
    monitoring algorithm, pacing

    algorithm, auxiliary therapy control, etc.

    2.6.4 Parameter Measurement System

    MPM modules and the CO2 module are used to provide
    parameter monitoring. However,

    ECG monitoring can also be implemented by the therapy module.
    .

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    22/168

    2-8

    2.7 External Device Connectors

    1. Hook

    2. Battery 2

    3. Battery 1

    4. External power input: It connects an AC power cord or a DC/AC
    adapter to run the

    equipment respectively on the external AC mains or DC power
    supply.

    5. Equipotential grounding terminal: When the
    defibrillator/monitor and other devices are

    to be used together, their equipotential grounding terminals
    should be connected

    together to eliminate the potential difference between them.

    6. USB connector: It connects the USB memory for data export.
    Data stored in the internal

    CF card can be transferred to the USB memory and then export to
    a PC via the data

    management software.

    7. Network connector: It is a standard RJ45 connector, through
    which software can be

    upgraded.

    8. Multifunctional connector: provides ECG analog output and
    defibrillation

    synchronization input.

    9. VGA connector: connects an external VGA display.

    1

    2

    4

    3

    6

    7

    8

    9

    5

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    23/168

    3-1

    3 Unpacking and InstallationThis chapter provides information
    you need to install a defibrillator/monitor ready for use.

    3.1 Unpacking the Equipment

    Open the package and take out the packing list. Check that all
    the articles included in the

     packing list are available and the quantity and
    specification are correct.

      All the optional parts purchased by the customer shall
    also be checked.

     Notify the supplier if provided components are not correct
    as compared to the packing

    list.

      In case of damage during transportation, keep the packing
    material and notify the

    supplier immediately.

      Keep the packing material till new equipment is
    accepted.

    The following pictures show the defibrillator/monitor and
    accessory packing.

    Main unit packing

    Accessory packing

    Main unitAccessory packing

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    24/168

    3-2

    3.2 Preparation for Installation

    3.2.1 Preparation for Installation Site

    1. Ensure that the site meets all safety, environmental and
    power requirements

    2. Check that required power sockets are available.

    3. Check that a network connector is available if the
    defibrillator/monitor needs to be

    connected to network.

    WARNING

      Only power cables provided with the system may be used.
    For reasons of safety,

    power (mains) extension cables or adapters shall not be
    used.

    Environmental Requirements

    WARNING

      To avoid explosion hazard, do not use the equipment in
    the presence of flammable

    anaesthetics, vapours or liquids.

    CAUTION

      The environment where the defibrillator/monitor will be
    used should be reasonably

    free from vibration, dust and corrosive substances. If these
    conditions are not met,

    the accuracy of the system may be affected and damage may
    occur.

    The environmental specification is as follows:

    Operating Temperature

    0 to 45℃  (without CO2 module)

    5 to 35℃  (with sidestream CO2 module)

    0 to 40ºC (with microstream CO2 module)

    Operating humidity 15% to 95%, (non-condensing)

    Operating altitude -381m to +4575 m (-1250 ft to 15000 ft, or
    106.2kPa to 57kPa)

    Storage temperature -30 to 70℃ 

    Storage humidity 10% to 95%, (non-condensing)

    Storage altitude -381m to +4575 m (-1250 ft to 15000 ft, or
    106.2kPa to 57kPa)

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    25/168

    3-3

    3.2.2 Electrical Requirements

    Check cables and power cords. Make sure that: 

    1. All system cables, power cords and power plugs are not
    damaged, and pins are not loose.Otherwise, remove it from use.

    2. The insulation of patient cables and leadwires is not
    damaged, and connectors are not

    loose.

    WARNING

      Only power sockets with protective grounding can be
    used.

    The electrical specification is as follows:

    Line voltage: 100 to 240VAC

    Current: 1.8 to 0.8 A

    Frequency: 50/60Hz

    3.3 Vehicle Mount Kit Installation

    Refer to BeneHeart D6/D5 Vehicle Mount Kit Instructions for
    Use for the detailed

    information on how to install the equipment on the vehicle mount
    kit.

    3.4 Installing Hook Kit or Conductive Gel Container Kit

    Refer to Hook Kit and Conductive Gel Container Kit
    Installation Guide to install the hook

    kit or conductive gel container kit.

    3.5 Preparation for Power On

    Before connecting the power cord to the defibrillator/monitor’s
    power input, check that

      The mains voltage meets the requirement.

      3-wire power cord is used. The power socket should be
    3-wire also. This ensures that the

    defibrillator/monitor is properly grounded. Do not use 2-wire
    power cord or socket.

      The equipotential grounding terminals should be connected
    together when the

    defibrillator/monitor and other devices are to be used
    together.

      The defibrillator/monitor is not placed under the
    infusion bag or placed where their

    might be liquid spillage. This protects the
    defibrillator/monitor from liquid ingress.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    26/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    27/168

    4-1

    4Testing and Maintenance

    4.1 Introduction

    To ensure the equipment always functions normally, qualified
    service personnel should

     perform regular inspection, maintenance and test. This
    chapter provides a checklist of the

    testing procedures for the equipment with recommended test
    equipment and frequency. The

    service personnel should perform the testing and maintenance
    procedures as required and use

    appropriate test equipment.

    The testing procedures provided in this chapter are intended to
    verify that the equipment

    meets the performance specifications. If the equipment or a
    module fails to perform as

    specified in any test, repairs or replacement must be done to
    correct the problem. If the

     problem persists, contact our Customer Service
    Department.

    CAUTION

      All tests should be performed by qualified service
    personnel only.

      Care should be taken to change the settings in
    [Installation Mode] and [Service

    Mode] menus to avoid loss of data.

      Before testing, service personnel should acquaint
    themselves with the test tools and

    make sure that test tools and cables are applicable.

      When testing monitoring parameters, move the Mode Select
    knob to Monitor to

    access the Monitor Mode.

      When performing therapy function tests, move the Mode
    Select knob to

    corresponding mode.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    28/168

    4-2

    4.1.1 Test Report

    After completing the tests, service personnel are required to
    record test results and report

    them to Mindray Customer Service Department.

    See the

    Test Report at the end of this chapter.

    4.1.2 Preventative Maintenance

    Below are preventative maintenance tests which need to be
    performed on the

    defibrillator/monitor. See the following sections for detailed
    maintenance procedures.

      Visual inspection

       NIBP test and calibration

      Microsteam and Sidestram CO2 test and calibration.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    29/168

    4-3

    4.1.3 Recommended Frequency

    Test itemAfter

    repair

    Function

    suspected

    6

    months

    12

    months

    24

    months

    Visual inspection × 

    Power-on Test × 

    User test × 

    Recorder check × 

    Charge/

    discharge

    Energy disarming

    Manual

    defibrillation

    tests Synchronous

    defibrillation

    Pacing test

    ×  ×  × 

    Performance test ×  ×  × ECG

    Module calibration ×  × 

    Resp Performance test ×  ×  × 

    SpO2 Performance test ×  ×  × 

    Temp Performance test ×  ×  × 

    Accuracy test

    Leakage test NIBP

    Module calibration

    ×  ×  × 

    Performance testIBP

    Pressure calibration

    Leakage test ×  ×  × CO2 

    Module calibration ×  ×  × 

     NIBP overpressure protection test ×  × 
    × 

    Analog out test × 

    Earth leakage current

    Patient leakage

    current

    Electrical

    safety tests as

     per

    IEC60601-1 Patient auxiliary

    current

    ×  × 

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    30/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    31/168

    4-5

    4. Compare the value of manometer with the value displayed on
    the equipment’s screen.

    The difference should be no greater than 3 mmHg.

    5. Raise the pressure in the metal vessel to 50 mmHg with the
    balloon pump. Repeat steps

    3 and 4.

    6. Raise the pressure in the metal vessel to 200 mmHg with the
    balloon pump. Repeat steps

    3 and 4.

    Note

      You can replace the balloon pump and manometer with a
    blood pressure simulator

    to form a test system.

    NIBP Leakage Test

    Tools required:

      An adult cuff

      An air tubing

      A correct sized cylinder

    To perform the leakage test:

    1. Connect the cuff to the equipment’s NIBP connector.

    2. Wrap the cuff around the cylinder as shown below.

    3. Press the Main menu button on the equipment’s front panel.
    Select [Others>>]→ 

    [Installation Mode>>]→ enter the required
    password→[Maintain NIBP]→ [Start 

    Leakage Test].

    After about 20 seconds, the equipment automatically deflates.
    This means the leakage test

    finishes.

    When the accuracy test is completed, the result will be
    displayed. If the message [NIBP

    Pneumatic Leak ] is displayed, it indicates that the NIBP
    airway may have leakages. Check

    the tubing and connections for leakages, and then perform a
    leakage test again.

    Defibrillator/monitor 

    Connector for 

      NIBP cuf 

    Tubing

    Cylinder 

    Cuff

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    32/168

    4-6

    Calibrating NIBP

    Tools required:

      T-shape connector

      Tubing

      Balloon pump

      Metal Vessel, volume 500±25 ml

      Calibrated manometer, accuracy higher than 1 mmHg

    1. Connect the equipment as shown below

    2. Before inflation, the reading of the manometer should be 0.
    If not, disconnect the

    airway and reconnect it until the readings is 0

    3. Press the Main menu button on the equipment’s front panel.
    Select [Others>>]→ 

    [Service Mode>>]→ enter the required
    password→[Calibrate NIBP].

    4. Calibrate pressure. To do so, set the calibration value to
    150 mmHg and adjust the pump

    output pressure to 150 mmHg. After the system is stable, click
    the [Calibrate] button to

    start calibration.

    5. Calibrate overpressure. To do so,

      set [Patient Cat.] to [Adu/Ped] and adjust pump output
    pressure to 330 mmHg.

    Click the [Calibrate] button and start calibration. Or

      set [Patient Cat.] to [Neo] and adjust pump output
    pressure to 165 mmHg. Click

    the [Calibrate] button and start calibration.

    All the calibration results will be displayed in the
    [Calibrating NIBP] screen. If the

    calibration fails, please check the connections and then perform
    a calibration again.

    Defibrillator/monitor

    Connector for

     NIBP cuff

    Manometer 

    Tubing

    Balloon pumpMetal vessel

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    33/168

    4-7

    4.2.3 CO2 Module Tests

    Leakage Check

    1. Access the [CO2 Setup] menu and set [Operating
    Mode] to [Measure].Wait for  CO2 

    warm-up. 

    2. Block the CO2 module gas inlet completely. This will
    cause different reactions from the

    Sidestream and Microstream CO2 modules.

      Sidestream: Check that alarm message [CO2 Filter Line
    Err] is displayed on the

    screen in 3s. Block the gas inlet for another 30s, if the alarm
    message does not

    disappear, the module does not leak.

      Microstream CO2 module: [CO2
    Purging…] is displayed in 3s. Block the gas
    inlet

    for another 30s, if the alarm message [CO2 Tubing Err] appears,
    the module does

    not leak.

    Module Calibration

    Test tools

      Gas cylinder, with 6% of CO2 and balance gas N2.

      T-shape connector

      Tubing

    For sidestream CO2 module, zeroing is required before
    calibration. Enter [CO2 Setup] menu

    and select [Zero] to perform zeroing.

    To calibrate the CO2 module, follow this procedure:

    1. Make sure that the CO2 module has been warmed up or
    started up.

    2. Connect the gas cylinder with the tubing using a T-shape
    connector as shown below.

    Check the airway and make sure there are no leaks.

    Defibrillator 

    /monitor

    Gas cylinder 

    To the air

    Gas valveTubing

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    34/168

    4-8

    3. Vent the tubing to the CO2 by opening the gas valve.

    4. Access the [Maintain CO2] menu. To do so, Press the Main Menu
    button on the

    equipment’s front panel. Select [Others>>]→[Installation
    Mode>>]→enter the

    required password→ [Maintain CO2].

    5. In the [Maintain CO2] menu, select a CO2 value equal to the
    vented CO2 

    concentration.

    6. In the [Calibrate CO2] menu, the measured
    CO2 concentration is displayed. Wait till

    the measured CO2 concentration becomes stable; select
    [Calibrate] to start CO2 

    calibrate.

    The message [Calibration Completed!] is displayed after a
    successful calibration. If the

    calibration failed, the prompt [Calibration Failed!] will be
    displayed. In this case, perform

    another calibration.

    4.2.4 Preventative maintenance test report

    Customer name

    Customer address

    Servicing person

    Servicing company

    Equipment under test

    (EUT)

    Model of EUT

    SN of EUT

    Hardware version

    Software version

    Test equipment Model/No. Effective date of calibration

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    35/168

    4-9

    Test items Test records Test results

    (Pass/Failed)

    Visual inspection

    The case, display screen, buttons, knob, SMR, modules, power

    cord, wall-mount bracket and accessories have no obvious
    signs

    of damage.

    The external connecting cables are not frayed and the
    connector

     pins are not loose and bent.

    The external connectors are not loose or their pins are not
    bent.

    The safety labels and data plate are clearly legible.

    NIBP test

    The difference is within ±3 mm when 0, 50 or 200 mmHg is set

    for NIBP accuracy test.

    There is no leakage with NIBP, or the manual leakage test
    result

    does not exceed 6mmHg/min.

    Sidestream CO2 test

    Block the gas inlet of the module or watertrap. The
    sidestream

    CO2 flowrate is slower than 10ml/min and an alarm of CO2

    Filterline Err is given. It indicates that there is no
    leakage.

    The displayed CO2 value is within 6±0.05%.

    Microstream CO2 test

    Block the gas inlet of the module or watertrap. An alarm of
    CO2

    Filterline Err is given. It indicates that there is no
    leakage.

    The displayed CO2 value is within 6±0.05%.

    4.3 Power On TestThis test is to verify that the defibrillator/
    monitor can power on normally. The test is passed if the

    defibrillator/ monitor starts up by following this
    procedure:

    1. Place the external paddles on paddle tray, insert the battery
    (install both if two batteries

    are configured) in the battery compartment, and then connect the
    equipment with AC

    mains. In this case, both the AC indicator and battery indicator
    shall light.

    2. Turn the Mode Select knob to Monitor. Check that the
    equipment passes the self test and

    is turned on properly.

    3. Check the display of technical alarm area, prompt area and
    battery status indicator on

    the upper right corner of the main screen to judge whether the
    equipment runs normally.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    36/168

    4-10

    4.4 User Test

    Follow this procedure to perform user test:

    1. If you use external paddles, place them on the paddle tray;
    if you use a pads cable,connect it to the 50 Ω test load.

    2. Insert the battery (2 if configured) into the equipment.
    Connect the AC mains if no

     battery is available.

    3. Select the Main Menu button on the equipment’s front panel.
    In the Main Menu, select

    [User Test>>]. Then a prompt “Enter user test?” pops up.
    Select “Yes” to enter the User

    Test screen.

    4. Check the test items you want to perform and select [Start]
    to start user test

    The test results indicate the condition of the system. If any
    item fails, the service indicator

    flashes.

    If you cannot pass User Test or the message “Connect paddles
    cable, and place paddles in

     paddle tray” is shown when paddle cable is connected and
    paddles are placed in paddle tray,

    check paddles status.

    Select the Monitor mode. Press and hold the [Event] hardkey, and
    then press the [Lead

    Select] hardkey on the front panel, the following screen
    appears.

    Observe the reading of “Lead Stat”:

      0 x 382: Paddles are properly placed in paddle tray.

      0 x 182: The travel switch indicating paddle status may
    fail, but impedance is correct.

      0 x 102 :Paddles are not properly placed in paddle tray
    and the impedance value is not

    correct.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    37/168

    4-11

    4.5 Password for Installation Mode

    Accessing installation mode is password protected. The required
    password is set to 888888

     before the equipment leaves the factory.

    4.6 Module Performance Tests

    4.6.1 Manual Defibrillation Test

    Test tools:

      Defibrillator/pacer analyzer

    Charge/Discharge

    1. Remove the batteries and connect the equipment with AC mains.
    Turn the Mode Select

    knob to Manual Defib.

    2. Connect the external paddles to the equipment and place the
    paddles on the

    defibrillator/pacer analyzer.

    3. Enter the Configuration-Main screen. From the Record Setup
    menu set [Shock Event]

    to [On] so that shock events can be recorded automatically if
    happened.

    4. Set the analyzer to Energy Measurement mode. In this case,
    the energy value should be

    displayed as 0 or blank.

    5. Select the energy level to 1J.

    6. Charge/discharge the equipment to verify the energies
    measured by the analyzer meet

    the following accuracy:

    Selected Energy (J) Measured Value (J)

    1 0 to 3

    100 85 to 115

    360 306 to 414

    7.  Set the energy to 100J and 360J respectively. Repeat
    Step 6.

    8. Disconnect the equipment from the AC mains. Run the equipment
    on fully charged

     battery. Move the Mode Select knob to Manual Defib. Repeat
    Steps 5 to 7.

    9. Use multifunctional electrode pads. Repeat Step 5 to Step
    7.

    10. Verify that the equipment records the shock events
    automatically and correctly.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    38/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    39/168

    4-13

    7. When charging finishes, press and hold the “Shock” button to
    deliver a shock.

    8. Verify that synchronous discharge succeeds and the delivery
    energy measured by the

    analyzer is 10J±2J.

    9. Verify that the delay time of synchronous defibrillation
    measured by the analyzer is less

    than 60ms.

    10. Verify that the synchronous discharge mark appears on the R
    wave.

    11. Verify that the prompt messages are correct during
    testing.

    12. Select lead II as ECG source and perform charging. Repeat
    steps 7 to 11.

    13. Use multifunctional electrode pads. Repeat steps 2 to
    12.

    4.6.2 Pacing Test

    Test tools:

      Defibrillator/pacer analyzer

    1. Run the equipment on fully charged battery. Move the Mode
    Select knob to Pacer. Set

    [Pacer Mode] to [Fixed].

    2. Connect the pads cable to the equipment and properly place
    the pads on the

    defibrillator/pacer analyzer.

    3. Set the analyzer to Pacing Measurement mode. Use test load of
    50Ω.

    4. On the equipment, set [Pacer rate] to [70ppm] and [Pacer
    Output] to [30mA].

    5. Press the [Start Pacing] soft key. Verify that the pacer rate
    measured by the analyzer is

    70 ppm±1ppm and the pacer output measured is 30 mA±5mA.

    6. Press the “Stop Pacing” soft key, and then set [Pacer rate]
    to [170ppm] and [Pacer

    Output] to [200mA].

    7. Press the [Start Pacing] soft key. Verify that the pacer rate
    measured by the analyzer is

    170 ppm±2ppm, and the measured current is 200 mA±10mA.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    40/168

    4-14

    4.6.3 ECG Test

    Performance Test

    Test tools

      ECG simulator

    1. Connect the simulator to the equipment’s ECG connector.

    2. Configure the simulator as HR=80 bmp.

    3. The displayed HR should not exceed 80±1 bmp;

    ECG Calibration

    1. Select the ECG parameter area to enter the [ECG Setup]
    menu.

    2. Select [Others>>]→[Calibrate]. A waveform signals
    appear on the screen and the

    message [ECG Calibrating] is displayed in the prompt information
    area in the lower

    left corner of the screen.

    3. Compare the amplitude of the waveform with the wave scale.
    The difference should be

    within 5%. If needed, you can also print out the waveform and
    the wave scale.

    4. After ECG calibration is completed, select [Stop
    Calibrating].

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    41/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    42/168

    4-16

    Pressure Calibration

    Tools required:

      Standard sphygmomanometer

      Balloon pump

      Tubing

      T-shape connector

    To perform a calibration:

    1. Connect the 3-way stopcock, the sphygmomanometer and the
    balloon pump through a

    T-shape connector, as shown below.

    2. Zero the transducer. Then open the stopcock to the
    sphygmomanometer.3. Press the Main menu button on the equipment’s
    front panel. Select [Others>>]→ 

    [Installation Mode>>]→ enter the required
    password→[Maintain IBP]. Then

    configure IBP calibration value.

    4. Inflate using the balloon pump until the reading of
    sphygmomanometer approximates

    the preset calibration value.

    5. Adjust the calibration value in the [Maintain IBP] menu until
    it is equal to the reading

    of sphygmomanometer

    6. Select the [Calibrate] button to start a calibration

    7. The message [Calibration Completed!] is displayed after a
    successful calibration. If

    the calibration failed, the prompt [Calibration Failed!] will be
    displayed.

    Sphygmomanometer

    T-shape connector 

    3-way stopcock 

    Pressure transducer 

    Pressure adapter cable

    IBP Module

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    43/168

    4-17

    4.6.6 SpO2 Test

    Test tool

    Patient simulator.

    1. Connect the patient simulator to the equipment’s SpO2
    connector  .

    2. Select the model and manufacturer of the SpO2 module under
    test. Configure the

     parameter as SpO2 96% and PR 80 bmp.

    3. The displayed SpO2 and PR values should be within the ranges
    listed below

    SpO2 (%) PR (bmp)

    Mindray 96% ±2% 80±3

    Masimo 96% ±2% 80±3

    MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I,

    MAX-FAST96% ±2%

    OxiCliq A, OxiCliq N, OxiCliq P, OxiCliq I 96% ±2.5%

    D-YS, DS-100A, OXI-A/N, OXI-P/I 96% ±3%

     Nellcor

    MAX-R, D-YSE, D-YSPD 96% ±3.5%

    80±3

    4.6.7 Temp Test

    Test tools

      Resistance box

    1. Connect the two pins of any Temp connector on equipment to
    the two ends of the

    resistance box using 2 wires.

    2. Set the resistance box to 1354.9Ω (corresponding
    temperature is 37 ). The displayed℃

    value on the equipment should not exceed 37±0.2 .℃

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    44/168

    4-18

    4.7 Analogue Output Test

    Test tool:

    Oscillograph

    1. Connect the patient simulator to the equipment under test
    using an ECG or IBP cable.

    2. Connect the oscillograph to the equipment’s multifunctional
    connector.

    3. Verify that the waveforms displayed on the oscillograph is
    identical with those

    displayed on the defibrillator/ monitor.

    4.8 Electr ical Safety Tests

    See A Electrical Safety Inspection for electrical
    safety tests..

    4.9 Recorder Check

    Tools required:

       None.

    1. Print ECG waveforms. The recorder should print correctly and
    the printout should be

    clear.

    2. Simulate some recorder problems, such as out of paper, paper
    jam, etc. the defibrillator/

    monitor should give corresponding prompt messages. After the
    problem is removed, the

    recorder should be able to work correctly.

    3. Switch automatic alarm recording for each parameter ON and
    then set each parameter’s

    limit outside set alarm limits. Corresponding alarm recordings
    should be triggered when

     parameter alarms occur.

    4.10 Factory Service

    4.10.1 Password for Service Mode

    Accessing service mode is password protected. The required
    password is set to 332888

     before the equipment leaves the factory.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    45/168

    4-19

    4.10.2 Accessing Service Mode Menu

    To access the factory service menu, Press the Main menu button
    on the equipment’s front

     panel. Select [Others>>]→ [Service
    Mode>>]→ enter the required passwords. The Service

    Mode-Main menu is shown below.

    4.10.3 Calibrating NIBP

    Refer to 4.2.2 NIBP Tests for calibrating NIBP.

    4.10.4 Calibrating/Zeroing Impedance

     Normally impedance calibration and zeroing is unnecessary.
    However, you can perform

    impedance checking after replacing the therapy module.

    1. If not pre-connected, connect the pads cable to the
    equipment.

    2. Connect a test load of 300 ohms to the pads cable.

    3. Start the equipment and select the Monitor mode. Press and
    hold the [Event] hardkey,

    and then press the [Lead Select] hardkey on the front panel, the
    following screen

    appears.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    46/168

    4-20

    4. Verify that the reading of “RT Imped” is between
    3000±450.

    NOTE

      If 300 ohms test load is not available, you can use a 50
    ohms test load to perform

    impedance checking. In this case, Verify that the reading of “RT
    Imped” is

    between 500±75.

    If the reading of “RT Imped” is not correct, perform impedance
    calibration/zeroing.

    1. Press the Main menu button on the equipment’s front panel.
    Select [Others>>]→ [Service Mode>>]→ enter
    the required passwords→[Calibrate/Zero Impedance] to

    enter the Calibrate/Zero Impedance screen.

    2. Connect a test load of 0 ohm to the pads cable; then select
    “Zero”. A message “Zero

    Completed” shall be shown. If the message “Zero Failed” is
    displayed, check the

    connection of pads cable.

    3. Connect a test load of 100 ohms to the pads cable; then
    select “Calibrate”. A message

    “Calibration Completed” shall be shown. If the message
    “Calibration Failed” is

    displayed, check the connection of pads cable.

    Replace the therapy module if impedance calibration/zeroing
    fails.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    47/168

    4-21

    4.10.5 Device Information

    Press the Main menu button on the equipment’s front panel.
    Select [Others>>]→ [Service

    Mode>>]→ enter the required passwords→ [Device
    Information]. In the Device

    Information list, you can view the device information such as
    software version, system status,

    and etc, as shown below.

    In the Device Information screen, you can select [Export] to
    export error codes and shock

    delivery data to a USB flash memory.

    4.10.6 Checking Failure Code

    Press the Main Menu button on the equipment’s front panel.
    Select [Others>>]→ [Service

    Mode>>]→ enter the required passwords→ [Failure
    Code] to check error codes. This helps

    the service personnel to identify failures.

    Refer to 6.8 Error Codes for the description of each error
    code.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    48/168

    4-22

    4.10.7 Inputt ing Serial Number

    Press the Main Menu button on the equipment’s front panel.
    Select [Others>>]→ [Service

    Mode>>]→ enter the required passwords→ [Input
    Serial Number] to input the equipment’s

    serial number.

    After inputting the serial number, you can view it by accessing
    Installation Mode and select

    [Version].

    4.10.8 Paddle Open Circuit Display

    This [Paddle Open Circuit Display] switch is for testing only.
    In normal operation, it should

     be set to [Off ]].

    4.10.9 Selecting Recorder Type

    BeneHeart D6 defibrillator/monitor can be configured with a 50
    mm recorder or an 80 mm

    recorder. You can choose the recorder type in the service mode
    by selecting [ Recorder Type]

    and toggle between [50 mm Recorder] and [80 mm Recorder].

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    49/168

    4-23

    Test Report

    Customer name

    Customer address

    Servicing person

    Servicing company

    Equipment under test

    (EUT)

    Model of EUT

    SN of EUT

    Hardware version

    Software version

    Test equipment Model/No. Effective date of calibration

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    50/168

    4-24

    Test items Test recordsTest results

    (Pass/Failed)

    Visual inspection

    The case, display screen, buttons, knob, modules, power cord,
    and

    accessories have no obvious signs of damage.

    The external connecting cables are not frayed and the connector
    pins

    are not loose and bent.

    The external connectors are not loose or their pins are not
    bent.

    The safety labels and data plate are clearly legible.

    Power-on test

    The power-on test is passed. The power indicator and alarm
    system

    work correctly and the equipment start up properly.

    Performance test

    Manual Defibrillation Test

    When running on AC mains and external paddles are used, the

    equipment can be properly charged and discharged; the energy

    delivered meets accuracy requirement, and shock information
    is

    correctly recorded. 

    When running on fully charged battery and external paddles are
    used,

    the equipment can be properly charged and discharged; the
    energydelivered meets accuracy requirement, and shock information
    is

    correctly recorded. 

    When running on AC mains and multifuncational electrode pads
    are

    used, the equipment can be properly charged and discharged;
    the

    energy delivered meets accuracy requirement, and shock
    information

    is correctly recorded. 

    When running on fully charged battery and multifuncational

    electrode pads are used, the equipment can be properly charged
    and

    discharged; the energy delivered meets accuracy requirement,
    and

    shock information is correctly recorded. 

    When external paddles are used, charge tone is correctly issued
    when

    the equipment is being charged. The prompt «Charged Removed»
    is

    shown on the screen and charge done tone stops when the
    Disarm 

    hotkey is pressed. The equipment does not discharge
    externally.

    When [Time to Auto Disarm] is set to [60s], the prompt
    «Charged

    Removed» is shown on the screen and the charge done tone
    stops

    after 60 seconds at the completion of charging. The equipment
    does

    not discharge externally. 

    When pads are used, the charge tone is correctly issued when
    theequipment is being charged. The prompt «Charged Removed» is

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    51/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    52/168

    4-26

    Test items Test recordsTest results

    (Pass/Failed)

    not exceed 6mmHg/min.

    Temp test

    The value displayed for each Temp channel of the monitor is
    within

    37±0.1ºC.

    IBP test

    The static pressure value displayed for each IBP channel is
    within

    200±2 mmHg.

    The ART and LV waves for each IBP channel are displayed
    correctly.

    Sidestream CO2 test

    Block the gas inlet of the module or watertrap. The sidestream
    CO2flowrate is slower than 10ml/min and an alarm of CO2 Filterline
    Err

    is given. It indicates that there is no leakage.

    The displayed CO2 value is within 6±0.05%.

    Miscrostream CO2 test

    Block the gas inlet of the module or watertrap. An alarm of
    CO2

    Filterline Err is given. It indicates that there is no
    leakage.

    The displayed CO2 value is within 6±0.05% 

    Analog output performance test

    The waves displayed on the oscillograph are identical with
    those

    displayed on the monitor.

    SpO2 test

    The displayed SpO2 and PR values should be within the
    specified

    ranges.

    Electrical safety tests

    Refer to Appendix A Electrical Safety Inspection. All the
    electrical

    safety tests should be passed.

    Recorder check

    The recorder can print ECG waves correctly and the printout is
    clear.

    Set the recorder to some problems such as out of paper, paper
    jam,

    etc. the equipment gives corresponding prompt messages. After
    the

     problem is removed, the recorder is able to work
    correctly.

    Automatic alarm recording for each parameter functions
    correctly

    when parameter alarms occur.

    Tested by: _________________________ Date:
    ________________________

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    53/168

    5-1

    5 Hardware and Software Upgrade

    5.1 Hardware Upgrade

    After upgrade the hardware, please upgrade corresponding
    software.

    5.1.1 Upgrading MPM module

    Upgrading MPM module configured with 3/5-lead ECG only

    You can upgrade MPM module from 3/5-lead ECG only to any of the
    following

    configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00117-00 Low cost MPM subassembly 12-Lead

    ECG upgrade package

    12-lead ECG

    801-0651-00118-00 12-Lead ECG/Mindray SpO2upgrade

     package

    12-lead ECG/Mindray SpO2

    801-0651-00049-00 NIBP upgrade package 3/5-lead ECG/NIBP

    801-0651-00050-00 Mindray SpO2 upgrade package 3/5-lead
    ECG/Mindray SpO2

    801-0651-00051-00 Masimo SpO2 upgrade package 3/5-lead
    ECG/Masimo SpO2

    801-0651-00052-00 Nellcor SpO2 upgrade package 3/5-lead
    ECG/Nellcor SpO2

    801-0651-00053-00 Mindray SpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Mindray SpO2

    /NIBP/TEMP

    801-0651-00056-00 Mindray SpO2/NIBP/IBP/TEMP

    upgrade package

    3/5-lead ECG/Mindray SpO2

    /NIBP/IBP/TEMP

    801-0651-00059-00 Masimo SpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Masimo SpO2

    /NIBP/TEMP

    801-0651-00060-00 NellcorSpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG /Nellcor SpO2

    /NIBP/TEMP

    801-0651-00061-00 Masimo SpO2/NIBP/IBP/TEMP upgrade

     package

    3/5-lead ECG/Masimo SpO2

    /NIBP/IBP/TEMP

    801-0651-00062-00 NellcorSpO2/NIBP/IBP/TEM upgrade

     package

    3/5-lead ECG/Nellcor SpO2

    /NIBP/IBP/TEMP

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    54/168

    5-2

    Upgrading MPM module configured with 3/5-leadECG/NIBP

    You can upgrade MPM module from 3/5-lead ECG/NIBP to any of the
    following

    configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00053-00 Mindray SpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Mindray SpO2

    /NIBP/TEMP

    801-0651-00056-00 Mindray SpO2/NIBP/IBP/TEMP upgrade

     package

    3/5-lead ECG/Mindray SpO2

    /NIBP/IBP/TEMP

    801-0651-00059-00 Masimo SpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Masimo SpO2

    /NIBP/TEMP

    801-0651-00061-00 Masimo SpO2/NIBP/IBP/TEMP upgrade

     package

    3/5-lead ECG/Masimo

    SpO2/NIBP/IBP/TEMP

    801-0651-00060-00 NellcorSpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Nellcor

    SpO2/NIBP/TEMP

    801-0651-00062-00 NellcorSpO2/NIBP/IBP/TEMP upgrade

     package

    3/5-lead ECG/Nellcor

    SpO2/NIBP/IBP/TEMP

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

    Upgrading MPM module configured wi th 3/5-lead ECG/Mindray
    SpO2

    You can upgrade MPM module from 3/5-lead ECG/Mindray SpO2 to any
    of the following

    configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00118-00 12-lead ECG/Mindray SpO2upgrade

     package

    12-lead ECG

    801-0651-00053-00 Mindray SpO2/NIBP/TEMP upgrade

     package

    3/5-lead ECG/Mindray

    SpO2/NIBP/TEMP

    801-0651-00056-00 Mindray SpO2/NIBP/IBP/TEMP upgrade

     package

    3/5-lead ECG/Mindray

    SpO2/NIBP/IBP/TEMP

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    55/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    56/168

    5-4

    Upgrading MPM module configured with 3/5-lead ECG/Mindray

    SpO2/NIBP/TEMP

    You can upgrade MPM module from 3/5-lead ECG/Mindray
    SpO2/NIBP/TEM to any of the

    following configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

    801-0651-00081-00 IBP upgrade package 3/5-lead ECG/Mindray

    SpO2/NIBP/IBP/TEMP

    Upgrading MPM module configured with 3/5-lead ECG/ Nellcor

    SpO2/NIBP/TEMP

    You can upgrade MPM module from 3/5-lead ECG/ Nellcor
    SpO2/NIBP/TEMP to any of the

    following configuration:

    PN Description of upgrade

    package

    Configuration after upgrade

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

    801-0651-00081-00 IBP upgrade package 3/5-lead ECG/Nellcor

    SpO2/NIBP/IBP/TEMP

    Upgrading MPM module configured with 3/5-lead ECG/ Mindray

    SpO2/NIBP/IBP/TEMP

    You can upgrade MPM module from 3/5-lead ECG/ Mindray
    SpO2/NIBP/IBP/TEMP to any

    of the following configuration:

    PN Description of upgrade package Configuration after

    upgrade

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    57/168

    5-5

    Upgrading MPM module configured with 3/5-lead ECG/ Masimo

    SpO2/NIBP/IBP/TEMP

    You can upgrade MPM module from 3/5-lead ECG/ Masimo
    SpO2/NIBP/IBP/TEMP to any

    of the following configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

    Upgrading MPM module configured with 3/5-lead ECG/Nellcor

    SpO2/NIBP/IBP/TEMP

    You can upgrade MPM module from 3/5-lead ECG/ Nellcor
    SpO2/NIBP/IBP/TEMP to anyof the following configuration:

    PN Description of upgrade package Configuration after
    upgrade

    801-0651-00116-00 12-lead ECG upgrade package 12-lead ECG

    5.1.2 MPM Module upgrade procedure

    1. Remove the MPM module assembly and parameter panel assembly
    as described in 7.3.8

     Disassembling the MPM Module.

    2. Replace the old parameter panel assembly and MPM module
    assembly with those in the

    upgrade package.

    3. Remove the watertrap receptacle unit, or microstream CO2
    connector assembly, or CO2

     panel as described in 7.8 Disassembling the Measurement
    Module Panel Assembly.

    4. Reassemble the watertrap receptacle unit, or microstream CO2
    connector assembly, or

    CO2 panel on the replacement parameter panel assembly in the
    upgrade package.

    NOTE

      If you need to upgrade 3/5-lead ECG to 12-lead ECG,
    insert the 12-lead ECG

    board into the MPM module. Make sure that the 12-lead ECG board
    should be in

    correct direction. Then fix the clip in place.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    58/168

    5-6

    5. Reassemble the equipment.

    6. Stick correct parameter panel overlay on the parameter panel
    according to theconfiguration you upgrade.

      If the upgrade MPM module is equipped with Masimo SpO2,
    you need to stick a

    Masimo label on the lower left corner of the front housing and a
    No Implied

    License label below the parameter panel, as indicated in the
    following pictures.

      If the upgraded MPM module is equipped with Nellcor SpO2,
    you need to stick a

     Nellcor label at the lower left corner of the front
    housing, as indicated in the

    following pictures.

    7. Perform the tests as described in 4.6 Module Performance
    Tests.

    Masimo label No Implied License label

     Nellcor label

    12-lead ECG board

    Clip 1Cli 2

    Clip here in place Clip here in place

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    59/168

    5-7

    5.1.3 Upgrade the Therapy Module

    You can use 801-0652-00039-00 pacer function upgrade kit to
    upgrade the therapy module so

    that equipment has pacing function. After upgrading the therapy
    module, choose Mode label

    (with pacing function) with the language you need.

    Follow this procedure to upgrade the therapy module:

    1. Remove therapy module as described in 7.3.7 Removing the
    Therapy Module

     High-voltage Board . Be noted that you need not to
    remove the parameter panel

    assembly and MPM module assembly.

    2. Take off the Mode Select knob. Peel off the Mode label.

    3. Replace the old therapy module with the one configured with
    pacing function in the

    upgrade package.

    4. Replace the old Mode Select knob and Mode label with those in
    the upgrade package.

    5. Reassemble the equipment.

    6. After upgrade the therapy module, perform the tests described
    in 4.6 Module

     Performance Tests.

    Mode Select knob

    Mode label

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    60/168

    5-8

    5.1.4 Upgrade the Equipment with CO2 Module

    You can use the following CO2 upgrade package to upgrade the
    equipment so as to have a

    CO2 monitoring function.

    PN Description of upgrade package

    801-0651-00079-00 M02B CO2 upgrade package (for adult/pediatric)
    (FRU)

    801-0651-00082-00 M02B CO2 upgrade package (for neonate)
    (FRU)

    801-0651-00080-00 Microstream CO2 upgrade package (for neonate)
    (FRU)

    Follow this procedure to upgrade the equipment:

    1. Remove the parameter module panel assembly as described in
    7.3.4 Removing the

     Measurement Module Panel .

    2. Remove the left capacitor sheet metal as described in 7.3.9
    Removing the CO2 Module.

    3. Install the required M02B CO2 module or microstream CO2
    module on the left

    capacitor sheet metal.

    Mindray CO2 module Microstream CO2 module

    4. Apply the CO2 insulating sheet on the MPM module mounting
    plate. Be careful to align

    the edge of insulating plate with the edge of the metal plate
    when stick the insulating

     plate to the metal sheet. Do not stick the screw holes on
    the metal sheet. Then bent the

    insulating plate properly to avoid being cut by the insulating
    plate during later assembly

     process.

    Make sure that the part number of

    connecting cable for the microstream

    CO2 connector fixing plate should

     be 0651-20-77166.

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    61/168

    5-9

    5. Peel off the overlay on the parameter module panel and
    install the connector fixing part

    of the microstream CO2 module or the watertrap receptacle kit
    and the gas outlet of the

    Mindray CO2 module on the parameter module panel. Choose an
    appropriate overlayfrom the upgrade package and apply it on the
    parameter module panel.

    6. Install the left capacitor metal sheet with CO2 module in the
    machine.

    7. Reassemble the machine. Route the gas pipes to avoid blocking
    them.

    8. Perform the tests as described in 4.2.3 CO2 Module
    Tests.

    Gas outlet

    Mindray CO2

    watertrap kit

    Microstream CO2

    connector fixing part

    CO2

    insulating plate

    Align the edge of

    insulating plate with the

    edge of the metal plate

    Bent here

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    62/168

  • 8/17/2019 BeneHeart D6 Service Manual V8.0 En

    63/168

    5-11

    5.2.1 Installing Mindray Patient Monitor Software Upgrade
    Tool

    1. Find the installation program and double click it to
    start

    installation.

    2. Select installation language.

  • Автоматический
  • Импульс дефибрилляции: бифазная усеченная экспоненциальная кривая
  • Режим работы: АНД
  • Каналы для измерения: ЭКГ
  • Разряды энергии: 200 Дж/400, 360 Дж/200
  • Дисплей: нет
  • Работа от встроенной батареи: 900 мин (15 ч)
  • Для взрослых и детей
  • Регистрационное удостоверение

Арт. 8129

Дефибриллятор Mindray BeneHeart C1A

  • Описание
  • Характеристики
  • Комплектация
  • Данные для транспортировки
  • Документация
  • Дефибриллятор Mindray BeneHeart C1A — автоматическое наружное устройство с подтверждением разряда нажатием кнопки, предназначенный для экстренного оказания помощи пациенту с нарушениями ритма сердца и проведения дефибрилляции при внезапной остановке сердца у взрослых и детей
  • Может быть использован в стоматологических клиниках, общественных местах, общеобразовательных учреждениях, тренажерных залах
  • В дефибриллятор встроен алгоритм анализа шокового ритма, с помощью которого можно получать сигналы с последующим анализом электрокардиограммы (ЭКГ) пациента для выявления необходимости разряда дефибрилляции. В случае обнаружения шокового режима прибор запрашивает подтверждение разряда, затем подает этот разряд, если обнаружен нешоковый ритм — аппарат переходит к режиму сердечно-легочной реанимации
  • Особенности дефибриллятора BeneHeart C1A:
    — импульс дефибрилляции: бифазная усеченная экспоненциальная (BTE) кривая с автоматической компенсацией в зависимости от импеданса пациента
    — диапазон энергии разряда от 10 до 360 Дж, энергия автоматически выставляется в зависимости от типа пациента и его импеданса (сопротивления)
    — возможность использования как для взрослых, так и для детей (выбор детского режима осуществляется переводом переключателя в соответствующее положение и не требует ключа)
    — функция самотестирования и индикации состояния дефибриллятора
    — встроенная память на 1 Гб с возможностью сохранения основных данных пациента и параметров дефибриллятора
    — автоматическая настройка громкости в соответствии с уровнем окружающего шума
    — неперезаряжаемая батарея обеспечивает до 200 разрядов с максимальной энергией 360 Дж или до 400 разрядов с энергией 200 Дж
    — возможен обмен данных по USB порту и не зависит от ОС на ПК
    — срок службы батареи в режиме ожидания до 5 лет
    — автоматическая отмена разряда при импедансе пациента менее 25 Ом и при импедансе более 300 Ом
    — срок годности электродов 3 года
    — режим СЛР (сердечно-легочная реанимация) может быть настроен на 30:2, 15:2 и только непрямой массаж сердца
    — простое переключение между режимом «взрослый/ребенок»
    — метроном СЛР
    — обратная связь в режиме реального времени
    — поддержка процесса сердечно-легочной реанимации: постоянная помощь спасателю в достижении цели спасения жизни
    — голосовые подсказки по проведению реанимационных действий на русском языке, выполняют роль персонального инструктора, который всегда находится рядом

Перед применением необходимо ознакомиться с инструкцией по применению (паспортом изделия) или получить консультацию специалиста

Способ управления

автоматический

Импульс дефибрилляции

бифазная усеченная экспоненциальная кривая

Режим работы

АНД (не синхронизирует разряды с ритмом сердца)

Энергия импульса воздействия для взрослых, Дж

100, 150, 170, 200, 300, 360

Энергия импульса воздействия для детей, Дж

10, 15, 20, 30, 50, 70, 100

Разряды энергии 200 Дж от заряженной батареи

400 (каждую мин 3 разряда)

Разряды энергии 360 Дж от заряженной батареи

200 (каждую мин 3 разряда)

Набор энергии 200 Дж, не более, с

8

Набор энергии 360 Дж, не более, с

15

Программное обеспечение

нет

Электроды

взрослые/детские одноразовые многофункциональные

Работа от встроенной батареи, мин

900 (15 ч)

Электропитание от сети переменного тока, В/Гц

нет

Электропитание от встроенной батареи, В

12

Возможность питания от сети автомобиля

нет

Гарантия производителя, мес.

12

Дефибриллятор Mindray BeneHeart C1A

Перед применением необходимо ознакомиться с инструкцией по применению (паспортом изделия) или получить консультацию специалиста

  • Блок основной BeneHeart C1A — 1 шт.
  • Батарея LM34S002A — 1 шт.
  • Электроды для дефибрилляции модель MR62 (взрослые/детские) — 1 пара
  • Руководство по эксплуатации

Габариты в упаковке, м

0,4х0,24х0,15

Количество изделий в упаковке

1

Упаковка

картонная коробка

Дополнительная упаковка

требуется

Понравилась статья? Поделить с друзьями:
  • Домик лего френдс для девочек инструкция
  • Стиральная машина beko wkb 61001 y инструкция по применению
  • Руководство по системной поведенческой психотерапии купить
  • Мануал по ремонту electrolux
  • Флостерон уколы инструкция по применению цена отзывы аналоги уколы цена